Sekonic L-246 Bedienungsanleitung


Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Sekonic L-246 (4 Seiten) in der Kategorie Lichtmessgerät. Dieser Bedienungsanleitung war für 10 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet

Seite 1/4
Specifications
Measuring Range of Illuminance:
Low: 0 -1000 Lx (Red scale)
Ordinary: 0-5000 Lx (Black scale)
High: 0-50000 Lx (by using the Slide of x10 multiplier)
Accuracy : ± 10% (tested by a standard parallel-light tungsten lamp of color
temperature 2854°K) without including the resolution for
the meter (the value deserves minimal scale 1/2)in this
accuracy.
Angle Allowance : 30'--- - --- -Iess than 3% 60° less than 10%
Correction of sight sensitivity
0- -1000 Lx (No correction light source factor used
0-5000 Lx (Correction used)
Power Source : No battery used
Condition of Use : Temperature: 5 - - 35° C Humidity: 45 85
Accessories : Lumidisc for low measurement
Lumidisc for ordinary measurement
Slide ( x10 multiplier)
Dimensions : 112X58X27mm Weight : 135g
Characteristics : Compact, light weight, handy to carry, and most easy - -to use
swivel-head system
Name of
Parts
How to operate the Illuminometer
1. Getting ready for measurement
All you do for preparation is just to insert the LUMIDISC into position. To insert it,
match the white dot (or the red dot for low measurement) on the lumidisc to the white
dot on the swivel-head of the meter. Turn the lumidisc clockwise lightly to about 45°
until it tightly fits.
2. Ordinary measurement: 0--5000 Lx
To measure the range of 0--5000 Lx, use the lumidisc for ordinary measurement (with
a white dot at the outer ring of the disc and a green filter attached at the back).
Insert it into position as mentioned in 1. "Getting ready". Then place the meter at the
surface to be measured so that the lumidisc is parallel to the surface. Read the BLACK
figure on the scale. The figure the pointer indicates is the correct lux to use.
3. Measurement of High range: 0 --50000 Lx
To measure the range of 0--50000 Lx, use the lumidisc for ordinary measurement (see
2. above). Insert it into position in the same procedure as in 2. Then INSERT THE
SLIDE (x10 multiplier) into the slit located at the top of the swivel-head. Measure in
the same way as mentioned in 2. "Ordinary measurement". Read the BLACK figure the
pointer indicates, then MULTIPLY the figure by 10 to get the correct lux.
4. Measurement of Low range: 0--1000 Lx
To measure the range of 0-1000 Lx, use the lumidisc for low measurement (with a red
dot at the outer ring of the disc and no filter at the back). Insert it into position and
measure in the same way as in 2. Read the RED figure on the scale. Then MULTIPLY
the figure by the correction factor of each light source shown in the table below.
Correction factor
Mercury arc lamp...............................1.1
Fluorescent lamp ...............................1.0
Incandescent light..............................1.0
Day light ..........................................0.8
Example
In case the pointer indicates 600 Lx
under mercury arc lamp, the correct
lux is, 600 Lx X 1.1-=660 Lx


Produktspezifikationen

Marke: Sekonic
Kategorie: Lichtmessgerät
Modell: L-246

Brauchst du Hilfe?

Wenn Sie Hilfe mit Sekonic L-246 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten