Quantum Photo-Meter LX Bedienungsanleitung
Quantum
Lichtmessgerät
Photo-Meter LX
Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Quantum Photo-Meter LX (5 Seiten) in der Kategorie Lichtmessgerät. Dieser Bedienungsanleitung war für 14 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet
Seite 1/5
QUANTUM PHOTO-METER
models PME, PMEX, PML, PMLX
OPERATING INSTRUCTIONS
1. MODEL DESIGNATIONS
PME: Photometer for general measurements. Measures in footcandles.
Optional accessory available: Luminance Receptor PM13 calibrated for footlamberts.
PML: Photometer for general survey measurements. Measures in lux.
Optional accessory available: Luminance Receptor PM13 calibrated for nits (candela/m2).
PMEX: Precision photometer for scientific measurements. Color corrected to the C.I.E photopic response.
Pre-calibrated for foot-candles and footlamberts.
Optional accessories available: Rotating Illuminance Receptor PM11, Rigid Fiber Optic Probe
PM10, and Flexible Fiber Optic Probe PM12.
PMLX: Precision photometer for scientific measurements. Color corrected to the C.I.E photopic response.
Pre-calibrated for lux and nits (candela/m2).
Optional accessories available: Rotating Illuminance Receptor PM11 for lux, Rigid Fiber Optic
Probe PM10 for nits, and Flexible Fiber Optic Probe PM12 for nits.
FOR USE WITH
PME / PML |
|
|
|
|
|
|
|
|
|
|
|
|
|
FOR USE WITH
PMEX / PMLX
PM-13 PM-10 PM-11 PM-12
2. MEASURE CONTROL
The Measure control is located on the right side of Photo-Meter. Press and hold the Measure control to take a
reading. Release the control to store the reading in memory and display it for a few seconds.
If the display shows only three dash lines, the light intensity is out of range. The Range control is on the left
side of Photo-Meter. Press it repeatedly, while holding Measure, until numbers appear in the display.
Over Range Typical Reading
The display will stay lit for a few seconds, after releasing Measure, with the last reading frozen in the display.
3. RANGE CONTROL
The Range button on the left side of Photo-Meter sets the readout range on the digital display. Each time the
Range button is pressed the meter switches to the next higher range. At the highest range, the next Range
button press switches to the lowest (most sensitive) range.
The table below illustrates how to cycle through the ranges. Each arrow signifies the action when Range is
depressed.
The quotes ("xxx") indicate a fast blinking display. When that occurs, multiply the display number by 1000
for the correct reading. For example, a blinking "6.43" signifies a reading of 6,430.
The units of measurement are indicated by the turret tabs, as read from the front of the Photo-Meter.
4. MEASURING MODE
For Models PMEX, PMLX:
The turret on top of the instrument may be set for illuminance
or luminance measurements. To change the mode of
measurement lift off the turret, rotate it 180°, and press it back
into place.
The turret has two labels on tabs that identify the mode of
measurement. The tab next to the front of the instrument is the
current mode. Model PMEX can be set for foot-candles or
foot-Lamberts. Model PMLX can be set for lux or nits
(candela/m2).
For Models PME, PML:
These models come with a single mode turret calibrated for illuminance: PME: foot-candles; PML: lux. They
may be calibrated also for luminance measurements using Luminance Receptor PM13 as an optional
accessory. Factory re-calibration is required.
5. MEASURE TECHNIQUE
Illuminance - foot-candles or lux
This is the amount of light energy incident upon a plane. The plane is defined by the geometric plane of the
white disc on the turret. Light sources within the (180°) hemisphere in front of the white disc will be sensed
according to the cosine law.
For best accuracy do not block light sources in the 180° hemisphere. The orientation of the white disc is
designed to minimize the effect of the operator's presence on the measurement.
COSINE LAW: RELATIVE SENSITIVITY
VERSUS ANGLE OF INCIDENT LIGHT
Luminance - foot-lamberts or nits (candela/m2)
Luminance is the photometric brightness of sources or light reflected from objects. Photo-Meters read the
overall luminance in approximately a 25° field of view, as shown.
RELATIVE SENSITIVITY
VERSUS ANGLE
When attempting luminance readings avoid casting shadows upon the subjects when those are reflecting
objects (i.e., not light sources).
6. MISCELLANEOUS
Photo-Meters are very sensitive instruments that respond to slight changes in illumination. This is especially
true when measuring fluorescent lights.
Fluorescent illumination varies in brightness at twice the rate of the AC line frequency (60 hz). This variation
is not apparent to the human eye, but it will be apparent in readings with Photo-Meters. A slight drift in the
reading of the instruments might indicate the condition of fluorescent illumination. The average of the highest
Produktspezifikationen
Marke: | Quantum |
Kategorie: | Lichtmessgerät |
Modell: | Photo-Meter LX |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Quantum Photo-Meter LX benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung Lichtmessgerät Quantum
24 Mai 2023
Bedienungsanleitung Lichtmessgerät
- Lichtmessgerät Voltcraft
- Lichtmessgerät Velleman
- Lichtmessgerät Pyle
- Lichtmessgerät Extech
- Lichtmessgerät Gossen
- Lichtmessgerät Sekonic
Neueste Bedienungsanleitung fĂĽr -Kategorien-
10 Juni 2024
31 März 2024
30 Januar 2024
22 Januar 2024
31 Dezember 2023
25 Dezember 2023
2 Dezember 2023
30 November 2023
18 November 2023
18 September 2023