Klein Tools MM2300 Bedienungsanleitung
Klein Tools
Multimeter
MM2300
Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Klein Tools MM2300 (9 Seiten) in der Kategorie Multimeter. Dieser Bedienungsanleitung war für 31 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet
Seite 1/9
CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
customerservice@kleintools.com
www.kleintools.com
ENGLISH
Instruction Manual
ENGLISH
• TRMS
• BACKLIGHT
• BAR GRAPH
• LEAD HOLDER
• MAX / MIN
• 3-5/6 DIGIT
6000 COUNT LCD
• AUTO / MANUAL
RANGE
• AUTO HOLD
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range Resolution Accuracy
600mV ˜ 600V 0.1mV ˜ 0.1V ± (0.5% + 4 digits)
1000V 1V ± (0.8% + 10 digits)
Overload Protection: Input Impedance (Nominal) 1000V : 10M , <100pFΩ
Input Impedance (Low Z): 3k , <200pF, up to 600V RMSΩ
AC Voltage Measurement
Range Resolution Accuracy
40Hz ~ 400Hz
Accuracy
400Hz ~ 5kHz
600mV 0.1mV
± (0.75% + 5 digits)
± (2.0% +3 digits)
6V ~ 600V ± (2.0% +3 digits)1mV ~ 0.1V
1000V ± (0.75% + 8 digits) ± (2.0% +3 digits)*1V
continued on next page.
Overload Protection: Input Impedance (Nominal) 1000V RMS : 10M , <100pF Ω
Input Impedance (Low Z): 3k , <200pF, up to 600V RMSΩ
Frequency Response: 40 to 5kHz Response: True RMS
*Accuracy for 400Hz ~ 1kHz range only.
DC Current Measurement
Range Resolution Accuracy
600μA ~ 400mA 0.1μA ~ 0.1mA ± (1.0% + 5 digits)
6A ~ 10A 1mA ~ 10mA ± (1.5% + 5 digits)
Overload Protection: µAmA: A Input: F440mA/1000V fuse / F11A/1000V fuse
AC Current Measurement
Range Resolution Accuracy
40Hz ~ 2kHz
600μA ~ 400mA 0.1μA ~ 0.1mA ± (1.2% + 5 digits)
6A ~ 10A 1mA ~ 10mA ± (1.8% + 5 digits)
Overload Protection: µAmA: A Input: F440mA/1000V fuse / F11A/1000V fuse
Frequency: Response: 40 to 2kHz True RMS
Resistance Measurement
Range Resolution Accuracy
600 Ω˜ 6M 0.1 Ω Ω ˜ 0.001M ± (1.0% + 5 digits)Ω
60M 0.01M ± (1.5% + 10 digits)Ω Ω
Overload Protection: 600V RMS
Capacitance Measurement
Range Resolution Accuracy
60nF 0.01nF
± (3.5% + 6 digits)
600nF ~ 6000μF 0.1nF ~ 1μF
Overload Protection: 600V RMS
Frequency Measurement
Range Resolution Accuracy
99.99Hz ˜ 499.9kHz 0.01Hz ˜ 0.1kHz ± (0.1% + 3 digits)
Overload Protection: Sensitivity: 600V RMS > 0.7V RMS
Duty Cycle Measurement
Range Resolution Accuracy
0.1 ˜ 99.9% 0.1% ± (0.2% per kHz + 0.1% + 5 digits)
Overload Protection: Sensitivity: 600V RMS < 50V
Frequency Range: <10kHz
Diode Test
Overload Protection Range Test Current Open Circuit Voltage
600V RMS 6.0V Appx. 0.25mA < 1.8V DC (typ.)
Continuity Test
Overload Protection Open Circuit Voltage Threshold (Appx.)
600V RMS < 0.44V <40Ω
Temperature Measurement
Range Resolution Accuracy
-58 ~ 600°F 0.1°F ± (1.0% + 5.4°F)
-50 ~ 600°C 0.1°C ± (1.0% + 3.0°C)
600 ~ 1832°F 1°F ± (1.0% + 5.4°F)
600 ~ 1000°C 1°C ± (1.0% + 3.0°C)
Overload Protection: Thermocouple Accuracy: 600V RMS Not specifi ed
Specifi cations subject to change
Do Not apply voltage to thermocouple.
• -58° to 1832°F (-50° to 1000°C)
8. Temperature:
Features:
Testing Fuse:
• When selecting uA or mA/A functions the meter will indicate
“ ” if the test leads are not properly connected to the meter. LEAd
• Note: This message remaining on the screen after properly
connecting the test leads is an indication that the fuse is open.
Rev B 04/14 139716
7. Frequency (Hz) / Duty Cycle: <500kHz
Features:
HOLD
6. Low Impedance (Low Z): < 600V
DO NOT use (LOW Z) mode at voltages greater than 600V.
• Press SELECT to choose AC or DC Voltage.
• Use Low Impedance (LOW Z) mode to reduce ghost voltages
• See AC/DC Voltage instruction
Capacitance Features:
HOLD
Safely discharge capacitor before
measurement.
• Reading may take up to 60 seconds
for large capacitors.
SYMBOLS USED ON LCD
~
AC Measurement DC Measurement
-
Negative DC Value
AT
Auto Range Active
O
.L.
Overload: Range Exceeded Apo
Auto Power-Off Active
Low Battery
HOLD
Hold Active
MIN
Minimum Reading
M
AX
Maximum Reading
%
Duty Cycle Mode Frequency Mode
V
Voltage Measurement
A
Current in Amps
Ω
Resistance in Ohms Diode Test
F
Capacitance in Farads Continuity Test
Relative / Zero Mode
n
Nano 10
-9
µ
Micro 10
-6
m
Milli 10
-3
k
Kilo 10
3
M
Mega 10
6
°F
Degrees Fahrenheit
°C
Degrees Celsius
A-HOLD
Auto-Hold Active
LOW
Z
Low Impedance
AUTO HOLD MAX/MIN RELHOLD WARRANTY
www.kleintools.com/warranty
CLEANING
Clean the instrument by using a damp cloth. Do not use abrasive cleaners
or solvents.
Take care to keep the sensor lens clean at all times. Clean the lens using a soft
cloth or cotton swab with water or rubbing alcohol only, and allow the lens to
dry before use.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifi cations
section, allow the instrument to return to normal operating conditions before
using it.
DISPOSAL / RECYCLE
Do not place equipment and its accessories in the trash. Items
must be properly disposed of in accordance with local regulations.
1000V
10A
MM2300
ENGLISH
FEATURE DETAILS
MM2300
Instruction Manual
A. Do not attempt to measure more than 1000V.
B. Do not attempt to measure more than 600V on (LOW Z) setting.
C. Do not attempt to measure more than 400mA.
D. Do not attempt to measure more than 10A.
E. Auto Power-Off (Apo)
• Device will power off after 30 minutes non-use.
• Press the select button to wake.
• Disabled during Max/Min function.
• Holding Select button while turning on disables Auto Power-Off.
F. Select Functionality Button
• Switch between AC and DC.
• Switch between “ ”, “Ω”, “ ”, and “ ”.
• Switch between Hz and %.
• Switch between °F and °C.
• Wake meter during Auto Power Off.
G. Relative Reading Mode
• Press to store current value.
• Display shows the difference between the stored and live readings.
• Press again to return to live reading.
H. Auto / Manual Range
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
• AT is displayed on LCD only during auto ranging mode.
I. Max/Min Hold
• Press to enter Max/Min mode; the largest and smallest values will be
saved while in this mode.
• Press repeatedly to alternative between the maximum and minimum
readings.
• Press for 2 seconds to return to live reading and clear the stored
maximum and minimum values.
J. Hold/Auto Hold/Backlight
• Hold captures the current displayed value.
• Auto Hold captures the first stable displayed value, until a new stable
value is measured. The meter will then capture the new value and emit
a beep (V, Ohm, Temperature and Amp functions).
• Press to hold current input on the display.
• Press again to enter Auto-Hold to capture stable values.
• Press again to return to normal operation.
• Press and hold to activate display backlight.
• Using lights drains the battery significantly.
K. L. Battery/Fuse Replacement
• When indicator is displayed on the LCD, batteries must be
replaced.
• Remove rubber boot, back screw and replace 2 x AAA batteries.
• This meter uses 440mA/1000V (69192) and 11A/1000V (69191)
fast blow fuses.
M. Magnetic Hanger Accessory (Optional, Sold Separately)
• Slide magnetic adapter into protective rubber boot.
• Attach instruments to metal for hands-free use
N. Bar Graph
• The bar graph shows an approximate analog representation
of a measurement.
• The bar graph responds much faster than the digital display.
• The scale of the bar graph is zero to the maximum reading of
the selected range.
FUNCTION INSTRUCTIONS
1. AC Voltage: <1000V. (Low Z) < 600V
Features:
MAX/MINRANGEHOLDREL
Do not attempt to measure more than 1000V
DO NOT use (LOW Z) mode at voltages greater than 600V.
• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "AC" Voltage position.
• Attach red lead to “V” input.
2. DC Voltage: <1000V. (Low Z) < 600V
Features:
MAX/MINRANGEHOLDREL
4. AC/DC Current (small): < 400mA
Features:
Do not attempt to measure more than 400mA
• Attach red lead to “mAμA” input.
• Select AC or DC current source.
5. Resistance/Diode/Continuity/Capacitance
AUTO/HOLD
0 10 20 30 40 50 60
MM2300
®
C
E
A
G
F
D
H
J
I
M
L
K
N
B
Continuity Features:
MAX/MINHOLD
• Display shows resistance.
• Buzzer sounds if less than 40Ω
Diode Features:
MAX/MINHOLD
Display shows:
• Forward voltage drop if forward biased.
• “O.L” if reverse biased.
AUTO HOLD
AUTO HOLD
AUTO HOLD MAX/MINRANGEHOLDREL
Resistance Features:
•
measure resistance on a live Do not
circuit.
• = < 60MΩ Ω
AUTO HOLDHOLDREL
RANGE MAX/MIN
5
1
8
3
4
6
7
2
Do not attempt to measure more than 10A
• Start with this setting if current level is unknown.
• Meter will automatically select "A" when red lead is connected to
"A" input. “ ” will show if connected but fuse is blown.LEAd
3. AC/DC Current (large): <10A
Features:
MAX/MINRANGEHOLDREL
Do not attempt to measure more than 1000V
DO NOT use (LOW Z) mode at voltages greater than 600V.
• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "DC" Voltage position.
• Attach red lead to “V” input.
AUTO HOLD
GENERAL SPECIFICATIONS
The Klein Tools MM1300 is an auto-ranging multimeter. It measures AC/DC
voltage, AC/DC current, resistance, capacitance, frequency, duty cycle, and
temperature. It can also test diodes and continuity.
• Operating Altitude: / 6572 ft 2000m
• Relative Humidity: <75%
• Operating Temperature: 32°F to 122°F (0°C to 50°C)
• Storage temperature: -4°F to 140°F (-20°C to 60°C) <80% R.H.
• Accuracy Temperature: 64°F to 82°F (18°C to 28°C)
• Temperature Coefficient: 0.1*(specified accuracy)/ °C
• Sampling Frequency: 3 samples per second
• Dimensions: 7" x 3.5" x 1.9" (178 mm x 89 mm x 229 mm)
• Weight: 14 oz. (397 g)
• Drop Protection: 2 m
• Ingress Protection: 20
• Calibration: Accurate for one year
• Safety Rating: CAT III 1000V, CAT IV 600V
• Accuracy: ± (% of reading + # of least significant digits)
WARNINGS
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe theses warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring a known
voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms, or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from the
metal probe contacts when making measurements.
• Do not open the meter to replace batteries or fuses while the probes
are connected.
• Use caution when working with voltages above 60V DC, or 25V AC
RMS. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock, replace
batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out power
before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use individual
protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.
SYMBOLS
~
AC Alternating Current Warning or Caution
DC Direct Current Dangerous levels
DC/AC Voltage or Current Double Insulated Class II
Ground AC Source
SERVICIO AL CLIENTE
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069, EE. UU.
1-877-775-5346
customerservice@kleintools.com
www.kleintools.com
Manual de
instrucciones
ESPAÑOL
• TRMS
• RETROILUMINACIÓN
• GRÁFICO DE
BARRAS
• PORTACABLES
• MÁX/MÍN
• PANTALLA LCD DE
3-5/6 DÍGITOS CON
RECUENTO DE 6000
• RANGO
AUTOMÁTICO/
MANUAL
• AUTORRETENCIÓN
ESPECIFICACIONES ELÉCTRICAS
Medición de voltaje de CD
Rango Resolución Precisión
600 mV ˜ 600 V 0,1 mV ˜ 0,1 V ± (0,5 % + 4 dígitos)
1000 V 1 V ± (0,8 % + 10 dígitos)
Protección contra sobrecarga: Impedancia de entrada (Nominal) 1000 V : 10 M , < 100 pFΩ
Impedancia de entrada (Low Z): > 3 k , < 200 pF, hasta 600 V RMSΩ
Medición de voltaje de CA
Rango Resolución Precisión
40 Hz ~ 400 Hz
Precisión
400 Hz ~ 5 kHz
600 mV 0,1 mV ± (0,75 % + 5 dígitos) ± (2,0 % + 3 dígitos)
6 V ~ 600 V ± (2,0 % + 3 dígitos)1 mV ~ 0,1 V
1000 V ± (0,75 % + 8 dígitos) ± (2,0 % + 3 dígitos)*1 V
Continúa en la página siguiente.
Protección contra sobrecarga: 1000 V RMS
Impedancia de entrada (Nominal): 10 M , < 100 pF Ω
Impedancia de entrada (Low Z): > 3 k , < 200 pF, hasta 600 V RMSΩ
Respuesta de frecuencia: 40 a 5 kHz Respuesta: RMS real
*Precisión solo para rango 400 Hz ~ 1 kHz.
Medición de corriente CD
Rango Resolución Precisión
600 μA ~ 400 mA 0,1 μA ~ 0,1 mA ± (1,0 % + 5 dígitos)
6 A ~ 10 A 1 mA ~ 10 mA ± (1,5 % + 5 dígitos)
Protección contra sobrecarga: µAmA: Entrada A: Fusible F440 mA/1000 V/ Fusible F11 A/1000 V
Medición de corriente CA
Rango Resolución Precisión
40 Hz ~ 2 kHz
600 μA ~ 400 mA 0,1 μA ~ 0,1 mA ± (1,2 % + 5 dígitos)
6 A ~ 10 A 1 mA ~ 10 mA ± (1,8 % + 5 dígitos)
Protección contra sobrecarga: µAmA: Entrada A: Fusible F440 mA/1000 V/ Fusible F11 A/1000 V
Frecuencia: Respuesta: 40 Hz a 2 kHz RMS real
Medición de resistencia
Rango Resolución Precisión
600 Ω ˜ 6 M 0,1 Ω Ω ˜ 0,001 MΩ ± (1,0 % + 5 dígitos)
60 M 0,01 M ± (1,5 % + 10 dígitos)Ω Ω
Protección contra sobrecarga: 600 V RMS
Medición de capacitancia
Rango Resolución Precisión
60 nF 0,01 nF ± (3,5 % + 6 dígitos)
600 nF ~ 6000 μF 0,1 nF ~ 1 μF
Protección contra sobrecarga: 600 V RMS
Medición de frecuencia
Rango Resolución Precisión
99,99 Hz ˜ 499,9 kHz 0,01 Hz ˜ 0,1 kHz ± (0,1 % + 3 dígitos)
Protección contra sobrecarga: Sensibilidad: 600 V RMS > 0,7 V RMS
Medición de ciclo de servicio
Rango Resolución Precisión
0,1 ˜ 99,9 % 0,1 % ± (0,2 % por kHz + 0,1 % + 5 dígitos)
Protección contra sobrecarga: Sensibilidad: 600 V RMS < 50 V
Intervalo de frecuencia: < 10 kHz
Prueba de diodo
Protección contra
sobrecarga Rango Corriente de
prueba Voltaje de circuito abierto
600 V RMS 6,0 V Aprox. 0,25 mA < 1,8 V DC (típ.)
Prueba de continuidad
Protección contra
sobrecarga Voltaje de
circuito abierto Umbral (aprox.)
600 V RMS < 0,44 V < 40 Ω
Medición de temperatura
Rango Resolución Precisión
-58 °F ~ 600 °F 0,1 °F ± (1,0 % + 5,4 °F)
-50 °C ~ 600 °C 0,1 °C ± (1,0 % + 3,0 °C)
600 °F ~ 1832 °F 1 °F ± (1,0 % + 5,4 °F)
600 °C ~ 1000 °C 1 °C ± (1,0 % + 3,0 °C)
Protección contra sobrecarga: Precisión del termopar: 600 V RMS no especifi cada
Especifi caciones sujetas a cambios
NO aplique voltaje al termopar.
• -58 °F a 1832 °F (-50 °C a 1000 °C)
8. Temperatura:
Funciones:
Fusible de prueba:
• Cuando se seleccionan las funciones μA o mA/A, el multímetro
indicará “ ” si los cables de prueba no están conectados LEAd
correctamente al multímetro.
• Nota: si este mensaje permanece en la pantalla después de haber
conectado los cables de prueba correctamente, indica que el
fusible está abierto.
Rev B 04/14 139717
7. Frecuencia (Hz)/Ciclo de servicio: < 500 kHz
Funciones:
HOLD
6. Baja Impedancia (Low Z): < 600 V
NO utilice el modo de baja impedancia (LOW Z) a voltajes
mayores que 600 V.
• Presione SELECT (SELECCIONAR) para elegir entre voltaje CA o CD.
• Utilice el modo de baja impedancia (LOW Z) para reducir los
voltajes fantasmas.
• Consulte las instrucciones sobre Voltaje CA/CD.
Funciones de capacitancia:
HOLD
Descargue el capacitor de manera
segura antes de la medición.
• La lectura puede tardar hasta
60 segundos en el caso de
capacitores grandes.
SÍMBOLOS QUE SE UTILIZAN EN LA PANTALLA LCD
~
Medición de CA Medición de CD
-
Valor negativo de CD
AT
Rango automático
activo
O
.L.
Sobrecarga: rango
excedido
Apo
Apagado automático
activo
Batería baja
HOLD
Retención activa
MIN
Lectura mínima
M
AX
Lectura máxima
%
Modo Ciclo de servicio Modo Frecuencia
V
Medición de voltaje
A
Corriente en amperios
Ω
Resistencia en ohmios Prueba de diodo
F
Capacitancia en
faradios Prueba de continuidad
Modo relativo/cero
n
Nano 10
-9
µ
Micro 10
-6
m
Mili 10
-3
k
Kilo 10
3
M
Mega 10
6
°F
Grados Fahrenheit
°C
Grados Celsius
A-HOLD
Autorretención activa
LOW
Z
Baja impedancia
AUTO HOLD MAX/MIN RELHOLD GARANTÍA
www.kleintools.com/warranty
LIMPIEZA
Limpie el instrumento con un paño húmedo. No utilice solventes ni limpiadores
abrasivos.
Mantenga la lente del sensor siempre limpia. Limpie la lente con un paño suave
o hisopo de algodón embebidos en agua, o frote solo con alcohol, y deje secar
la lente antes del uso.
ALMACENAMIENTO
Retire las baterías si no va a utilizar el instrumento durante un tiempo
prolongado. No lo exponga a la humedad ni a altas temperaturas. Luego de un
período de almacenamiento en condiciones extremas que sobrepasen los límites
mencionados en la sección Especifi caciones, deje que el instrumento vuelva a
las condiciones de funcionamiento normales antes de utilizarlo.
ELIMINACIÓN/RECICLAJE
No arroje el equipo ni sus accesorios a la basura. Los elementos
se deben desechar correctamente de acuerdo con las regulaciones
locales.
1000 V
10 A
MM2300
Produktspezifikationen
Marke: | Klein Tools |
Kategorie: | Multimeter |
Modell: | MM2300 |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Klein Tools MM2300 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung Multimeter Klein Tools
28 August 2024
27 August 2024
19 August 2024
1 Mai 2024
31 März 2024
15 März 2024
Bedienungsanleitung Multimeter
- Multimeter Basetech
- Multimeter Gembird
- Multimeter Brennenstuhl
- Multimeter Clas Ohlson
- Multimeter Profile
- Multimeter REV
- Multimeter Voltcraft
- Multimeter Bosch
- Multimeter Laserliner
- Multimeter Velleman
- Multimeter Powerfix
- Multimeter Alecto
- Multimeter EMOS
- Multimeter TFA
- Multimeter FERM
- Multimeter Topcraft
- Multimeter Schneider
- Multimeter Tacklife
- Multimeter Abus
- Multimeter Elro
- Multimeter Stanley
- Multimeter Perel
- Multimeter Wago
- Multimeter Workzone
- Multimeter Somogyi
- Multimeter Amprobe
- Multimeter Metrix
- Multimeter Maxwell
- Multimeter Silverline
- Multimeter Yato
- Multimeter Brandson
- Multimeter Keithley
- Multimeter Hager
- Multimeter Kyoritsu
- Multimeter GW Instek
- Multimeter Monacor
- Multimeter Brüder Mannesmann
- Multimeter Milwaukee
- Multimeter JUNG
- Multimeter Testo
- Multimeter Elma
- Multimeter IDEAL
- Multimeter Digitus
- Multimeter Kingcraft
- Multimeter Testboy
- Multimeter Projecta
- Multimeter Greenlee
- Multimeter Multimetrix
- Multimeter Extech
- Multimeter BENNING
- Multimeter Metrel
- Multimeter Högert
- Multimeter APPA
- Multimeter Fluke
- Multimeter FLIR
- Multimeter Aldi
- Multimeter Gossen Metrawatt
- Multimeter Hubinont
- Multimeter Joy-it
- Multimeter Kewtech
- Multimeter Mastech
- Multimeter Megger
- Multimeter Micronta
- Multimeter PCE Instruments
- Multimeter Plieger
- Multimeter Sanwa
- Multimeter Skandia
- Multimeter Sonel
- Multimeter Steren
- Multimeter Uni-T
- Multimeter Wiha
- Multimeter Vimar
- Multimeter Owon
- Multimeter PeakTech
- Multimeter Amiko
- Multimeter Rigol
- Multimeter HT Instruments
- Multimeter Beha-Amprobe
- Multimeter Chauvin Arnoux
- Multimeter Aim TTi
- Multimeter Etekcity
- Multimeter Rohde & Schwarz
- Multimeter Strex
- Multimeter MBS
- Multimeter Kurth Electronic
- Multimeter Weidmüller
- Multimeter CEM
- Multimeter Dasqua
- Multimeter Horex
- Multimeter Enovalab
- Multimeter Qian
- Multimeter Digi-tool
- Multimeter MGL Avionics
- Multimeter Cosinus
- Multimeter Weltron
- Multimeter EEVBlog
- Multimeter Caltek
- Multimeter Elworks
- Multimeter Bearware
- Multimeter Noyafa
Neueste Bedienungsanleitung für -Kategorien-
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
13 Oktober 2024
9 Oktober 2024
5 Oktober 2024
5 Oktober 2024
3 Oktober 2024