Rossmax SB210 Bedienungsanleitung
Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Rossmax SB210 (2 Seiten) in der Kategorie Impuls. Dieser Bedienungsanleitung war für 7 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet
Seite 1/2
• Finger is too large or too small to fit into the device.
• Poor pulse quality
• Venous pulsations
• Anemia or low hemoglobin concentrations.
• Cardiogreen and other intravascular dyes
• Carboxyhemoglobin
• Methemoglobin
• Dysfunctional hemoglobin
• Artificial nails or fingernail polish
• On fingers with anatomical changes, oedemas, scars or burns.
- Using the device for long periods may cause pain for people
with circulatory disorders. Reposition the device at least once
every 4 hours to allow the patient’s skin to breath and to check
patient’s condition regularly.
- Do not use the device near flammable or explosive gas
mixtures.
- Do not use the device during an MRI or CT scan.
- The device may not work when circulation is reduced. Warm or
rub the finger, or re-position the device.
- This device is a precision electronic instrument and must be
repaired by qualified technical professionals. Field repair of the
device is not possible. Do no attempt to open the case or repair
the electronics. Opening the case may damage the device and
void the warranty.
- Do not overextend the device’s spring.
- A functional tester cannot be used to access the accuracy of a
pulse oximeter monitor.
- Do not self-diagnose or self-medicate on the basis of the
measurements without consulting your doctor. In particular, do
not start taking any new medication or change the type and/
or dosage of any existing medication without prior approval.
- Do not look directly inside the housing during the
measurement. The red light and the invisible infra-red light in
the pulse oximeter are harmful to your eyes.
- This device is not intended for use by people (including
children) with restricted physical, sensory or mental skills or a
lack of experience and/or a lack of knowledge, unless they are
supervised by a person who has responsibility for their safety
or they receive instructions from this person on how to use the
device. Children should be supervised around the device to
ensure they do not play with it.
- Neither of the displays for the pulse wave and pulse bar allows
the strength of the pulse or circulation to be evaluated at the
measurement site. Rather, they are exclusively used to display
the current signal variation at the measurement site and do
not enable reliable diagnostics for the pulse.
- The maximum temperature of sensors which the user will touch
might reach 43°C when operating in the 40°C environment.
Cleaning
Troubleshooting
Specification
1. Please clean the surface of the device before using. Wipe
the device with medical alcohol (70% isopropyl alcohol)
first, and then let it dry in air or clean it by dry clean fabric.
When cleaning the device with water, the water temperature
should be lower than 60°C
2. Using the medical alcohol to disinfect the product after use,
prevent from cross infection for next time use.
3. The best storage environment of the device is -25°C to 70°C
ambient temperature and not higher than 90% relative
humidity.
Symptoms Check points Corrections
SpO2 or pulse
rate cannot
displayed
Applied nger
improperly.
Place the nger
properly and try again.
SpO2 is too low to
detect
Try again; go to
consult with your
physician if you are
sure the device works
well.
SpO2 or pulse
rate are not
displayed stably
Applied nger
improperly.
Place the nger
properly and try again.
Finger is shaking or
body is moving. keep body steady
No display
when button is
pressed
Batteries run down Replace with new
batteries
Batteries not
inserted correctly. Re-insert batteries
The display
disappears
suddenly
The device will auto
power o when it
gets no signal.
Normal
Low battery Replace with new
batteries
Note: If the unit does not work, return it to your dealer.
Under no circumstance should you disassemble and
repair the unit by yourself.
SpO2
Measuring range 35%~99%, (the resolution is 1%).
Accuracy 70%~99%: ±2%, Below 35~69%: unspecied.
Optical Sensor Red light (wavelength is 660nm), Infrared
(wavelength is 905/880nm)
Pulse
Measuring range 30bpm~250bpm (the resolution is 1 bpm)
Accuracy ±3bpm
Power source AAA × 2 (Alkaline)
Battery life Continually for 16 hours with two alkaline
batteries
Operating
Condition
Temperature: 5°C~40°C (41°F ~ 104°F),
Relative Humidity: 15-95% (non condensing),
Atmospheric pressure: 700hPa ~ 1060hPa,
Attitude: -1,280 to 12,000 feet (-390m to
3,658m)
Storage /
Transportation
Condition
Temperature: -25°C~+70°C (-13°F ~ 158°F),
Relative humidity: 15-90%(non condensing),
Atmospheric pressure: 700hPa ~ 1060hPa,
Attitude: -1,280 to 12,000 feet (-390m to
3,658m)
Dimensions 63.5(L) × 34W) × 35(H) mm
Weight About 37g (without the batteries)
Standards IEC60601-1-2, Class B, IEC60601-1, Type BF,
ISO80601-2-61, IEC60601-1-11
Type BF applied parts
IP Classication IP22: Protection against harmful ingress of
water and particulate matter
Note: 1. Do not sterilize, autoclave or immerse this device in
liquid. Do not pour or spray any liquids onto the device.
2. Do not use caustic or abrasive cleaning agents, or
any cleaning agent containing ammonium chloride or
isopropyl alcohol.
Maintenance
Recommends user to return this device to the manufacturer
perform the following checks every 24 months.
- Inspect the equipment for mechanical and functional damage
or deterioration.
- Ensure all user interface keys and accessories function normally.
Model: SB210
www.rossmax.com
WARNING: The symbol on this product means that it's an
electronic product and following the European directive
2012/19/EU the electronic products have to be dispose on
your local recycling centre for safe treatment.
EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and the ME equipment
The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the Finger-tip pulse oximeter can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the Finger-tip pulse
oximeter as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmit-
ter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz , d=[3.5/
V1]√P 80 MHz to 800 MHz , d=[3.5/
E1]√P 800 MHz to 2,5 GHz , d=[3.5/
E1]√P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.7 3.7 7.37
100 11.67 11.67 23.33
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING and
are specied for use only in a shielded location
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6 3 Vrms
150 kHz to 80 MHz 3V Portable and mobile RF communications equipment should be used
no closer to any part of the EQUIPMENT or SYSTEM including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter. Interference
may occur in the vicinity of equipment marked with the following
symbol.
Radiated RF
IEC 61000-4-3 3 V/m
80 MHz to 2.5 GHz 3V/m
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air ±6 kV contact
±8 kV air Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be
at least 30 %.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power sup-
ply lines
±1 kV for input/output
lines
±2 kV for power sup-
ply lines Mains power quality should be that of a
typical commercial or hospital environment.
Surge IEC 61000-4-5 ± 1kV dierential mode
± 2kV common mode ± 1kV dierential mode
± 2kV common mode Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT(>95% dip in UT)
for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip in UT)
for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the EQUIPMENT or SYSTEM
requires continued operation during power
mains interruptions, it is recommended that
the EQUIPMENT or SYSTEM be powered
from an uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz) magnetic
eld IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic environment specied below. The customer or the user
of the Finger-tip pulse oximeter should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
CE emissions CISPR11 Group 1 The Finger-tip pulse oximeter uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The Finger-tip pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage uctuations/icker emissions
IEC 61000-3-3 Complies
Rossmax InnoTek Corp.
12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo N° 18, CP 29006, Málaga, Spain
Fingertip Pulse Oximeter
- The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of
such marks by Rossmax International Ltd. is under license. Other trademarks and trade names are those of their
respective owners.
- The Pulse Oximeter uses Bluetooth®(Bluetooth® low energy technology)
- Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a
service mark of Apple Inc.
- Google Play and the Google Play logo are trademarks of Google Inc.
IN0SB210000000021
RI_IB_SB210_EN_2Yr_
TP_ver2003
Warranty Card
This instrument is covered by a 2 year guarantee from the date of
purchase, batteries and accessories are not included. The guarantee
is valid only on presentation of the guarantee card completed by
the dealer confirming date of purchase or the receipt. Opening or
altering the instrument invalidates the guarantee. The guarantee
does not cover damage, accidents or non-compliance with the
instruction manual. Please contact your local seller/dealer or www.
rossmax.com.
Customer Name: ____________________________________
Address: ___________________________________________
Telephone: _________________________________________
E-mail address: ______________________________________
Product Information: ________________________________
Date of purchase: ____________________________________
Store where purchased:
____________________________________________________
RI_IB_SB210_EN_2Yr_TP_ver2003.indd 1 2020/8/27 下午 01:51:04
Introduction
ACT (Artery Check Technology)
Error code for your reference
Installing Batteries
Attaching the lanyard
How to use
Rossmax Fingertip Pulse Oximeter SB210 is used to measure
arterial oxygen saturation (% SpO2) of hemoglobin and pulse
rate, an important indicator of your respiratory function. It is
non-invasive device intended for spot-check of adult and
pediatric whose age is over 3 at home, hospital and clinics.
Attention: Consult the accompanying documents. Please
read this manual carefully before use. Please be sure to
keep this manual.
ACT processes the SpO2 signal and determines the elasticity
of blood vessel based on the derived wave form. It further
classifies the vascular status into 6 levels and presents the result
in an intuitive graphical interface.
Artery and blood circulation in good condition
Artery and blood circulation in good to aver-
age condition
Artery and blood circulation in average condi-
tion
Artery and blood circulation in below average
condition
Artery and blood circulation in poor condition
Artery and blood circulation in critical condi-
tion
Note: the classification of artery and blood circulation
condition is for reference only, Please consult with
your physician for further advice.
SENSOR ERROR:
Sensor cannot be detected, return the
device to your local distributor or ser-
vice centre.
MEASURE ERROR:
Signals cannot be detected, turn the de-
vice off and measure again.
AC ERROR:
Weak signal for artery check, turn the
device off and measure again.
1. Use thumb to slide battery cover out
2. Insert or replace 2 “AAA” sized batteries down with the correct
electrical polarity.
You need to replace the batteries when
1. LOW BATTERY appears on display
2. The function button is pressed and nothing appears on
display
Caution: Batteries may leak or explode if used or disposed of
improperly. Remove batteries if the device will be
stored for long time. Do not use different types or
brands of batteries at the same time. Do not mix fully
charged and partially charged batteries at the same
time.
1. Insert the narrow end of the lanyard through the holder.
2. Draw the other end of the lanyard through the loop at the
narrow end and tighten.
1. Open the clip; press the Power On/Off button as . Bluetooth
turns on automatically after power on.
2. Information of software version appears. Insert one finger(left
hand middle finger is recommended), nail side up, into the
finger opening of the pulse oximeter as .
Note: If no finger insert, the device will auto shut off after 30
seconds
3. The display shows , pulse oximeter begins its measurement.
Note: Make sure the finger is lying flat. Do not shake and keep
body steady during measurement.
Note: If artery check result cannot be detected, "" will
appear on the screen.
5. Remove the finger, the screen will show artery check level
as .
6. Press button shortly to switch the display to the 3 parameters
(artery check, SpO2 and pulse rate) as .
Note:
1. The SpO2 sensor and photoelectric receiving tube should be
arranged in a way with the subject’s arteriole in a position
there between.
2. Make sure the optical path is free from any optical obstacles
like rubberized fabric.
3. Excessive ambient light may affect the measuring result. It
includes fluorescent lamp, dual ruby light, infrared heater,
direct sunlight and etc.
4. Strenuous action of the subject or extreme electrosurgical
interference may also affect the accuracy.
5. The device will reflect the visual and audio signal when the
measurement of SpO2 is lower than the default value of 90%.
Cautionary Notes
Data Transfer via Bluetooth®
- This device is to be operated by trained personnel only.
- The device has intended only for spot-checking, but not
medical result evaluation.
- This device is designed to determine the percentage of arterial
oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of
the measurement include the following:
• Do not apply the pulse oximeter on the same arm as a blood
pressure cuff, arterial catheter or infusion line(s)
• Excessive light, such as sunlight or direct home lighting.
• Not steady at the site of application (e.g. trembling)
• Moisture in the device
• Improperly applied device
Pairing the Pulse Oximeter
with your Smartphone.
1. Bluetooth® turns on automati-
cally after device power on.
2. Download and install the free
APP onto your smartphone.
3. To pair this device with your
smartphone for data transmis-
sion.
APP
APP
Name/ Functions of each part
1. OLED display
2. Pulse strength
3. Pulse search icon
4. Artery check icon
5. Alarm icon
1
2
3
4
5
6
7
10
6. Battery indicator
7. Power On/O Button
8. SpO2 icon
9. Pulse icon
10. Battery cover
8
9
4. Your SpO2 and pulse rate values will appear on the screen
after few seconds and artery check result will appear on
screen after 30 - 60 seconds as .
RI_IB_SB210_EN_2Yr_TP_ver2003.indd 2 2020/8/27 下午 01:51:06
Produktspezifikationen
Marke: | Rossmax |
Kategorie: | Impuls |
Modell: | SB210 |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Rossmax SB210 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung Impuls Rossmax
21 Juni 2024
16 Juni 2024
16 Juni 2024
16 Juni 2024
15 Juni 2024
15 Juni 2024
30 April 2024
Bedienungsanleitung Impuls
- Impuls Medion
- Impuls Beurer
- Impuls Dexford
- Impuls Medisana
- Impuls Braun
- Impuls Polar
- Impuls Pyle
- Impuls Garmin
- Impuls Mio
- Impuls ProfiCare
- Impuls Homedics
- Impuls Bushnell
- Impuls Omron
- Impuls Terraillon
- Impuls IHealth
- Impuls Suunto
- Impuls Sigma
- Impuls Christopeit
- Impuls Domyos
- Impuls CicloSport
- Impuls Runtastic
- Impuls Magellan
- Impuls TomTom
- Impuls IGPSPORT
- Impuls Gima
- Impuls Oregon Scientific
- Impuls Digi-tech
- Impuls VirtuFit
- Impuls Ciclo
Neueste Bedienungsanleitung für -Kategorien-
12 Juni 2024
11 Juni 2024
11 Juni 2024
9 Juni 2024
18 Mai 2024
17 Mai 2024
29 April 2024
20 April 2024
16 April 2024
13 April 2024