Philips Respironics DreamStation Auto CPAP apparaat Bedienungsanleitung
Philips Respironics
CPAP-Geräte
DreamStation Auto CPAP apparaat
Lesen Sie kostenlos die đ deutsche Bedienungsanleitung fĂźr Philips Respironics DreamStation Auto CPAP apparaat (32 Seiten) in der Kategorie CPAP-Geräte. Dieser Bedienungsanleitung war fĂźr 7 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet
Seite 1/32
User manual
DreamStation
CPAP
CPAP Pro
Auto CPAP
Š Koninklijke Philips N.V., 2018. All rights reserved.
Table of Contents
Intended Use 1 ...........................................................................................................................................................
Important 1 .................................................................................................................................................................
Warnings 1 ..................................................................................................................................................................
Cautions 2 ...................................................................................................................................................................
Contraindications 2 ..................................................................................................................................................
Symbols Glossary 3 ...................................................................................................................................................
System Contents 4 ....................................................................................................................................................
How to Contact Philips Respironics 4 .................................................................................................................
System Overview 5 ...................................................................................................................................................
Installing/Replacing the Air Filters 6 ......................................................................................................................
Where to Place the Device 7 .................................................................................................................................
Supplying AC Power to the Device 7 ....................................................................................................................
Connecting the Breathing Circuit 8 ......................................................................................................................
Navigating the Device Screens 9 ............................................................................................................................
Starting the Device 9 ................................................................................................................................................
Menu Navigation (Therapy ON) and Optional Humidiîcation Settings .................................................10
Ramp Feature 10 ........................................................................................................................................................
Menu Navigation (Therapy OFF) 11 ......................................................................................................................
Wireless Technology .........................................................................................................................15
Check Mask Fit 16 .....................................................................................................................................................
Sleep Progress 16 ......................................................................................................................................................
Altitude Compensation 16 ......................................................................................................................................
Device Alerts ........................................................................................................................................................17
Troubleshooting 21 ...................................................................................................................................................
Accessories 22 ...........................................................................................................................................................
Traveling with the System 24 ..................................................................................................................................
Cleaning the Device 24 ............................................................................................................................................
Caring for the Filters 24 ..........................................................................................................................................
Cleaning the Tubing ..............................................................................................................................................25
Service 25 ....................................................................................................................................................................
Additional Notices 25 ...............................................................................................................................................
Speciîcations ........................................................................................................................................................26
Disposal 27 ..................................................................................................................................................................
EMC Information 27 .................................................................................................................................................
Limited Warranty ................................................................................................................................... Back Page
1User Manual
Caution: U. S. federal law restricts this device to sale by or on the order of a physician.
Intended Use
The Philips Respironics DreamStation systems deliver positive airway pressure therapy for the treatment of Obstructive Sleep Apnea in
spontaneously breathing patients weighing over 30 kg (66 lbs). It is for use in the home or hospital/institutional environment.
Important
The device is to be used only on the instruction of a licensed physician. Your home care provider will make the correct pressure
settings and device conîgurations including accessories, according to your health care professionalâs prescription.
Several accessories are available to make your OSA treatment with the DreamStation system as convenient and comfortable as
possible. To ensure that you receive the safe, effective therapy prescribed for you, use only Philips Respironics accessories.
Warnings
⢠This manual serves as a reference. The instructions in this manual are not intended to supersede the health care professionalâs
instructions regarding the use of the device.
⢠The operator should read and understand this entire manual before using the device.
⢠This device is not intended for life support.
⢠The device should be used only with masks and connectors recommended by Philips Respironics or with those recommended by the
health care professional or respiratory therapist. A mask should not be used unless the device is turned on and operating properly.
The exhalation port(s) associated with the mask should never be blocked. The device is intended Explanation of the Warning:
to be used with special masks or connectors that have exhalation ports to allow continuous îow of air out of the mask. When
the device is turned on and functioning properly, new air from the device îushes the exhaled air out through the mask exhalation
port. However, when the device is not operating, enough fresh air will not be provided through the mask, and exhaled air may be
rebreathed. Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suffocation.
⢠If you are using a full face mask (a mask covering both your mouth and your nose), the mask must be equipped with a safety
(entrainment) valve.
⢠When using oxygen with this system, the oxygen supply must comply with local regulations for medical oxygen.
⢠Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open îame.
⢠When using oxygen with this system, turn the device on before turning on the oxygen. Turn the oxygen off before turning the
device off. This will prevent oxygen accumulation in the device. Explanation of the Warning: When the device is not in
operation and the oxygen îow is left on, oxygen delivered into the tubing may accumulate within the deviceâs enclosure. Oxygen
accumulated in the device enclosure will create a risk of îre.
⢠When using oxygen with this system, a Philips Respironics Pressure Valve must be placed in-line with the patient circuit between
the device and the oxygen source. The pressure valve helps prevent the backîow of oxygen from the patient circuit into the
device when the unit is off. Failure to use the pressure valve could result in a îre hazard.
⢠Do not connect the device to an unregulated or high pressure oxygen source.
⢠Do not use the device in the presence of a îammable anaesthetic mixture in combination with oxygen or air, or in the presence
of nitrous oxide.
⢠Do not use the device near a source of toxic or harmful vapors.
⢠Do not use this device if the room temperature is warmer than 35° C (95° F). If the device is used at room temperatures warmer
than 35° C (95° F), the temperature of the airîow may exceed 43° C (109° F). This could cause irritation or injury to your airway.
⢠Do not operate the device in direct sunlight or near a heating appliance because these conditions can increase the temperature of
the air coming out of the device.
⢠Contact your health care professional if symptoms of sleep apnea recur.
⢠If you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, if it has been
dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, disconnect the power cord and
discontinue use. Contact your home care provider.
⢠Repairs and adjustments must be performed by Philips Respironics-authorized service personnel only. Unauthorized service could
cause injury, invalidate the warranty, or result in costly damage.
⢠Do not use any accessories, detachable parts, and materials not recommended by Philips Respironics. Incompatible parts or
accessories can result in degraded performance.
⢠Use only approved cables and accessories. Misuse may affect EMC performance and should be avoided.
⢠The Health Industry Manufacturers Association recommends that a minimum separation of six inches be maintained between
a wireless phone and a pacemaker to avoid potential interference with the pacemaker. The DreamStation on-board
communication should be considered a wireless phone in this regard.
⢠Use only power cords supplied by Philips Respironics for this device. Use of power cords not supplied by Philips Respironics may cause
overheating or damage to the device and may result in increased emissions or decreased immunity of the equipment or system.
⢠The device should not be used while stacked or in close approximation to other non-approved devices.
⢠Do not pull or stretch the tubing. This could result in circuit leaks.
Produktspezifikationen
Marke: | Philips Respironics |
Kategorie: | CPAP-Geräte |
Modell: | DreamStation Auto CPAP apparaat |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Philips Respironics DreamStation Auto CPAP apparaat benĂśtigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten