Kyoritsu 1110 Bedienungsanleitung
Kyoritsu
Multimeter
1110
Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Kyoritsu 1110 (2 Seiten) in der Kategorie Multimeter. Dieser Bedienungsanleitung war für 10 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet
Seite 1/2
INSTRUCTION MANUAL
M O D E L 1 1 1 0
KYORITSU MULTIMETER
KYORITSU ELECTRICAL INSTRUMENTS
WORKS,LTD.
1. Safety Warnings
This instrument has been designed and tested according to IEC Publication 61010:
Safety Requirements for Electronic Measuring Apparatus. This instruction manual
contains warnings and safety rules which must be observed by the user to ensure
safe operation of the instrument and retain it in safe condition. Therefore, read
through these operating instructions before using the instrument.
# WARNING
● Read through and understand instructions contained in this manual before
starting using the instrument.
● Save and keep the manual handy to enable quick reference whenever
necessary.
● Be sure to use the instrument only in its intended applications and to follow
measurement procedures described in the manual.
● Be sure to understand and follow all safety instructions contained in the
manual.
Failure to follow the instructions may cause injury, instrument damage and/or
damage to equipment under test. Kyoritsu is by no means liable for any
damage resulting from the instrument in contradiction to this cautionary note.
The symbol indicated on the instrument means that the user must refer to related #
parts in the manual for safe operation of the instrument. Be sure to carefully read
the instructions following each symbol in this manual. #
# DANGER is reserved for conditions and actions that are likely to cause
serious or fatal injury.
# WARNING is reserved for conditions and actions that can cause serious or
fatal injury.
# CAUTION is reserved for conditions and actions that can cause injury or
instrument damage.
# DANGER
● Make sure to set the function selector switch to the appropriate position
before making measurement. Use extreme caution not to avoid applying
voltage to the instrument with the range selector switch set to a current or
resistance range.
● Do not attempt to make measurement in the presence of flammable gasses,
fumes, vapor or dust. Otherwise, the use of the instrument may cause
sparking, which can lead to an explosion.
●Never attempt to use the instrument if its surface or your hand is wet.
● Do not exceed the maximum allowable input of the selected measuring
range.
●Never open the instrument when making measurement.
●The instrument should be used only in its intended applications or
conditions. Otherwise, safety functions equipped with the instrument do not
work, and instrument damage or serious personal injury may be caused.
●Verify proper operation on a known source before use or taking action as a
result indication of the instrument.
# WARNING
● Never attempt to make any measurement if any abnormal conditions are
noted, such as broken case, cracked test leads and exposed metal parts.
● Do not turn the function selector switch with test leads connected to the
instrument.
● Do not install substitute parts or make any modification to the instrument.
Return the instrument to Kyoritsu or your distributor for repair or
re-calibration.
● Do not try to replace the batteries or fuse if the surface of the instrument is
wet.
● Always set the function selector switch to the OFF position before opening
the instrument for battery replacement.
● Stop using the test lead if the outer jacket is damaged and the inner metal or
color jacket is exposed.
# CAUTION
● Always make sure to insert each plug of the test leads fully into the
appropriate terminal on the instrument.
● Be sure to set the function selector switch to the OFF position after use.
When the instrument will not be in use for a long period of time, place it in
storage after removing the batteries.
● Do not expose the instrument to the direct sun, extreme temperatures or
dewfall.
● Use a damp cloth and detergent for cleaning the instrument. Do not use
abrasives or solvents.
● This instrument isn't dust & water proofed. Keep away from dust and water.
● Keep your fingers and hands behind the protective fingerguard during
measurement.
Ranges
DC.V
AC.V
DC.A
Ω
LED 10mA approx. at 0 (at 3V of battery voltage)Ω
Buzzer beeps below about 100Ω
TEMP. 2※
BATT.TEST 1.5V 0.7 〜2.0V(Load resistance:10
Ω
approx.)
0 100〜℃
Ranges other than the above(-20〜150℃)
0.3V
3V
12V
30V
120V
300V
600V
12V
30V
120V
300V
600V
60 Aμ
30mA
300mA
×1
×10
×100
0〜0.3V (16.7kΩ/V)
0 3V〜
0 12V〜
0〜30V (20kΩ/V)
0 120V〜
0 300V〜
0 600V〜
0 12V 〜 ( )9kΩ/V
0 30V〜
0 120V 〜 ( )9kΩ/V
0 300V〜
0 600V〜
0 60 A〜 μ
(Terminal voltage:0.3V approx.)
0 30mA〜
(Terminal voltage:0.4V approx.)
0 300mA〜
(Terminal voltage:1V approx.)
※
1
0〜3kΩ(30Ωatmid-scale)
0〜30kΩ(300Ωatmid-scale)
0〜300kΩ(3kΩatmid-scale)
Measuring Range Accuracy
±3% of full scale value
±4% of full scale value
±3% of full scale value
±3% of full scale value
±3% of full scale length
±3% of scale length
±3% of scale length
±4% of scale length
2. Specifications
● ±Measuring ranges and accuracy (23 5 , 75%RH or less)℃
1 : Small differences may result depending on the resistance of the fuse.※
2 : The temperature probe (M-7060) has been discontinued. ※
●Standards: :IEC 61010-1
Measurement voltage CAT III 300V, CAT II 600V pollution
degree 2
IEC 61010-031
IEC 61326-1 (EMC), EN50581(RoHS)
● φFuse: Fast acting type(F500mA/600V), 6.3x32mm
●Overload Protection: AC/DC600V: fused (Current/Resistance/0.3V/BATT. TEST
1.5V ranges)
AC/DC720V: 10 seconds (600/300/120V ranges)
AC/DC120V: 10 seconds (30/12V ranges)
AC/DC30V:10 seconds(3V range)
●Withstand Voltage: AV3470V / 5 sedonds between internal circuit and housing
case
●Location for use: Altitude 2000m or less, Indoor use
●Drop Protection: From a height of 1m onto concrete floor
●
Operating Temperature
& Humidity: 0-40
℃
, relative humidity up to 85% without condensation
●Storage Temperature
& Humidity: -10-50
℃
, relative humidity up to 85% without condensation
● × ×Dimensions: 94(L) 140(W) 39(D)mm
●Weight: Approx. 280g(including batteries)
●Power Source: Two R6P(1.5V) or equivalent batteries
●Accessories: Test leads MODEL 7066A …………………1
R6P battery …………………………………2
Fuse (F500mA/600V) 2……………………
Carrying case 1………………………………
Instruction manual 1…………………………
3. Instrument Layout
Meter Cover
Pointer
Scale Plate
Meter Zero
Adjust Screw
Function Selector Switch
Front Panel Carrying Case
-Terminal
+Terminal
Ohm Zero
Adjust Knob
Test Leads
Protective fingerguard
Test Lead Tips
Measurement Category:
To ensure safe operation of measuring instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as O to CAT
IV, an d called me asurement c ategories . Higher-nu mbered cate gories
correspond to electrical environments with greater momentary energy, so a
measuring instrument designed for CAT III environments can endure greater
momentary energy than one designed for CAT II.
O : Circuits which are not directly connected to the mains power supply.
CAT II : Electrical circuits of equipment connected to an AC electrical outlet by
a power cord.
CAT III : Primary electrical circuits of the equipment connected directly to the
distribution panel, and feeders from the distribution panel to outlets.
CAT IV : The circuit from the service drop to the service entrance, and to the
power meter and primary over-current protection device (distribution
panel).
Protective fingerguard:It is a part providing protection against electrical shock
and ensuring the minimum required air and creepage distances.
Caution:Keep your fingers and hands behind the Protective fingerguard during
measurement.
Cap:
Uncapped condition for CAT II environment
Capped condition for CAT III/ IV environments
The Cap shuld be firmly attached to the probes.
4. How to Read Scales
※ :1 The thick portion of the scale indicates the allowable range of
voltage of a battery. (The lower limit of voltage for a 1.5V dry
battery specified by IEC 60086 is 0.9V.) Note that satisfactory
indication on the meter may not mean the battery has enough
power for high load (high current consumption) application.
Range
DC.V
AC.V
DC.A
Ω
BATT.TEST 1.5V 1 ※
TEMP.
LED
0.3V
3V
12V
30V
120V
300V
600V
12V
30V
120V
300V
600V
60 Aμ
30mA
300mA
×1
×10
×100
B
B
C
B
C
B
C
C
B
C
B
C
C
B
B
A
A
A
E
D
30
30
12
30
12
30
6
12
30
12
30
6
6
30
30
Ω
Ω
Ω
2.0
-
20 150〜
×0.01
×0.1
×1
×1
×10
×10
×100
×1
×1
×10
×10
×100
×10
×1
×10
×1
×10
×100
×1
×1
Scale Used
How to Read Scale
Regardless of indicated value
A
B
C
D
E
5. Preparation
●Checking Test Leads and Fuse
Plug the red test lead into the + terminal and the black test lead into the
-
terminal,
and set the function selector switch to a position in the area. Then, short the test Ω
lead tips. When the meter pointer deflects to the right, proceed to measurement.
When the meter pointer does not deflect, replace the fuse with the spare fuse,
which is supplied with the instrument. If there is still no deflection, the test leads
may have an open. Replace the test leads.
●Adjusting Meter Pointer Zero
Turn the zero adjust screw to set the meter pointer to the "0" mark on the
extreme left of the scale for accurate measurement.
●Checking Function Selector Switch Setting
Make sure to set the function selector switch to the appropriate position.
Otherwise, intended measurement cannot be made, or injury or instrument
damage may result.
When the order of voltage or current under test is unknown, first make
measurement on the highest range, and then select the appropriate range.
6. How to Make Measurements
Voltage Measurements (ACV, DCV)
# DANGER
●Do not make measurement on a circuit above 600V AC or DC.
●Do not apply voltage that exceeds the rated voltage of the selected range.
●Do not turn the function selector switch during measurement.
●Do not touch the metal parts of the test leads during measurement.
● When the order of the voltage under test is unknown, make measurement on
the highest range.
●Keep your fingers and hands behind the Protective fingerguard during
measurement.
(1)
Plug the red test lead into the + terminal and the black test lead into the
-
terminal.
(2)Set the function selector switch to the appropriate DCV or ACV position.
(3) Connect the test leads to the circuit under test so the instrument is in parallel
with the circuit. In DCV measurement, the meter pointer deflects to the right
(normal direction) when the red test lead is connected to the positive side of the
circuit under test and the black test lead to the negative side of the circuit.
Connecting the test leads the other way will reverse the pointer deflection.
(4)Take reading on the appropriate scale.
Current Measurements (DCA)
# DANGER
●
Exercise caution not to apply voltage to the instrument set to a current range.
●Do not turn the function selector switch knob during measurement.
● Make sure to firmly connect the test leads to the circuit under test so that the
connections will not become loose. The test leads must be connected to or
removed from the circuit under test with the circuit powered off.
● When the order of the current under test is unknown, make measurement on
the highest range.
●Keep your fingers and hands behind the Protective fingerguard during
measurement.
(1)
Plug the red test lead into the + terminal and the black test lead into the
-
terminal.
(2)Set the function selector switch to the appropriate DCA position.
(3)Power off the circuit under test.
(4) Connect the test leads to the circuit under test so the instrument is in series with
the circuit. In DCA measurement, the meter pointer deflects to the right (normal
direction) when the red test lead is connected to the positive side of the circuit
under test and the black test lead to the negative side of the circuit. Connecting
the test leads the other way will reverse the pointer deflection.
(5)Power on the circuit under test.
(6)Take reading on the appropriate scale.
Resistance/Continuity Check
# DANGER
●
Exercise caution not to apply voltage to the instrument set to a resistance range
.
●Make sure to power off the circuit under test.
●Keep your fingers and hands behind the Protective fingerguard during
measurement.
— — Resistance Measurement
(1)
Plug the red test lead into the + terminal and the black test lead into the
-
terminal.
(2)Set the function selector switch to the appropriate resistance position.
(3) Short the test lead tips. Turn the Ohm Zero Adjust Knob to set the meter
pointer to the "0" mark on the extreme right of the scale for accurate
measurement. Make this adjustment whenever the function selector switch is
turned to a different resistance position. When the zero adjustment cannot be
made, replace the batteries.
(4)Connect the test leads to the circuit under test.
(5)Take reading using the appropriate multiplying factor.
Note: Note that keeping the test lead tips shorted will exhaust the internal batteries.
— — Continuity Check
(1)
Plug the red test lead into the + terminal and the black test lead into the
-
terminal.
(2)Set the function selector switch to the ' ' position.
(3)Connect the test leads to the circuit under test.
(4)Check if there is a sound of the buzzer. The buzzer beeps below about 100Ω.
Note: The meter does not read on this range.
— — LED Check
(1)
Plug the red test lead into the
+
terminal and the black test lead into the - terminal.
(2)Set the function selector switch to the p1-x10 position.
(3)Connect the test leads to the LED under the light-up test.
(4)When the LED does not light up, connect the test leads the other way.
Note: Connect the red test lead to the anode of the LED and the black test lead
to the cathode. The meter deflection has no meaning on this range.
— Temperature (TEMP.) Measurements —
(1)
Plug the red test lead into the
+
terminal and the black test lead into the - terminal.
(2)Set the function selector switch to the p1-x10 position.
(3)
Short the test lead tips. Turn the Ohm Zero Adjust Knob to set the meter pointer
to the "0" mark on the extreme right of the scale for accurate measurement.
(4)Remove both red and black test reads from the instrument.
(5) Plug the red lead of temperature probe Model 7060 to the + terminal and the
black lead to the
-
terminal.
(6) Touch the object under test with the tip of the temperature probe and take
reading on the TEMP scale.
Note: Can not measure with temperature probe discontinued.
Battery Test (BATT. TEST 1.5V)
This range measures the voltage of a battery, applying a load similar to that used
in common applications (load resistance: 10 ). Ω
# DANGER
●Do not apply voltage above the rated voltage for the battery test.
●Do not turn the function selector switch during a test.
●Keep your fingers and hands behind the Protective fingerguard during
measurement.
(1)
Plug the red test lead into the + terminal and the black test lead into the - terminal.
(2)Set the function selector switch to the 'BATT. TEST 1.5V' position.
(3) Connect the red test leads to the positive side of the battery and the black test
lead to the negative side of the battery.
(4)Take reading on the BATT TEST scale.
Note: The less power a battery has, the lower the reading on this range becomes
compared to the reading on the 3V DC range.
7. Battery & Battery Replacement
# WARNING
● To avoid electric shock hazard, be sure to set the function selector switch to
the OFF position and remove the test leads from the instrument.
●Replacement fuse must have the following rating.
Fast acting type, F500mA/600V, 6.3 32mmφ ×
# CAUTION
●Do not mix new and old batteries.
●Install the batteries observing correct polarity shown inside the instrument.
(1) Remove the test leads from the terminals on the instrument and set the function
selector switch to the OFF position.
(2) Loosen the screw on the bottom of the instrument and remove the bottom case
from the instrument.
(3)Replace the batteries or fuse with new one(s). Use;
Two R6P dry batteries
A fast acting type fuse: F500mA/600V, 6.3 32mmφ ×
(4)Mount the bottom case and tighten the screw.
Note: Use the spare fuse stored inside the instrument. After using the spare fuse,
store a new one as a new spare.
-+
+-
Spare Fuse
Bottom Case
Fuse
DISTRIBUTOR
6-18 92-1430F
Kyoritsu reserves the rights to change specifications
or designs described in this manual without notice
and without obligations.
Cap
Protective fingerguard
Produktspezifikationen
Marke: | Kyoritsu |
Kategorie: | Multimeter |
Modell: | 1110 |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Kyoritsu 1110 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung Multimeter Kyoritsu
27 September 2024
20 September 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
29 August 2024
28 August 2024
9 August 2024
5 August 2024
Bedienungsanleitung Multimeter
- Multimeter Basetech
- Multimeter Gembird
- Multimeter Brennenstuhl
- Multimeter Clas Ohlson
- Multimeter Profile
- Multimeter REV
- Multimeter Voltcraft
- Multimeter Bosch
- Multimeter Laserliner
- Multimeter Velleman
- Multimeter Powerfix
- Multimeter Alecto
- Multimeter EMOS
- Multimeter TFA
- Multimeter FERM
- Multimeter Topcraft
- Multimeter Schneider
- Multimeter Tacklife
- Multimeter Abus
- Multimeter Elro
- Multimeter Stanley
- Multimeter Perel
- Multimeter Wago
- Multimeter Workzone
- Multimeter Somogyi
- Multimeter Amprobe
- Multimeter Metrix
- Multimeter Maxwell
- Multimeter Silverline
- Multimeter Yato
- Multimeter Brandson
- Multimeter Keithley
- Multimeter Hager
- Multimeter GW Instek
- Multimeter Monacor
- Multimeter Brüder Mannesmann
- Multimeter Milwaukee
- Multimeter JUNG
- Multimeter Testo
- Multimeter Elma
- Multimeter IDEAL
- Multimeter Digitus
- Multimeter Kingcraft
- Multimeter Testboy
- Multimeter Projecta
- Multimeter Greenlee
- Multimeter Multimetrix
- Multimeter Extech
- Multimeter BENNING
- Multimeter Metrel
- Multimeter Högert
- Multimeter APPA
- Multimeter Fluke
- Multimeter FLIR
- Multimeter Aldi
- Multimeter Gossen Metrawatt
- Multimeter Hubinont
- Multimeter Joy-it
- Multimeter Kewtech
- Multimeter Mastech
- Multimeter Megger
- Multimeter Micronta
- Multimeter PCE Instruments
- Multimeter Plieger
- Multimeter Sanwa
- Multimeter Skandia
- Multimeter Sonel
- Multimeter Steren
- Multimeter Uni-T
- Multimeter Wiha
- Multimeter Vimar
- Multimeter Owon
- Multimeter PeakTech
- Multimeter Amiko
- Multimeter Rigol
- Multimeter HT Instruments
- Multimeter Beha-Amprobe
- Multimeter Chauvin Arnoux
- Multimeter Aim TTi
- Multimeter Etekcity
- Multimeter Rohde & Schwarz
- Multimeter Strex
- Multimeter MBS
- Multimeter Kurth Electronic
- Multimeter Weidmüller
- Multimeter CEM
- Multimeter Klein Tools
- Multimeter Dasqua
- Multimeter Horex
- Multimeter Enovalab
- Multimeter Qian
- Multimeter Digi-tool
- Multimeter MGL Avionics
- Multimeter Cosinus
- Multimeter Weltron
- Multimeter EEVBlog
- Multimeter Caltek
- Multimeter Elworks
- Multimeter Bearware
- Multimeter Noyafa
Neueste Bedienungsanleitung für -Kategorien-
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
13 Oktober 2024
9 Oktober 2024
5 Oktober 2024
5 Oktober 2024
3 Oktober 2024