How to Prime Pump the Test
Before you connect the to the Unit Under T), the Pump Test (UU
Pump must be lled with uid and primed for use. Follow the
directions below to prepare the for use. Pump
1. Fill the with uid: Pump
a. Remove the Reservoir Filler Cap (item .)
b. Fill the Hydraulic Fluid Reservoir approximately -thirds full two
(66 with distilled water, mineral oil, or hydraulic oil. cc)
Note
If water is used in the , use only DISTILLED water Pump
and drain the water out of the after use. Pump
c. Install the Reservoir Filler Cap (item .) Tighten by hand.
2. the : Prime Pump
a. Adjust the Vernier Fine Adjustment Knob (item ) to the
approximate dle-range.mid
b. Turn the Pressure Bleed Knob (item ) clockwise until it
stops in the full closed position.
c. Squeeze the handles together to pump uid. uid until Pump
uid shows at the . Pump Port
d. Attach the uid hose to the . Pump
e. Squeeze the handles together until uid shows at the end of
the hose.
f. Attach the uid hose to the Unit Under (UUT) and Test
squeeze the handles until pressure increases. Pump
g. Bleed the trapped air from the Pump. Pressurize the Pump to
1,000 psi then slowly turn the Pressure Bleed Knob (item )
counter-clockwise to release pressure.
Notes
• As the pressure it will be hard to squeeze the rises
handles together. Use the Handle Limiter (item ) to
limit the travel of the handles which will make them
easier to squeeze together. toggle the Handle To
Limiter on or o, squeeze the handles together and
push the Handle Limiter (item
) in or out.
• The rate at which the pressure increases is directly
related to the test conguration (hose length and
volume of UUT.)
Replacement Parts
Rebuild Kit, Fluke PN 4097836
Hydraulic Fluid Reservoir Rebuild , Fluke PN 2844341 Kit
Handle Assembly Rebuild , Fluke PN 2844352 Kit
LIMITED WARRANTY AND OF LIABILITYLIMITATION
This Fluke product will be from defects in material and workmanship for free
one year from the date of purchase. This warranty does not cover fuses,
disposable batteries, or damage from accident, neglect, misuse, alteration,
contamination, or conditions of operation or . Resellers are abnormal handling
not authorized any obtain to extend other warranty on Fluke’s behalf. To
service during the warranty period, contact your nearest Fluke authorized
service r cente to obtain return authorization information, then send the product
to that with a description of the problem.Service Center
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES,
SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR
IMPLIED. FLUKE SPECIAL, INDIRECT, IS NOT LIABLE FOR ANY
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not
allow the exclusion or of an implied warranty or of incidental or limitation
consequential damag not apply to you.es, this limitation of liability may
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11/99
Fluke 700HTP- -2
Hydraulic Test Pump
Instruction Sheet
Introduction
The Fluke 700HTP-2 Hydraulic (the or ProductTest Pump Pump ) is a
portable source of high pressure.
Safety
W Warning
To prevent personal injury se the Product only as , u
specified, or the protection supplied by the Product can be
compromised.
Pump Sp ecifications
• Maximum pressure: 10,000 psi (690 bar)
• raulic media: distilled water, mineral based hydraulic oil, up Hyd or
to hydraulic oil 100 (30 W).
• Wetted materials: 303 stainless steel, aluminum, polyurethane,
PTFE, nitrile, and nylon.
• Weight: 1.29 lb (0.58 kg)
• Dimensions: length 9.5 41 mm), width 6.1 in (155 mm), depth in (2
2.6 in (66 mm)
Box Contents
• Model 700HTP-2 Hydraulic with a 1/4- NPT tee Test Pump inch
installed.
• 1/4- NPT male to 1/4- female adapters. Two inch inch BSP
• One 1/4- NPT male to 1/4- NPT male tting. inch inch
• One 1/4- NPT male to 1/8- NPT male tting. inch inch
• Instruction Sheet
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
• Technical Support USA: 1-800-44-FLUKE
(1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE
(1-888-993-5853)
• Canada: 1-800-36-FLUKE
(1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-3-6714-3114
• Singapore: +65-738-5655
• Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit . http://register.uke.com
To view, print, or download the latest manual supplement, visit
http://us.uke.com/usen/support/manuals.
Features
gsp001.eps
No. Item Function
Master
Instrument Port
Port for Fl uke 700 Series Pressure
Module, 1/4-inch NPT female.
Test Pressure
Port
Port for pressure instrument under
test, 1/4-inch NPT female.
Pressure Relief
Valve Port
Port for an optional Fluke 700PRV-1
Pressure Relief Valve (plugged).
Vernier Fine
Adjustment Knob
Lets you adjust the applied pressure
in small increments.
Hydraulic Fluid
Reservoir
Holds 100 cc of mineral based
hydraulic oil, distilled water, or
hydraulic oil.
1/4-inch NPT Tee
Fitting For master and test ports.
Handle Limiter Limits the travel of the handles.
Reservoir Filler
Cap
Cap which seals the hydraulic uid
reservoir. Install before you pump.
Pressure Bleed
Knob
Lets you release pressure in small
increments.
Reservoir Nut Remove nut to clean reservoir.
gsp002.eps
Figure 1. Test Configuration
How to Operate the Test Pump
W Warning
To prevent a violent release of pressure, open the
Pressure ) very slowlyBleed Knob (item before you
disconnect pressure lines. Do not connect the Pump to
an external pressure source.
W Caution
To prevent damage hen odule or w a Pressure M Pressure
I nstrument is connected that could be damaged by
pressures of 690 bar (10,000 psi), install and correctly
set a relief valve in the system or an optional Fluke
700PRV-1 Pressure Relief Valve on the . Pump
1. Connect a 700 Series Pressure Module (the Pressure Module)
to the Master Instrument (item .) To connect to a 700P Port
High Pressure Module (700P29, P30, or P31) install a male
1/4-inch NPT Adapter. Use Teon
® tape or equivalent on the
NPT thread connections to prevent leaks.
2. Fill and the . See “How to the Prime Pump Prime Test Pump”.
3. Connect the Pressure Module to the calibrator as shown in
Figure 1.
4. Connect the input of the Pressure Unit Under Test U to the (U T)
Test Port Test Pressure (item ) on the . Pump
5. Loosen the Pressure Bleed Knob (item ) to release pressure
from the . Pump
6. Use the Calibrator to z the Pressure Module. Refer to the ero
Users Manual for instructions.
7. Turn the Adjustment Knob (item ) to the Vernier Fine
middle-range.
8. Turn the Pressure Bleed Knob (item ) clockwise until it stops
in the full closed position.
9. Squeeze the handles together to pump increase fluid which will
pressure. See the Notes in step 2 of the “How to the Prime Test
Pump” procedure.
W Caution
To prevent damage, do not let the pressure increase to
more than the pressure rating of the equipment under
test. Pressure can increase quickly if the pressurized
volume is small.
10. incremental pressure adjustments, use the Vernier Fine For
Adjustment Knob.
11. Once the test is done, slowly turn the Pressure Bleed Knob
(item ) counter-clockwise to release pressure.
Produktspezifikationen
Marke: | Fluke |
Kategorie: | pomp |
Modell: | 700HTP-2 |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Fluke 700HTP-2 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung pomp Fluke
21 September 2024
23 August 2024
23 Juli 2024
22 Juli 2024
22 Juli 2024
20 Juli 2024
Bedienungsanleitung pomp
- pomp Crivit
- pomp SilverCrest
- pomp SKS
- pomp Makita
- pomp Vonroc
- pomp Siemens
- pomp Powerfix
- pomp Black And Decker
- pomp Ryobi
- pomp Coleman
- pomp Einhell
- pomp Worx
- pomp Bartscher
- pomp Ultimate Speed
- pomp Eheim
- pomp Eurom
- pomp Hyundai
- pomp Fuxtec
- pomp Hazet
- pomp Metabo
- pomp Kärcher
- pomp Güde
- pomp Scheppach
- pomp Fieldmann
- pomp Wetelux
- pomp Westfalia
- pomp Vaillant
- pomp Husqvarna
- pomp Enermax
- pomp MSW
- pomp Germania
- pomp Fluval
- pomp Vetus
- pomp Jabsco
- pomp Sanibroyeur
- pomp Topeak
- pomp Hayward
- pomp Gre
- pomp Milwaukee
- pomp Toolcraft
- pomp Osram
- pomp Bestway
- pomp AL-KO
- pomp Gardena
- pomp Melinera
- pomp Powerplus
- pomp Blackburn
- pomp Lezyne
- pomp Alpha Tools
- pomp Sicce
- pomp Bulex
- pomp Danfoss
- pomp Comet
- pomp Hozelock
- pomp Ozito
- pomp Blumfeldt
- pomp Nibe
- pomp Sun Joe
- pomp Nefit
- pomp Oase
- pomp POLARIS
- pomp Velda
- pomp Vivaria
- pomp Waldbeck
- pomp WHALE
- pomp AstralPool
- pomp Pedrollo
- pomp Esotec
- pomp T.I.P.
- pomp Panduit
- pomp Heissner
- pomp DAB
- pomp Abac
- pomp Baracuda
- pomp Ubbink
- pomp JANDY
- pomp Liberty Pumps
- pomp Monzana
- pomp Lifan
- pomp Elpumps
- pomp For_Q
- pomp Simpson
- pomp Pontec
- pomp BluGarda
- pomp Grundfos
- pomp Zoeller
- pomp B-Air
- pomp Anova
- pomp Perlick
- pomp Fiap
- pomp Little Giant
- pomp Ebara
- pomp Basement Watchdog
- pomp Barwig
- pomp Wayne
- pomp Ribimex
- pomp Maruyama
- pomp ESPA
- pomp Hollex
- pomp Sauermann
- pomp Eco-Flo
- pomp Nowax
- pomp EvoHeat
- pomp Zéfal
- pomp OSIP
- pomp Franklin Electric
Neueste Bedienungsanleitung für -Kategorien-
16 Oktober 2024
16 Oktober 2024
16 Oktober 2024
15 Oktober 2024
15 Oktober 2024
15 Oktober 2024
13 Oktober 2024
7 Oktober 2024
2 Oktober 2024
1 Oktober 2024