Introduction
qsg Q10562 Copyright © 2020 Aluratek, Inc. All Rights Reserved.
aluratek.com
model ADPM01F
User Manual
Home Health Digital Pulse Oximeter
If you need further assistance, please contact the Aluratek support
department for troubleshooting prior to returning this device.
E-mail: support@aluratek.com
Web: www.aluratek.com/helpdesk
Local (Irvine, CA): 714-586-8730
Toll free: 1-866-580-1978
Thank you very much for your purchase! This manual is written and
complied in accordance with the council directive MDD93/42/EEC for
harmonized standards. The Manual is written for the Pulse Oximeter you
purchased. In case of modifications and software upgrades, the information
contained in this document is subject to change without notice. The Manual
describes, in accordance with the Pulse Oximeter’s features and
requirements, main structure, functions, specifications, correct methods for
transportation, installation, usage, operation, repair, maintenance and
storage, etc. as well as the safety procedures to protect both the user and
equipment. Refer to the respective chapters for details.
Please read the Manual very carefully before using this equipment. These
instructions describe the operating procedures to be followed strictly, failure
to follow these instructions can cause measuring abnormality, equipment
damage and personal injury. The manufacturer is NOT responsible for the
safety, reliability and performance issues and any monitoring abnormality,
personal injury and equipment damage due to user’s negligence of the
operation instructions.
WARNING:
• It is recommended that the Pulse Oximeter should not be applied to
the same finger for over 2 hours.
• The infrared light used in this devicecan be harmful to the eyes. Do
not stare at the at the light. (The LED display is not harmful)
• This device is not intended for treatment of any kind.
• For best results do not use with dark nail polish as it may interfere
with accurate readings.
• This manual is published by Aluratek, Inc. All rights reserved.
Aluratek warrants this product against defect in material or
workmanship for from the date of purchase.1 YEAR
For more information, please visit:
https://aluratek.com/warranty-return-policy
You can register your product online at:
https://aluratek.com/product-registration
1) Insert the two batteries ensuring that the + (plus) and – (minus) terminals
are aligned correctly, and then replace the cover.
2) Open the clip as shown in Figure 4.
3) Insert the finger into the rubber cushions of the clip (make sure the finger
is in the right position)
4) Press the power button once on front panel. Please keep the finger/hand
steady during the measurement.
5) Result will be shown on the display.
The Power button has two main functions. When the device is powered off,
press the button to power it on. When the device is powered on, quick press
the button to change the display orientation of the LED display.
Fingernail should be facing up toward the LED Display
1. Safety
1.1 Instructions for Safe Operation
Check the digital pulse oximeter periodically to make sure there is no
visible damage to the product. It is recommended that the device should
be inspected once a week. Any necessary maintenance must be performed
by qualified service engineers ONLY. Do not attempt to fix or repair the
oximeter in any way.
1.2 Warnings
Explosive hazard – DO NOT use the oximeter in an environment with
concentrated flammable gases.
- DO NOT use the oximeter during an MRI or CT scan.
- DO NOT use if you are allergic to rubber.
- Disposal – Make certain your follow your local laws and regulations when
disposing of this device (including battery, plastic bags, foam and paper
boxes).
1.3 Suggestions
Keep the oximeter away from dust, excessive vibration, corrosive
substances, explosive materials, high temperature and moisture. If the
oximeter gets wet, please stop operating it until it has completely dried.
- When it is carried form extreme temperatures such as very cold
environment to warm/hot environments please do not use it immediately
allowing the product to acclimate to the new temperature for best results.
- Be careful not to scratch the surface of the LED or sensor with sharp
materials.
- High temperature or high pressure steam disinfection of the oximeter is
not permitted. Refer to User Manual in the relative chapter for instructions
on the cleaning and disinfection of the device.
- Do not have the oximeter completely submerged in any liquid. When it
needs cleaning, please wipe its surface with medical alcohol and a soft cloth.
- Do not spray any liquid directly on the device.
- When cleaning the device with water, the temperature should be lower
than 60°C (140°F)
2. Overview
The pulse oxygen saturation is the percentage of HbO2 in the total Hb in
the blood, so-called O2 concentration in the blood. It is an important
bio-parameter for respiration. For the purpose of measuring the SpOz more
easily and accurately, our company developed the Pulse Oximeter. At the
same time, the device can measure the pulse rate simultaneously. The
Pulse Oximeter features in small volume, low power consumption,
convenient operation and being portable. It is only necessary for patients to
put one of their fingers into a fingertip photoelectric sensor for diagnosis,
and a display screen will directly show measured value of Hemoglobin
Saturation and Pulse rate.
2.1 Classification:
Class II b, (MDD93/42/EEC IX Rule 10)
3. Principle and Caution
2.2 Features
Operation of the product is simple and convenient.
• The product is small and light weight (total weight is about 50g/0.11lbs
including batteries) and convenient for carrying.
• Power consumption of the product is extremely low. Two AAA batteries
can power the device continuously for an estimated 20 hours.
• The product will automatically be powered off when not in use after 5
seconds to conserve the battery life.
2.3 Major Applications and Scope of Application
The Pulse Oximeter can be used to measure human Hemoglobin Saturation
and pulse rate through the finger and indicate the pulse intensity.
The product is not suitable for use in continuous supervision of
patients. If suffering from toxicosis which caused by carbon
monoxidethe device is not recommended for use
2.4 Environment Requirements
Storage Environment
a) Temperature: - 40°C + 60°C
b) Relative humidity: <95%
c) Atmospheric pressure: 500hPa – 1060hPa
Operating Environment
a) Temperature: 10°C- 40°C/50°F-105°F
b) Relative Humidity: <75% (for best results)
c) Atmospheric pressure: 700hPa – 1060hPa
3.1 Principle of Measurement
Principle of the Oximeter is as follows: An experience formula of data process
is established taking use of Lambert Beer Law according to Spectrum
Absorption Characteristics of Reductive Hemoglobin (Hb) and Oxyhemoglobin
(HbO2) in glow & near-infrared zones. Operation principle of the instrument
is: Photoelectric Oxyhemoglobin Inspection technology is adopted in
accordance with Capacity Pulse Scanning & Recording Technology, so that
two beams of different wavelength of lights can be focused onto human nail
tip through perspective clamp finger-type sensor. Then measured signal can
be obtained by a photosensitive element, information acquired through which
will be shown on screen through treatment in electronic circuits and
microprocessor.
Figure 1 Operating Principle
Photo Detector
Emitters
3.2 Caution
1. The finger should be placed properly (see the attached illustration of this
manual, Figure 4), or else it may cause an inaccurate measurement.
2. The SpO2 sensor and photoelectric receiving tube should be arranged in a
way with the subject’s arteriole in a position there between.
3. The SpO2 sensor should not be used at a location or limb tied with arterial
canal or blood pressure cuff or receiving intravenous injection.
4. Make sure the optical path is free from any optical obstacles like
rubberized fabric.
5. Excessive ambient light may affect the measuring result. It includes
fluorescent lamp, dual ruby light, infrared heater, direct sunlight and etc.
6. Strenuous action of the subject or extreme electrosurgical interference
may also affect the accuracy.
7. Do not use enamel or other dark nail polishes or paint for more accurate results.
4. Technical Specifications
1) Display Format: OLED display
SpO2 Measuring range: 0% - 100%
Pulse Rate measuring range: 30bpm – 250bpm;
Pulse Wave display: Columniation display and the waveform display.
2) Power Requirements: 2x1.5V AAA alkaline battery (not included)
3) Power consumption: Less than 30mA
4) Resolution: 1% for SpO
2 and 1bpm for Pulse rate.
5) Measurement Accuracy: +-2% in stage of 70%-100% SpO
2, and the
meaningless when stage being smaller than 70% +- 2bpm or +-2%
(select larger) for Pulse rate.
6) Measurement Performance: SpO
2 and pulse rate can be shown correctly
when pulse-filling ratio is 0.4%. SpO
2 error is +-4%, pulse rate error is
+-2bpm or +-2%.
7) Resistance to surrounding light: The deviation between the value
measured in the condition of man-made light or indoor natural light and
that of darkroom is less then +-1%.
8) It is equipped with a function on/offpower switch. The oximeter will power
off in case no finger is the oximeter within 5 seconds to preserve battery life.
9) Optical sensor: Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
5. Installation
5.1 View of the Front Panel
5.2 Battery
Step 1. Refer to figure 3. And insert the two AAA size batteries properly in the
right direction.
Step 2. Replace the cover.
Please take care when you insert the batteries for the improper
insertion may damage the device.
Power buttonPulse bar
graph
Pulse waveform
Battery indicator
Pulse rate
SpO2
Figure 2 Front View
6. Operating Guide
7. Repairing and Maintenance
• Please change the batteries when low-voltage is displayed on the screen or
the device no longer powers on.
• Please clean the surface of the device before using. Wipe the device with
medical alcohol first, and then let it dry in air or clean it by dry clean fabric.
• Using the medical alcohol to disinfect the product after use, prevent from
cross infection or next time use.
• Please take out the batteries if the oximeter is not going to be used for a
long period of time.
9. Technical Support and Warranty
8. Radio and TV Interference Statement
Quick Reference Guide
Please thoroughly read the manual prior to operations of
the Pulse Oximeter:
1. Insert 2 AAA batteries (ncluded)
2. Power on
3. Insert finger
Note: If you receive afinger out” message please reinsert
or try a different finger or thumb.
Tips: If your hands/fingers are cold or if you are wearing
dark nail polish you may not be able to get an accurate
reading. Please warm your hands/fingers and/or remove
the nail polish and try again. If the sensors are dirty you
may not be able to get a reading. Gently clean the sensor
with rubbing alcohol and try again.
- For fingers that have poor blood circulation due to being cold or for really
thin fingers best results may be yielded from the thumb.
- Do not use the device on infant or neonatal patients.
- The product is best suited for children over four years old and adults
whose weights do not exceed 300 lbs. - The device may not work for all
patients. If you are unable to achieve stable readings, discontinue use.
- During proper working conditions results should begin to be displayed
within 5 seconds. It is common for pulse and Oxygen levels to vary
during use.
- The device does not have a low-voltage alarm function, it only shows the
low-voltage, please change the batteries if the device no longer powers on.
- This does not have an alarm function. Do not use the device in situations
where alarms are required.
- It is highly recommended that the batteries be removed if the device is
going to be stored for extended periods of time of more than one month,
or else batteries may leak.
- A flexible circuit connects the two parts of the device. Do not twist, pull or
try to disassemble.
WARNING!!! This equipment has been tested and found to comply with
the limits of a Class B digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the instruction, may
cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on,
the user is encouraged to try to correct the interference by whatever
measures may be required to correct the interference.
Figure 3 Battery Installation
Figure 4 Finger position
(palm down measurement is more accurate)
!
!
!
!

Produktspezifikationen

Marke: Aluratek
Kategorie: Messung
Modell: ADPM01F
Breite: 60 mm
Tiefe: 36 mm
Gewicht: 57 g
Produktfarbe: Weiß
Akku-/Batterietyp: AAA
Höhe: 31 mm
Energiequelle: Akku
Schnellstartübersicht: Ja
Display: LED
Anzahl unterstützter Akkus/Batterien: 2
Genauigkeit: 2 BPM
Messbereich Sauerstoffsättigung: 0 - 100 %
Messbereich Pulsfrequenz: 30 - 250 BPM