Aluratek ADIT01F Bedienungsanleitung


Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Aluratek ADIT01F (2 Seiten) in der Kategorie Nicht kategorisiert. Dieser Bedienungsanleitung war für 22 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet

Seite 1/2
The icon for human
body mode”
Non-Contact Forehead Thermometer
Lot label
Product label
Probe
Handle
Measurement
button
Setting buttons
Battery lid
LCD display
136
76.5 38
Resolution 0.1℃/0.1℉
Working condition
15℃~40℃(59.0℉~104℉)
RH≤95%Non-condensing
“ERR”displays when it’s not used under
working condition
Storage condition -25℃~55℃ (-13℉~131℉)
RH≤95%Non-condensing
6.Illustration
Temperature value of previous
measurement
Memory
Temperature value
Reading display
Fahrenheit reading
Celsius reading
Battery power is sufficient.
and it functions properly.
When it is
visible
Reading scale
Object mode
Human Body mode
The battery is exhausted
and thermometer can not
function properly. Please
replace battery immediately
When it
flashes
The battery is in low level,
but the thermometer is still
functional properly.Please
replace battery asap
When it is
visible
Measurement mode
Battery level
DetailsIconFunction definition
7.Display & icons
4.Features
Any object can generate certain proportion of infrared radiant energy
as per its own temperature. The radiant energy and its wavelength
distribution are subjected to its surface temperature. Based on this
principle, this thermometer is designed to detect infrared radiation
at 5~14um by highly precise infrared sensor. By adopting this high
quality sensor plus special calculation and calibration, this thermometer
is able to take accurate body temperature.
2.Working principle
3.Safety precautions
This is a Non-Contact Forehead Thermometer applicable to forehead
measurement. The thermometer measures body temperature by
collecting heat radiation emitting from forehead. New probe
structure is adopted in this thermometer. It's simply operational,
hygeian, reliable and highly accurate. Users can get precise reading
within one second by one touch. This thermometer is cost-effective
and is widely used in schools, customs, hospitals and for family use.
Intended Use & Indication for use: The Non-Contact Forehead
Thermometer is an infrared thermometer intended for the intermittent
measurement of human body temperature in people of all ages.
This thermometer is classified as a Class IIa(for CE)/class II(for FDA)
Medical Device, sorted as internally- powered equipment and type
BF application device. It's prohibited to use this thermometer in
flammable anesthetic gas or gas mixture of air and oxygen or
nitrous oxide. This is a continuous operation equipment.
1.Introduction & classification
Notice:
1.Temperature under human body mode is obtained from dynamic compensation
of environmental temp and forehead surface temp.
2. Object temperature mode is to test surface temperature of an object. The
temperature get from forehead under this mode is merely temperature of forehead
surface but not body temperature.
(fig.10.1) (fig.10.2) (fig.10.4)
5-8cm
(fig.10.3)(fig.10.5)
Press measurement button to turn on thermometer and it displays boot
screen (fig.10.1). After POST and two beeps, it will display value of last
reading and be ready for measurement (fig.10.2).
Make sure the thermometer is under body mode.
Keep distance at 5cm to 8cm from thermometer probe to the middle of
forehead (fig.10.3). Press measurement button and then it gives a beep to
indicate measurement is finished and value will be displayed (fig.10.4). If
measurement value exceeds alarm value(Defaulted value is 38℃),it gives
beep. beep. beep as a indication.
After measurement, if the thermometer is idle for 30 seconds, it will display
OFF (fig.10.5) and gives a beep and shut off automatically.
10.1 Body temperature
10.Measurement
When thermometer is on, it displays current Reading scale. Press theC/F”button
to select reading scale.
Reading scale setting:
( fig.9.1) ( fig.9.2)
When thermometer is on, it displays current measurement
mode ( fig.9.1). Press the “Mode” button to change
measurement mode( fig.9.2).
Measurement mode setting:
User can change reading scale between Celsius and Fahrenheit, and
change measurement mode between human body mode and object
mode .
9.Setting
To switch unit of temperature readingC/F
To track last 10 readings MEM
To switch measurement mode between
human body and object
Mode
DescriptionButtons
8.Function definition of buttons
Mode MEM C/F
Power supply d.c. 3V 2*AAA Batteries
Power consumption When off≤10uW
When measurement≤30mW
Shelf life 5 Year
Memory Storage of last 10 readings
Display 3 Colors Backlight LCD (red, green, orange)
Reading scale Celsius or Fahrenheit
Automatic shut off In 30 seconds
Dimensions 136mm×76.5mm×38mm
Net weight 75g
Operating Instructions Instruction Manual
1. Take out the thermometer and open the battery
lid to install 2 *AAA batteries
2. Keep distance of 2-3 inches from thermometer
probe to the middle of forehead, please do not
contact forehead directly
3. Select measurement mode and click the
measurement button to start taking temperature
4. Reading displays on the screen after successful
measurement
2-3 IN.
Measurement button
The icon for object
mode”
Distinguish the positive and negative
poles of the battery
Notice:
1.
The value under this mode is object surface temperature instead of core temperature.
2.The default value of edinfrared emissivity is 0.95. The reading will deviate from the
real temperature because of different emissivity. For example, the reading on stainless
steel is obviously lower than real temperature. BE CAUTIOUS FOR SCALDING.
(fig.10.8)(fig.10.7)(fig.10.6)
Press measurement button to turn on thermometer (fig.10.6).
Make sure the thermometer is under object mode.
Keep vertical distance at 5cm to 8cm from object to measurement probe.
Press measurement button and then it gives a“beepto indicate measurement
is finished and value will be displayed (fig.10.7).
After measurement, if the thermometer is idle for 30 seconds, it will display
“OFF”
(fig.10.8) and gives abeep and shut off automatically.
10.2 Object temperature
Measuring accuracy(at
laboratory conditions)
for forehead temperature:
±0.2℃/0.4℉ during 34.0℃-42.0℃;
±0.3℃/0.5℉ during 42.1℃-43.0℃.
ADIT01F
5cm~8cm (2in~3in)
Measurement
distance
Human body
Measurement
34.0~43.0 (93.2~109.4℉)
<34.0℃/93.2℉,LO”displays
>43.0℃/109.4℉,“HI”displays
Object
0℃~93℃ (32.0℉~199.4℉)
<0℃/32℉,“LO”displays
>93℃/199.4℉, “HIdisplays
5.Technical parameters
Warning:
Using this thermometer is not intended as a substitute for
consultation with your physician.It is dangerous for users to
perform a self-evaluation and self-treatment based on the
measuring result.please follow doctor’s instruction.
Keep the thermometer out of reach of children ,please consult
the doctor at once in case of accidental swallow of battery or
other component.
Don’t throw the battery into fire
Notice:
The device is precision instrument, don’t drop, tramp or
impose any vibration or impact on the thermometer.
Do not touch the lens of the probe with your fingers and
disassemble the device by yourself.
Please make sure your forehead is clean before measuring
forehead temperature.
Please stay still indoors about 30 minutes after exercise, eating
or bathing before measuring.
Please place the thermometer indoors for about 30 minutes if
ambient temperature varies a lot before using.
Please collect the record of Individual temperature under the
good condition of body in usual days as a reference for checking
fever or not.
Do not measure the sites of scarred tissue or tissue
compromised by skin disorders because they will affect the
accuracy of measurement.
Do not measure if patient is treated with certain drug therapies
because body temperature may rise in the drug within the effort
time limit.
Do not immerse the device into water or any other liquid, do not
expose to the sun.
Do not use a mobile or cordless hone near the thermom
measuring. Do not use the thermometer near a mobile or
cordless hone.
Please don’t measure body temperature in strong electromag-
netic interference environment (such as microwave, high
frequency equipment operation environment) to ensure the
accuracy of measurement data.
This thermometer only a personal device, please do not share
with others.
Please store the thermometer according to the technical
specifications.
Keep the sensor and probe cavity clean before use and after use.
Recommendations:
The materials (ABS) of contact with patient has passed the ISO
10993-5 and ISO 10993-10 standard test, no toxicity, allergy and
irritation reaction. They are compliant with the MDD
requirements based on the current science and technology, and
other potential allergic reactions are unknown.
The patient can measure, read data and replace battery under
normal circumstances and maintain the device and its
accessories according to the user manual.
The PATIENT is an intended OPERATOR.
Don’t use this thermometer for other purposes.
It is forbidden to leave the product exposed to any chemical
solvent, direct sunshine or high temperature in case of
damaging the product or the battery.
Do not measure while talking on the phone.
Please report to MANUFACTURER if any unexpected operation
or events occur.
Measurement mode for human body and object available,one
button to switch measurement mode;
Beeper for high temperature;
℃/℉ reading available;
Automatic shut off;
Storage of last 10 readings;
3 Colors Backlight LCD for easy reading.
Please read this manual instruction thoroughly before use.
Version: V2.0 Date: 2020-06-03
Compliant to IEC 60601-1, IEC60601-1-11, IEC 60601-1-2,
IEC 62304, ISO80601-2-56, ASTM E1965-98
Manual instruction

and gives “beep. beep.beep.beep”,with the color of backlight turning red.

and gives beep. beep.beep.beep”,with the color of backlight turning red.󶅂
Object mode:

and gives“beep. beep.beep.beep”, with the color of backlight turning red.
Body mode:

and gives“beep. beep.beep.beep”, with the color of backlight turning red.
10.3 Exceeding measurement range
Notice:

it displays 
Under this condition, it’s not allowed to use this thermometer or accuracy is not assured.
(
fig.10.13
)(

) (
fig.10.10
) (
fig.10.11
) (
fig.10.12
)
The measurement results are probably fluctuating due to improper measurement ways.
Please practice adequate measurements in order to improve your skill.
The measurement results can not supersede a doctor's diagnosis.
Special maintenance is unnecessary for this thermometer. Please contact distributor or
manufacturer in case of malfunction.
13. Trouble shooting
Description Solutions
LCD displayLO
orHI
1.Breeze, water, sweating, cosmetic on forehead may affect
measurement.
2.While if the testing environmental temp changes so
enormously or if the thermometer is used directly from
high -temp object to very low-temp one, the measurement
difference will happen.The thermometer should be kept in
a relatively stable environment for 10 minutes to get heat
balance before starting a new measurement.
3.Ensure measurement distance is 5cm to 8cm.
No response when
pressing measurement
button 1.Take out and reassemble battery.
No display or
improper display
Shut off right after
switching on
1.Take out battery and install battery again.
1.Check battery level or take out the battery and install the
battery again.
EMC Declaration
15. Technical Support and Warranty
12. Maintenance & tips
Make sure the sensor and probe cavity is clean otherwise it will affect accuracy.
Cleaning method for probe:
1.Use the cotton stick or soft cloth with water or alcohol to wipe the casing.
2.Use the cotton stick or soft cloth with alcohol to wipe the sensor surface or probe
cavity gently. Don't use thermometer before alcohol is vaporized.
Read this manual book thoroughly before use. Make sure battery is well installed.
It is not allowed to put the thermometer in any liquid or expose to strong sunlight
or extremely low temperature.
Strong crash or hit to the product will cause its damage.
Do not dismantle this thermometer by yourself.
Open the battery lid and take out exhausted battery.
Put into 2 AAA batteries and close up battery lid. After new battery is installed,
the color of backlight turns in the order of green , orange and red , each of them
flashing one time respectively, with abeep heard. If no beep is heard, please check

Notice:
1.Take out battery in case the thermometer is not used for a long time. Don't put the
battery to fire.
2.Dispose battery according to local regulations.
11. Battery replacement
Keep the thermometer from children's reach.
Do not use the thermometer under circumstance of strong electromagnetic interfere.
(
fig.11.1
)
Standard list
Medical device software - Software life-cycle processes
ISO 80601-2-56
Medical electrical equipment part 2-56: particular
requirements for basic safety and essential performance
of clinical thermometer for body temperature measurement
Disposal
NOTES:
Please act according to the native law to proceed to handle the battery and wastes.
Take out the battery if you are not going to use the unit for a long time.
IEC 62366 Medical devices Application of usability engineering to
medical devices
Biological evaluation of medical devices - Part 1:
Evaluation and testing within a risk management process
IEC 15223-1

Symbols for use in the labeling of medical devices
IEC 60601-1-11
Medical electrical equipment Part 1-11: General requirements
for basic safety and essential performance Collateral standard:
Requirements for medical electrical equipment and medical
electrical systems used in home healthcare environment
IEC 60601-1-2
Medical electrical equipment -- Part 1-2: General requirements
for basic safety and essential performance - Collateral standard:
Electromagnetic compatibility - Requirements and tests
IEC 60601-1 Medical electrical equipment Part 1: General requirements
for basic safety and essential performance
Information supplied by the manufacturer with medical devices
Standard Specification for Infrared Thermometer for
Intermittent Determination of Patient Temperature
IEC 60601-1-6
Medical electrical equipment Part1-6: General requirements
for basic safety and essential performance – Collateral
standard: Usability
Dispose of the device in accordance with the regulation applicable at the place
of operation. Dispose of at public collection point in the EU countries
 WEEE Directive.
If you have any queries, please refer to the local authorities responsible for
waste disposal.
Dispose of empty battery at your retail store or at appropriate collection sites
according to national or local regulations to protect environment.

Read instruction manual before use
Type BF applied part
Batch
Serial number

LOT
SN
Manufacturer information



IP code of the device: this devices grade of against ingress of solid



IP22

parliament and of the council of 8 June 2011
RoHS
COMPL IANT
Follow operating instructions
Guidance and manufacturer´s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage fluctuations
flicker emissions
IEC 61000-3-3
Class B
N/A
N/A
Emissions test
Group 1
Electromagnetic environment - guidance
The Infrared Thermometer uses RF energy
only for its internal function. There for, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The Infrared Thermometer suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
The infrared thermometer is intended for use in the electromagnetic environment
specified below.
The customer or the user of infrared thermometer should assure that it is used in
such an environment..
The infrared thermometer is intended for use in the electromagnetic environment
specified below.
The customer or the user of infrared thermometer should assure that it is used in
such an environment.
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
Guidance and manufacturer´s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
NOTE UT is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturer´s declaration – electromagnetic immunity
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV,
± 15 kV air
Floors should be wood,
concrete or
ceramic tile. If floors are
covered with
synthetic material, the
relative humidity
should be at least 30 %.
Compliance level
IEC 60601
test level
Electromagnetic
environment-guidance
Electrostatic
transient /
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC
61000-4-11
± 2 kV for
power
supply lines
± 1 kV for
input/output
lines
± 1 kV
differential
mode
± 2 kV
common mode
0 % UT; 0,5
cycle g) At 0°,
45°, 90°, 135°,
180°, 225°,
270° and 315°
0 % UT; 1
cycle and 70 %
UT; 25/30
cycles Single
phase: at 0°
0 % UT;
250/300 cycle
N/A
N/A
30 A/m 30 A/m
N/A
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the Infrared
thermometer requires
continued operation
during power mains
interruptions, it is
recommended that the
Infrared thermometer be
powered from an
uninterruptible power
supply or a battery.
The infrared thermometer is intended for use in the electromagnetic environment
specified below.
The customer or the user of infrared thermometer should assure that it is used in
such an environment.
For transmitters rated at a maximum output power not listed above the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
Rated
maximum
output of
transmitter
W
150 kHz to 80 MHz
outside ISM and
amateur radio bands
150 kHz to 80 MHz
in ISM and amateur
radio bands
180 MHz to 800
MHz
800 MHz to 2.7 GHz
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM -for
EQUIPMENT and SYSTEMS
Separation distance according to frequency of transmitter
m
Recommended separation distances between
portable and mobile RF communications equipment and the Infrared thermometer
The Infrared thermometer is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
Infrared thermometer can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Infrared thermometer as recommended below, according to
the maximum output power of the communications equipment
b. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the Infrared thermometer is used exceeds the applicable RF
compliance level above, the Infrared thermometer should be observed to verify
normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the Infrared thermometer.
c. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3V/m.
d=[ ] p
V1
3.5 d=[ ] p
V
2
12
0.01
1
0.1
10
100
0.12
1.2
0.38
3.8
12
0.20
2.00
0.63
6.32
20.00
0.035
0.35
0.11
1.10
35
0.07
0.70
0.22
2.21
70
NOTE 1 At 80 MHz and 800 MHz, the higher frequency ra nge applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is
affected by absorption and reflection from structures, objects and people.
a. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz
are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283
MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz
and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7
MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17
MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and
50,0 MHz to 54,0 MHz.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
6 V in ISM and
amateur radio
bands between
0,15 MHz and
80 MHz
10 V/m
80 MHz to 2.7
GHz
385MHz-5785M
Hz Test
specifications for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communication
equipment
(Refer to table 9
of IEC
60601-1-2:2014)
N/A
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY to
RF wireless
communication
equipment (Refer to
table 9 of IEC
60601-1-2:2014)
80 MHz to 800 MHz
800 MHz to 2.7 GHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Infrared thermometer, including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
where p is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).b
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Recommended separation distance
Immunity test Compliance level
IEC 60601
test level
Electromagnetic
environment-guidance
d=[ ] p
V
1
3.5
d=[ ] p
V2
12
d=[ ] p
E1
3.5
d=[ ] p
E1
7
Normalized symbols
1) Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
2) Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this the Infrared Thermometer could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
3) Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the Infrared Thermometer, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
If you need further assistance, please contact the Aluratek support department
for troubleshooting prior to returning this device.
E-mail: support@aluratek.com
Web: www.aluratek.com/helpdesk
Local (Irvine, CA): 714-586-8730
Toll free: 1-866-580-1978
Aluratek warrants this product against defect in material or workmanship for
1 YEAR from the date of purchase.
For more information, please visit: https://aluratek.com/warranty-return-policy
You can register your product online at: https://aluratek.com/product-registration


Produktspezifikationen

Marke: Aluratek
Kategorie: Nicht kategorisiert
Modell: ADIT01F

Brauchst du Hilfe?

Wenn Sie Hilfe mit Aluratek ADIT01F benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten




Bedienungsanleitung Nicht kategorisiert Aluratek

Bedienungsanleitung Nicht kategorisiert

Neueste Bedienungsanleitung für -Kategorien-