Tripp Lite HCINT350SNR Bedienungsanleitung

Tripp Lite Batterie HCINT350SNR

Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Tripp Lite HCINT350SNR (8 Seiten) in der Kategorie Batterie. Dieser Bedienungsanleitung war für 37 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet

Seite 1/8
23-08-312-932965 REV E
HC350S, HCINT350S Series Power Module Application Guidelines
Series AGSM5396, AGTR0829
SAVE THESE INSTRUCTIONS – This manual contains IMPORTANT SAFETY
INSTRUCTIONS for that should be followed during installation and maintenance of the
UPS and batteries.
The HC350S and HCINT350S series of hospital cart power modules are designed to
provide AC voltage output power from either utility provided AC input voltage or from
inverting 12VDC power from a sealed lead acid battery. To ensure proper operation the
following guidelines should be observed:
1. The power module can support AC loads up to a maximum of 400VA or 300W.
2. The power module charger is intended to be used with a 12V sealed lead acid
absorbed glass mat (AGM) battery with a 33Ah to 66Ah rating. Batteries rated at
less than 33AH are not recommended since the charger current of 12A would
reduce the battery cycle life of smaller batteries.
3. The appliance inlet dongle serves as a disconnect device and shall be easily
accessible.
4. Alternately, the socket- ll be installed near theoutlet of the end-product sha
equipment and shall be easily accessible or disconnect device provided within
the end-product.
5. Adequate provision for airflow needs to be provided at both the inlet and outlet
vents of the power module. There should be at least 0.5 inches of clear space on
the fan end and on the vented end of the power module. A baffle or air deflector
is recommended at the fan end so that the air intake is raised at least 6 inches
above floor level to prevent dust accumulation inside the power module.
6. The power module may be mounted in any orientation. Four 8-32 threaded
inserts are provided on the surface with the AC input cord to facilitate mounting.
7. The AC voltage input and output connections are made with user supplied cables
to the IEC input and output connectors on the power module. Use only with
approved plug/cord sets according to the equipment electrical ratings. No
external fusing is needed since safe operation is assured by internal non-
replaceable 10A fuses. However, if a customer wishes to prevent the internal
fuses from blowing in the event of an extreme overload, then an external fuse or
circuit breaker rated at 6A to 10A (120VAC models) or 3A to 5A (230V models)
should be provided on the AC voltage input.
8. The 12V battery input is made with a user supplied cable to the Anderson DC
connector on the power module. Proper polarity must be observed. The
recommended wire size is 8AWG in order to keep the voltage drop in the cable to
less than 0.1V. The length of cable between the battery and the power module
should be less than 30 inches. A user supplied 40A to 50A automotive style fuse
is required in the positive lead of the battery (within 18 inches of the battery
positive terminal).
1000 Eaton Boulevard
Cleveland, OH 44122
United States
TrippLite.Eaton.com
Tripp Lite is now part of Eaton.
23-08-312-932965 REV E
9. Additional bonding is needed to connect the battery negative terminal to the case
of the power module. The wire (or metal frame of the cart) should be short (less
than 18 inches) and preferably of the same gauge as the other battery wires
(although a minimum gauge of 16AWG has been acceptable). This bonding
connection is necessary to prevent internal damage to the power module in the
event of any accidental reversed polarity connections or accidental connections
between the battery positive wiring and the case of the power module.
10. A Remote User Interface (RUI) module must be connected to the power module
using two CAT-5 cables. The COMM1 jacks should be connected together and
the COMM2 jacks should be connected together. If the cables are not connected
or if they are interchanged, the power module will not operate. The RUI may be
either a Tripp Lite RUI or a customer designed RUI that meets Tripp Lite
specifications.
11. The power module includes a low voltage cutoff (LVC) function that will remove
any load on the battery when the battery is depleted. The LVC level is about
10.5V to assure optimum battery cycle life. If any other DC loads are present in
the system, they need to include their own LVC function in order to prevent
damage to the battery due to over discharge.
12. The power module will not turn on unless connected to a battery that has a
voltage greater than 9.6V. If the battery voltage is less than 9.6V, it is assumed
to be defective and the charger is inhibited in order to prevent possible
outgassing of the battery.
13. The power module reports % Battery Capacity through its USB port. This data is
most accurate when a 55Ah sealed lead acid absorbed glass mat (AGM) battery
is used. Batteries of other capacities may be used, but the accuracy of the
% Battery Capacity report will be reduced.
14. After connecting a battery to the power module, the power module should be
allowed to charge the battery to full capacity in order to initialize the % Battery
Capacity report and the % Battery Capacity LED’s (on the RUI). This will take at
least 4 hours, even with fresh battery. The % Battery Capacity report and LED’s
will not be accurate until after the initial charge has been completed. After
initialization, the power module will properly indicate the % Battery Capacity as
the battery is repeatedly discharged and recharged. However, each time the
battery is replaced, the initialization will need to be repeated.
15. CAUTION: Double Pole/Neutral Fusing.
Technical Details
T
y
pe of electrical suppl
y
s
y
stem: TN
In
g
ress protection: IP20
Stora
g
e environment: 0 to 95% RH non-condensin
g
, 32-122Fº
(
0-50ºC
)
.
A
mbient operatin
g
temperature ran
g
e: 32-104Fº
(
0-40ºC
Relevant standards for manufacture, test, or use: IEC/EN 62040-1, IEC 60601-1
23-08-312-932965 REV E
Guías de Aplicación para los Módulos de Potencia de las Series
HC350S y HCINT350S
Series AGSM5396, AGTR0829
GUARDE ESTAS INSTRUCCIONES – Este manual contiene INSTRUCCIONES DE
SEGURIDAD IMPORTANTES que deben seguirse durante la instalación y el
mantenimiento del UPS y las baterías.
Los módulos de potencia para carros de hospital de las series HC350S y HCINT350S
están diseñados para proporcionar una potencia de salida de voltaje de CA
suministrada por la red pública o a partir de la inversión de potencia de 12VCD de una
batería de plomo-ácido sellada. Para garantizar el funcionamiento correcto, deben
seguirse las siguientes guías:
1. El módulo de potencia puede admitir cargas de CA hasta un máximo de 400VA o
300W.
2. El cargador del módulo de potencia está diseñado para usarse con una batería
de fibra de vidrio absorbente [AGM] sellada de 12V con especificación de 33Ah a
66Ah . No se recomienda el uso de baterías con especificación menor que 33Ah
ya que la corriente del cargador de 12A reduciría el ciclo de vida de la batería en
el caso de las baterías más pequeñas.
3. El dongle de entrada del aparato sirve como dispositivo de desconexión y
deberá ser fácilmente accesible.
4. Alternativamente, el tomacorriente del producto final se debe instalar cerca del
equipo y debe ser fácilmente accesible o se proporcionará un dispositivo de
desconexión dentro del producto final.
5. Es necesario proporcionar un flujo de aire adecuado en las rejillas de entrada y
de salida del módulo de potencia. Debe haber al menos 1.3 cm [0.5"] de espacio
libre en el extremo del ventilador y en el extremo ventilado del módulo de
potencia. Se recomienda colocar un deflector de aire en el extremo del ventilador
de modo que la entrada de aire se eleve al menos 15 cm [6"] sobre el nivel del
suelo para evitar la acumulación de polvo en el módulo de potencia.
6. El módulo de potencia puede instalarse en cualquier orientación. Para facilitar la
instalación, se proporcionan cuatro insertos roscados 8-32 sobre la superficie
con el cable de alimentación de CA.
7. Las conexiones de entrada y salida del voltaje de CA están hechas con cables
suministrados por el usuario a los conectores IEC de entrada y salida en el
módulo de potencia. Utilice solo con una clavija y cable de alimentación
aprobados según las especificaciones eléctricas del equipo. No se necesitan
fusibles externos p3-ya que la operación segura está garantizada por fusibles
internos de 10A no reemplazables. Sin embargo, si un cliente desea evitar que
los fusibles internos se quemen en caso de una sobrecarga extrema, entonces
debe instalar un fusible externo o breaker de 6A a 10A (modelos de 120VCA) o
3A a 5A (modelos de 230V) en la entrada de voltaje de CA.
8. La entrada de la batería de 12V está hecha con un cable suministrado por el
usuario al conector de CD Anderson en el módulo de potencia. Debe respetarse
la polaridad correcta. El calibre de cable recomendado es 8AWG a fin de


Produktspezifikationen

Marke: Tripp Lite
Kategorie: Batterie
Modell: HCINT350SNR
Breite: 152.4 mm
Tiefe: 292.1 mm
Gewicht: 6530 g
Produktfarbe: Grau
Höhe: 88.9 mm
Stecker: C13; C14
Verpackungsbreite: 221 mm
Verpackungstiefe: 350.5 mm
Verpackungshöhe: 188 mm
Paketgewicht: 7300 g
Warentarifnummer (HS): 85044095
Ursprungsland: China
Menge pro Packung: 1 Stück(e)
Gehäusematerial: Stahl
LED-Anzeigen: Power, Status
Anzahl Produkte pro Versandkarton: 1 Stück(e)
Gewicht Versandkarton: 7300 g
Länge des Versandkartons: 350.5 mm
Breite des Versandkartons: 221 mm
Höhe des Versandkartons: 188 mm
Netzteiltyp: Drinnen
Anzahl der USB-Anschlüsse: 1
Ausgangsspannung: 230 V
Ausgangsleistung: 350 W
Eingangsspannung: 230 V
Nachhaltigkeitszertifikate: RoHS
Eingangsstrom: 15 A
DC-Ausgang Volt: +12V
Eingangsfrequenz: 50 Hz
Ausgangskapazität: 0.35 kVA
Output Frequenz: 50 Hz