Megohmmeter
PCE-MO 1500
Safety Requirements & Safety Warnings:
This series of insulation resistance tester is designed and produced according to the safety
standard IEC1010. Strictly follow the safety standards of double insulation CAT III 1000V
CAT IV and pollution degree 2. To protect your safety, please read the manual before using
it.
Precautions:
Safety Symbol Description:
Caution! Refer to the user’s manual before using this meter. Double Insulation
Warning! Dangerous Voltage (Risk of electric shock).
Danger: To avoid serious or fatal damage that may be caused by certain states and wrong
operations.
Warning: indicates avoiding the risk of electric shock.
Caution: indicates avoiding damage to the instrument and making accurate measurements.
· Do not test in flammable locations, sparks may cause an explosion.
· Do not operate the instrument if the instrument surface is wet or if the operator's hands
are wet.
· When testing voltage, there is a possibility of personal injury due to accidental
short-circuit between metal parts and test leads.
· Do not exceed the maximum allowable range of the measuring range during measure-
ment.
· Do not open the battery cover during measurement.
· When performing insulation measurement, do not touch the line under test.
Warning
· If the instrument is abnormal, please stop using it. For example: Broken instrument or
exposed metal parts, please do not use it in this situation.
· When the test lead is inserted into the instrument interface and connected to the circuit
for testing, do not press the function button, but exit the measurement first, and then
change the test item. ·
Do not install substitute parts or make any unauthorized changes to the instrument.
· Do not replace the battery when the instrument is wet.
· Make sure all test leads are securely connected to the meter's test sockets.
· Make sure the instrument is turned off before opening the battery cover.
Caution:
· Before measuring, confirm that the function button is selected to the appropriate
position.
· After use, press and hold the “OFF” button to turn off the unit. If you do not use it
for a long time, please take out the battery and storage it.
·Do not storage for a long time in high temperature, humidity, places where condensation
may occur, or in direct sunlight.
· Please use a damp cloth or detergent to clean the instrument case, do not use abrasives
or solvents to wipe the instrument case and accessories.
· When the instrument is wet, please dry it before use or storage.
General:
Insulation resistance tester (or high voltage insulation meter) is a special instrument for
measuring insulation resistance. It is an ideal test instrument for measuring insulation
resistance of large transformers, transformers, generators, high-voltage motors, power
capacitors, power cables, arresters, etc.
· Automatic release voltage function.
· Suitable for measuring insulation resistance of various electrical equipment and
insulating materials such as transformers, motors, cables, switches, electrical appliances, etc.
It is suitable for the maintenance, repair, test and verification of various electrical
equipment.
· Rated output voltage variable
The rated output voltage can be switched by the function buttons "▲" "▼":
1000V switches between 50V, 100V, 250V, 500V, 1000V
2500V switches between 250V, 500V, 1000V,2000V, 2500V.
5000V switches between 500V, 1000V, 2500V, 5000V.
· High voltage indication
There is a red LED light in the machine indicating that there is high voltage output in the
instrument, there is a danger of electric shock, the operator should pay special attention to
safety.
· Low battery voltage indication " ".
· Battery-driven, easier to use.
· 2000 digits with bar graph display, easy to read the data.
Measuring range:
1000V: 0.1MΩ 20GΩ~
2500V: 1MΩ~200GΩ
5000V: 10M~200GΩ, automatic range conversion.
·Easy to operate and carry.
· Strong load capacity, <1.8mA output short-circuit current.
· Complete protection circuit, not afraid of voltage counterattack, output short
circuitprotection.
· It can measure AC/DC voltage 10V-600V, AC voltage frequency 40Hz ~ 70Hz.
· Dust-proof and moisture-proof structure, suitable for outdoor operation.
· Dimensions: 180X140X70 (mm)
· Weight: about 800g (including battery)
A. Technical Data
Working humidity conditions: 0°C~35°C, relative 75% or less, 2000 meters above sea
level
Output voltage: (output ±10% load >=1000M )Ω
Short circuit current: about less than 1.8mA;
Low battery indication: Display shows
Insulation resistance: >=500MΩ (1000V);
Withstand voltage: AC 2KV 50Hz 1 minute;
Working temperature and working humidity: temperature 0°C~40°C, relative humidity
85% or less;
Table 2. 2500V technical parameters
Output voltage: ±10%
250V/500V/1000V/2000V/2500V
B. Appearance
1. Insulation resistance measurement high voltage output jack L
2. Positive terminal of AC/DC voltage measurement
3. Voltage measurement input negative shielding end G (shielding end)
4. Insulation resistance measurement sampling jack E
5. Display LCD screen
6. Insulation resistance measurement high voltage start button
7. Insulation resistance voltage selection button "▲"
8. Insulation resistance voltage selection button "▼"
9. Data hold/backlight button,
10. Data read/save button
11. Function selection button; can choose between insulation resistance/AC voltage/DC
voltage functions
12. Polarization index DAR and absorption ratio PI buttons
13. Power switch button: Press and hold for more than 2 seconds to turn on, and then press
and hold for more than 2 seconds to turn off.
C. Safety Precautions
1. Be sure to read this manual carefully before use, and follow the instructions in sequence.
2. Do not use accessories not provided by the original factory to avoid danger.
3. During the test, there is a DC high voltage output on the measuring terminals E and L of
the instrument. It is strictly forbidden to touch the human body to avoid electric shock.
4. In order to avoid the error caused by the insulation leakage of the test rod itself, the test
rod connected to the measuring end L of the instrument should be suspended as much as
possible, so as not to touch the external objects or the test wire of the E end.
5. When the insulation resistance value of the measured object is high, and the measure-
ment has serious word jumping phenomenon, you can connect the shield terminal "G" of
the instrument. For example: when measuring the insulation between the cable core and the
cable shell, in addition to connecting the two ends of the object to be measured to the "E"
and "L" ends, then connect the inner layer of insulation between the cable shell and the core
to the shield of the instrument Terminal "G" to eliminate measurement errors due to surface
leakage (as shown in below picture). The method of adding a shielding box can also be
used, that is, placing the object under test in a metal shielding box, and then connecting the
shielding box to the shield "G" terminal of the instrument.
6. After the test is completed, remove the connecting wire.
7. The instrument should not be used and stored in sunlight, high temperature and high
humidity places.
8. The long-term short-circuit of the measuring end shall not exceed 60s;
9. The measurement continuous working time should not exceed 30 minutes;
D. Insulation resistance measurement
▲Attention
a. When measuring insulation resistance, please place the two test leads strictly apart, do no
twist them together.
b. Do not short-circuit the two test leads and the high-voltage output in the high-voltage
output state before measuring the insulation resistance.
c. Do not take measurements if the battery cover is open.
Plug the red test lead into the meter's "L" input jack and the black test lead into the "E" inpu
jack.
1. Test connection
Connect the wiring of the "E" end jack of the meter (black test lead with alligator clip) to
one end of the object to be tested; clip the wiring of the "L" end jack of the meter (red test
lead with alligator clip) to the other end of the tested object, You can also hold the test rod
with one hand to make the test needle contact the other end of the object to be tested.
When operating with one hand, the other hand or other parts of the body should not be in
contact with any part of the test circuit. One is for safety, and the other is not to affect the
test line. The accuracy of the measurement; the wiring of the G end jack of the instrument is
a shielded wire (with a large
Test clip), connected to the surface of the DUT to prevent surface leakage from affecting
the test impedance (as shown in Figure A and Figure B).
2.Selection of rated voltage
Press the "▲""▼"key on the instrument panel to select the appropriate voltage.
2500V can choose 250V/500V/1000V/2000V/2500V
3.Insulation resistance measurement
Press and hold the " " button for >=2 seconds to turn on the machine. After turning on the
machine, it is recognized as insulation resistance measurement, and the display shows"---".
Press keys the "▲" and "▼" on the instrument panel to select the appropriate voltage, and
the selected voltage is displayed in the lower right corner of the display, connect the test lead
of the test rod to the two ends of the test object, press the "TEST" button on the instrument
panel to output the insulation resistance test voltage, and the test light in the button emits a
red warning, the lower right corner of the display screen displays the insulation resistance test
voltage, and the display .The measurement time is displayed at the bottom left of the screen,
and the insulation resistance value is displayed in the middle of the screen. Press the "TEST"
button again to end and the measurement, the test light in the "TEST" button goes out, and
the measured value of insulation resistance remains in the middle of the display. Notice
*Before testing, make sure there is no electricity in the circuit under test, do not measure liv
equipment or insulation with circuits.
*After the measurement, do not touch the circuit with your hands, the stored capacitance of
the circuit can cause clicks.
*Do not measure if the battery cover is open.
E.Absorption ratio DAR measurement
When than the insulation resistance measurement time is greater 1 minute, the measure-
ment is ended. Press the the "DAR/PI" key on the instrument panel and display will display
the calculate "DAR" symbol. The instrument will automatically the absorption ratio and
display it in resistance the The lower left corner of the display; ratio of the insulation value at
1 minute to the insulation resistance value at 30 seconds.
F. Polarization Index PI Measurement
After the the insulation 10 resistance measurement time is greater than minutes, measure-
ment is ended. Press the "DAR/PI" key on the instrument until panel the display shows
the The "PI" symbol. instrument will automatically calculate the polarization index value
and display it in the lower left corner of the The display; "Polarization" index PI" is the
ratio of the insulation resistance value value at 10 minutes to the insulation resistance at 1
minute.
Insulating materials and all have the absorption process polarization process of charge after
adding high be voltage. The power system requires that the polarization index should
measured in the the insulation test of main transformer, cable,
motor, judge etc., and this data can be used to the quality of insulation, quality and aging
problems.
G. AC voltage test
1. After power on, press the "SELECT" key on the instrument panel to select AC voltage
measurement, the display shows the voltage "AC" unit symbol V.
2. Insert the red test lead into the "AC/DC600V" jack, and the black test lead into the "G"
jack.
3. Touch the two test leads to the two ends of the voltage to be measured, and the display
shows the measured AC voltage in volts.
Notice:
*The measured voltage should not exceed the rms value of 600V AC voltage, and the
frequency of the AC voltage is 40Hz to 70Hz. The data error measured beyond this
frequency range will exceed the technical specifications of the machine.
*After completing all measurements, disconnect the test leads from the circuit under test
and remove the test leads from the instrument input
*Do not measure if the battery cover is open.
I. DC Voltage Test
1. After power on, press the "SELECT" key on the instrument panel to select DC voltage
measurement, the display shows the voltage "DC" unit symbol V.
2. Insert the plug end of the red test lead into the "AC/DC600V" jack, and the plug end of
the black test lead into the "G" input jack.
3. Put the two test leads in contact with the two ends of the voltage to be measured, and the
measured DC voltage in volts will be displayed on the display.
Notice:
*The measured voltage should not exceed the rms value of 600V DC voltage.
*After completing all measurements, disconnect the test leads from the circuit under test
and remove the test leads from the instrument input
*Do not measure if the battery cover is open.
J.Data Hold/Backlight Operation
1. Press the "HOLD/*" button for more than 2 seconds, the backlight will be on; press the
"HOLD/*" button again for more than 2 seconds, the backlight will be off;
2. Press "HOLD/*" to hold data, press again to exit data hold.
H. Store data/read data
1. Store data
After measuring the insulation resistance value, press the "HOLD/*" key to keep the
measurement data first, the display shows the "HOLD" symbol, and then press " " to save
the data, the "MEM" symbol is displayed at the bottom of the display to store the current
measurement data, Long press the " " key to exit the storage, repeating this operation,
the instrument can store 100 sets of data.
2.Read data After power on, press the " " button, the "READ" symbol will be displayed at
the bottom of the display to enter the reading data, press the button "▲""▼" to read up and
down other stored data; long press " " to exit the data reading.
3. Delete data
After power on, press the" " button, the "READ" symbol will be displayed at the bottom
of the display to enter the data reading mode, press " " to delete the current data, after the
data is deleted, the display will display "--", press the "▲""▼" button to go up Scroll down
to other groups of data, and then press the " " key to delete the data to be deleted in other
groups; press
"DAR/PI" to delete all data.
I. Shut down
After the test is completed, press hold and the
" " key for more seconds, than 2 the LCD
screen will not display, and the power of the
instrument will be off. turned For capacitive
loads, the the residual charge on test product first prevent should be discharged to the
residual charge remove discharge from hurting people. Then the test leads.
J. Replacement battery
If "E3" is LCD displayed on the during use, the battery capacity is insufficient, please replace
the battery. battery, sure are Before replacing the make the test leads disconnected and the
power is turned off.
1. Stop all test work and disconnect the test leads from the item under test.
2. Disconnect the test leads from the meter.
3. Remove the battery cover from the back of the meter.
4. Put in 8 batteries. The model is (AA) /1.5V, pay attention to the position and polarity of
the batteries.
5. Install the battery cover and screw on the battery cover.
PI Polarization Index
Measurement
10 minutes insulation resistance / 1 minute insulation
resistance
PI Polarization Index
Measurement
4
4---2
2.0--1.0
Judgement standard
best
good
warning
DAR absorption ratio
measurement
1 minute insulation resistance / 15 seconds insulation
resistance
DAR absorption ratio
measurement
≥1.4
1.25--1.0
Judgment standard
best
good


Produktspezifikationen

Marke: PCE Instruments
Kategorie: Messgeräte
Modell: PCE-MO 1500

Brauchst du Hilfe?

Wenn Sie Hilfe mit PCE Instruments PCE-MO 1500 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten




Bedienungsanleitung Messgeräte PCE Instruments

Bedienungsanleitung Messgeräte

Neueste Bedienungsanleitung für -Kategorien-