Differential Pressure Transmitter
PCE-DDMU 10
INSTRUCTION MANUAL
Table f ontents o c
1 General nformation. i
2 Product dentification. i
3 Installation.
4.Initial tart s -up
5 Operation.
6 Dimensions Fixing. /
7 Maintenance.
8 Return.
9 Warranty conditions.
Operating Manual
Dierential Pressure Transmitter PCE-DDMU 10
Menu button
1 General information
Do not blow into the pressure ports!
This causes damage to the device.
- This operating manual contains important informa-
tion on proper usage of the device. Read thisthe
operating manual carefully before installing and
starting up the pressure measuring device.
- c-Adhere to the safety notes and operating instru
tions which are given in the operating manual. Addi-
tionally applicable regulations regarding occupa-
tional safety, accident prevention as well as national
installation standards and engineering rules must
be complied with!
- This operating manual is part of the device. It must
be kept near its location and always be accessible
to all employees.
- This operating manual is copyrighted. The contents
of this operating manual reflect the version available
at the time of printing.
WARNING! To avoid operator hazards and dam-
age to he device, the following instructions t may
only be carried out by qualified technical perso n-
nel.
Limitation of liability: In -case of non
observance of the operating manual, inappropri-
ate use, modifications or damage, no liability is
assumed and warranty claims will be excluded.
Symbols used:
: Warning
!: Caution
: Note
– T –echnical modifications reserved
2 Product identification
2.1 Intended use
-
-
T differential pressurhe PCE-DDMU 10 e
transmitter has developed for t been he
measurement of pressur difference dry, non e of -
aggressiv gases pressuriz aire and ed and an bec
us for wide r of different HVACed a ange
applications.
Its robust desi means it c us i the gn an be ed n
laboratory under industrial conditions. and
Preferred e e n areas of us ar i heating, ventilation
and ng ean air conditioni systems; cl room and
medical technology, filter technology draft and
meteri checksng .
It s he operator's responsibility to check and verify i t
the suitability f he device or intended o t f the applica-
tion. f any bts emain, ease contact our les I dou r pl sa
department n order o ensure proper se. i t u
PCE Instruments is not liable for any incorrect
selections and heir t effects!
WARNING! – Danger by inappropriate use!
2.2 Identification
- The device can be identified by its manufacturing
label. It provides the most important data. T he
product can be clearly identified by the ordering
code.
Fig. 1: Manufacturing label
!The manufacturing label may not be removed
from the device!
2 .3 Package contents
Please verify that all listed parts are included in the
delivery undamaged and comply with your order :
1 Differential ressure Transmitter P PCE-DDMU 10
1 Operating manual
2.4 Transport and storage
Permissible temperature for storage: 10 °C to + 70 °C-
For transport make sure that both pressure inputs of
differential pressure sensors are open. The air transport
of absolute pressure sensors should only be done in a
pressure compensated cabin.
2.5 Declaration of onformity c / CE
The delivered evice all egal equirements. d fulfills l r
T applicablehe directives, harmonised standards d an
documents are listed in the EC eclaration of onformity d c .
The operational safety is additionally confirm by the ed
CE si on he manufacturing label. gn t
3 Installation
3.1 Mounting nd afety a s instructions
WARNING! This device may only be installed by
qualified technical personnel who have read and
understood the operating manual!
WARNING! Install the device only when depres-
surized and current free- !
!Handle this device with care, both in ed and packag
unpackaged condition!
!No modifications/changes should be made on the
device.
!Do not throw the device!
!Do not use any force when installing the device, to
prevent damage of the device and the plant!
!Avoid of sources interference (transformers,
transmitters, motors etc.) or sources of heat in the
surrounding area .
!Shocks or vibrations at the mounting connection
can cause distortions to the output signals.
3.2 Installation steps
- Carefully remove the pressure measuring device
from its package and dispose of the package prop-
erly.
- Mount the device in a suitable location using both
straps. The corresponding dimensions are provided
under "6".
- Connections should be vertical, i. e. the pressure
connections should point downwards. The sensors
are calibrated in the factory for these mounting con-
ditions. In addition, the formation of condensation in
the pressure tubing of the sensor is also reduced.
- Use the appropriate hoses for the connection.
Fig. 2: PCE-DDMU 10
3.3 installation Electrical
WARNING! This device may only be installed by
qualified technical personnel who have read and
understood the operating manual!
- Open the top cover and establish the electrical
connection of the device according to the follow-
ing description. Screw the top cover onto the box
again.
- Guide externally the power supply line through
the right so that the wires can cable fitting be
connected with the right terminal block (SUP-
PLY) without any problem.
- Next, tighten the cable gland by hand. Make
sure, however, that the cable is strain-relieved.
3.4 Pressure port
- Connect the hose with the higher pressure to the
right hand (positive) pressure port.
- Connect the hose with the lower pressure to the
left hand (negative) pressure port.
4 - Initial start up
After switching on the power supply, the output
signal can be measured. Variations in the output
signals may have two possible causes:
1. The warm up time of the sensor is about 30min- .
After this period, the sensor signal should be
stable for zero differential pressure and constant
ambient temperature.
2. For small pressure ranges, a slight deviation in
the zero point due to ambient conditions can oc-
cur. This error can be corrected by adjusting the
zero-point potentiometer of the sensor after the
warm- up time. (Set the output signal of the sen-
sor with both pressure inputs open to the nomi-
nal value.)
5 Operation
5 .1 General information
WARNING! - check Before start up, the user must
to ensure proper installation and for any visible de-
fects.
WARNING! The device may only be started and
operated by authorized personnel who have read
and understood the operating manual!
WARNING! The device may only be used within
the technical specifications the data in the (see
data sheet)!
WARNING! – Do not touch the bonding wires this
ca . n destroy the device
5.2 Setting of damping
P otentiometer for setting the damping is located
adjacent to the display on the left (see image 2).
The damping of the device can be set by turning a
size 2 Phillips screwdriver in the area of 0 to
5000ms.
5.3 Menu buttons
Zeroing: secondpush left menu button for at least 1
5.4 Configuration switch
3- - :wire system
1 2 3 4
Off / 0 ... 10 V 0 ... 20 mA
On 5 V0 ... / 4 ... 20 mA
Off Automatic zero adjustment
off
On Zero adjustment active at
start and for 24h
Nominal pressure range
Off Off 4 1,6 10 40 250 1000
Customized ranges
6 On Off 1,0 2,5 25 60 400
Off On 160 600
2- -wire system:
1 2 3 4
Off Off Automatic zero adjustment
off
On O ff Zero adjustment active at
start
O ff On Zero adjustment active at
start and for 24h
On On Zero adjustment active at
start, then every 7 days
Nominal pressure range
Off Off 4 1,6 10 40 250 1000
Customized ranges
On Off 6 1,0 2,5 25 60 400
Off On 160 600
The switches 1 and 2 don't have any function
at special pressure ranges.
terminal block
2- -wire system
Ordering code
Connector pinout
Type designation
Supply
Nominal
pressure
range
Sig nal
Serial
number
Display
(optional)
Potentiometer for adjust-
ing damping
P - hillips screwdriver
size 2
Configuration
switching
Negative
pressure connection
Positive
pressure connection
Cable gland M16x1.5
terminal block
3 wire system
3- -s 3- -wire ystem )(current / voltage wire system ( )current / voltage
w ith two outputs
supply terminal-: 7 - : VS
supply +: terminal 8: VS+
signal: 9 terminal : IOUT
signal: terminal 10: VOUT
Output 2 NO2: terminal 1 : NO2
Output 2 C2: terminal 2 : C2
Output 2 NC2 terminal 3 : NC2
Output 1 NO1 terminal 4 : NO1
Output 1 C1 terminal 5 : C1
Output 1 NC1 terminal 6 : NC1
supply -: terminal 7: VS-
supply +: terminal 8: VS+
signal: terminal 9 : IOUT
signal: terminal 10: VOUT
2- -s 2- - wire ystem (current) wire system (current) with two outputs
supply +: terminal 1: VS+
supply terminal-: 2 - : VS
supply +: terminal 1 + : VS
supply terminal-: 2 - : VS
output 1: terminal 3: S1
output 2: terminal 4: S2