Thank you for purchasing the OMRON RS6 Wrist Blood Pressure
Monitor.
The OMRON RS6 is a compact and easy to use blood pressure
monitor, operating on the oscillometric principle. It measures your
blood pressure and pulse rate simply and quickly. For comfortable
controlled inflation without the need of pressure pre-setting or
re-inflation the device uses its advanced “IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse
rate of people within the range of the designated wrist cuff, following
the instructions in this instruction manual.
It is mainly designed for general household use. Please read the
Important Safety Information in this instruction manual before using
the unit.
Please read this instruction manual t horoughly before using the
unit.
Please keep for future reference.
For specific information about your own blood pressure,
CONSULT YOUR DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with
arrhythmia or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results
and self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should
consult a doctor before using the unit, as cuff inflation can cause
internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
(General Usage)
• Do not leave the unit unattended wit h infants or persons who
cannot express their consent.
• Do not use the unit for any purpose other than measuring blood
pressure.
• Do not use a mobile phone or other devices that emit
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit.
• Do not disassemble the unit or wrist cuff.
Do not operate the unit in a moving vehicle (car, airplane).
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse
with plenty of clean water.
• Use only two “AAA” alkaline (LR03) batteries with this unit. Do not
use other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
Replace old batteries with new ones immediately. Replace both
batteries at the same time.
• Remove the batteries if the unit will not be used for three months or
more.
Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol or
caffeine, smoking, exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your
wrist.
• Read and follow the “Important information regarding Electro
Magnetic Compatibility (EMC)” in the Technical Data Section.
Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any used
accessories or optional parts.
Main unit:
Display:
2.1 Installing/Replacing the Batteries
• If the battery low symbol ( ) appears on the display, turn off
the unit then replace both batteries at the same time.
• The measurement values continue to be stored in memory
even after the batteries are replaced.
Disposal of used batteries should be carried out in accordance
with the national regulations for the disposal of batteries.
2.2 Setting the Date and Time
Notes:
• If the batteries have been removed, the date and time setting
will need to be reset.
If the date and time are not set, “-:--” appears during or after
measurement.
3.1 Applying the Wrist Cuff
Do not apply over clothing.
3.2 How to Sit Correctly
Relax your wrist and hand. Do not bend your wrist back, clench
your fist, or bend your wrist forward.
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button at
any time during measurement.
• Remain still while taking a measurement.
After the Positioning symbol lights for more than 2 seconds,
the wrist cuff will start to inflate automatically.
Cuff Wrapping Guide
The Cuff Wrapping Guide is a unique feature that
indicates if the cuff is not wrapped tightly enough
around the wrist. Even when the is displayed, a
blood pressure reading will be taken.
Note: This reading is NOT reliable due to the incorrect
wrapping of the cuff. Please wrap the cuff again, taking
care to wrap it correctly and take the measurement
again. When the is displayed, the cuff is correctly
wrapped tightly enough on the wrist and the reading is
accurate and reliable.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after two minutes.
This criteria is for home blood pressure measurement.
If the irregular heartbeats cause the measurement to be
invalid, no result is shown. If the irregular heartbeat symbol
( ) is shown after you have taken a measurement, repeat
the measurement. If the irregular heartbeat symbol ( ) is
shown frequently, please make your doctor aware of it.
The blood pressure can differ between the right arm and the
left arm, and therefore also the measured blood pressure
values can be different. OMRON recommends to always use
the same arm for measurement. If the values between the
two arms differ substantially, please check with your
physician which arm to use for your measurement.
Warning:
Self-diagnosis of measured results and treatment are
dangerous. Please follow the instructions of your doctor.
3.4 Setting the Positioning Sensor
Wrist Blood
Pressure Monitor
Model RS6
Instruction Manual
IM-HEM-6221-E-01-08/2012
1883267-9A
Introduction
Important Safety Information
A. Wrist cuff D. Memory ( ) button
B. Batt ery compartment E. Date/Time setting ( ) button
C. Display F. START/STOP button
G. Memory symbol
H. Systolic blood pressure
I. Diastolic blood pressure
J. Battery low symbol
K. Heartbeat symbol
(Flashes during measurement)
L. Pulse display and Memory
number
M. Irregular heartbeat symbol
N. Movement error symbol
O. Deflation symbol
P. Average value symbol
Q. Date/Time display
R. Blood pressure level indicator
S. Positioning indicator
T. Positioning symbol
U. Cuff wrapping guide
1. Remove the battery cover.
2. Insert two 1.5V “AAA”
alkaline (LR03) batteries
as indicated in the battery
compartment and then
replace the battery cover.
Notes:
1. Press the button.
2. Set the monitor to the correct date and time
before taking a measurement for the first time.
3. Press the START/STOP button to store the setting.
1. Overview
A
E
F
C
D
B
G
P
I
J
K
LMNO
U
T
S
Q
R
H
2. Preparation
Year
Month
Day
Hour
Minute
1) Push the button to change.
Hold down to advance the digit
rapidly.
2) Push the button to confirm.
Next setting appears.
1. Put your arm through the
cuff loop.
Your palm should face upward.
2. Wrap the wrist cuff around
your wrist.
Wrap the wrist cuff securely
around the wrist for taking
accurate measurements.
Make sure that the wrist cuff
does not cover the protruding
part of the wrist bone (ulna) on
the outside of the wrist.
Notes:
• You can take a measurement on
either your left or right wrist.
• If the wrist cuff comes off,
assemble as shown in the figure.
To take a measurement, you need
to be relaxed and comfortably
seated, under comfortable room
temperature. No bathing, drinking
alcohol or caffeine, smoking,
exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the
floor.
• Sit upright with your back straight.
The cuff should be at the same level
as your heart.
1. Press the START/STOP button.
All the symbols appear on the display.
While your position is improper, the Positioning symbol and the
Positioning indicator flash, when you are in proper position the
Positioning symbol lights on the display.
The monitor is in
proper position.
The monitor is higher
than the heart.
The monitor is lower
than the heart.
3. Using the Unit
1
2
Cuff Wrapping Guide
START INFLATE DEFLATE COMPLETE
Alternating date/time display
2. Undo the wrist cuff and remove the unit.
3.
Press the START/STOP button to turn off the monitor.
Important:
• Recent research suggests that the
following values can be used as a
guide to high blood pressure for
measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
Your blood pressure monitor includes an
irregular heartbeat feature. Irregular
heartbeats can influence the results of the
measurement. The irregular heartbeat
algorithm automatically determines if the
measurement is usable or needs to be
repeated. If the measurement results are
affected by irregular heartbeats but the result
is valid, the result is shown together with the
irregular heartbeat symbol ( ).
If you move during measurement, the
movement error symbol ( ) will appear on the
display. Keep still and repeat the measurement.
Notes:
• Wait 2 - 3 minutes before taking another
blood pressure measurement. Waiting
between readings allows the arteries to
return to the condition prior to taking the
blood pressure measurement.
1. Press and hold the
button to reset the
Positioning sensor.
The Positioning sensor
is set "on" as default.
2. Press the START/STOP button to store the setting.
3.5 Using the Memory Function
The monitor automatically stores the result up to 90 sets.
It can also calculate an average reading based on the
measurements from the last three readings taken within 10
minutes. If there are only two readings in memory for that
period, the average will be based on two readings. If there is
one reading in memory for that period, the average will be
based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest
readings.
• When viewing the reading taken without setting the date and
time, “-:--” is displayed instead of the date and time.
To View the Average Value
To View the Readings Stored in Memory
To Delete All the Values Stored in Memory
Note: You cannot partially delete the stored readings.
4.1 The Icons and Error Messages
4.2 Troubleshooting
4.3 Maintenance
To protect your unit from damage, please observe the
following:
Do not subject the main unit and the cuff to
extreme temperatures, humidity, moisture or
direct sunlight.
Do not disassemble the unit.
Do not subject the unit to strong shocks or
vibrations (for example, dropping the unit on the
floor).
Do not use volatile liquids to clean the main unit.
Do not wash the cuff or immerse it in water.
Do not use petrol, thinners or similar solvents to
clean the cuff.
Do not carry out repairs of any kind yourself. If a
defect occurs, consult your OMRON retail outlet
or distributor as mentioned on the packaging.
Do not store the unit in the following situations:
If the unit is wet.
Locations exposed to extreme temperatures,
humidity, direct sunlight, dust or corrosive
vapours.
Locations exposed to vibrations, shocks or
where it will be at an angle.
Calibration and Service
The accuracy of this blood pressure monitor has
been carefully tested and is designed for a long
service life.
It is generally recommended to have the unit
inspected every two years to ensure correct
functioning and accuracy. Please consult your
authorised OMRON dealer or the OMRON
Customer Service at the address given on the
packaging or attached literature.
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC
(Medical Device Directive).
• This blood pressure monitor is designed according to the European
Standard EN1060, Non-invasive sphygmomanometers Part 1:
General Requirements and Part 3: Supplementary requirements
for electromechanical blood pressure measuring systems.
• This OMRON product is produced under the strict quality system of
OMRON HEALTHCARE Co. Ltd., Japan. The Core component for
OMRON blood pressure monitors, which is the Pressure Sensor, is
produced in Japan.
Made in China
Press the button.
Note: If there are no measurements
results stored in the memory, the
screen to the right is displayed.
1. Press the button, while the average value is
displayed.
The Memory number
appears for a second
before the pulse rate is
displayed. The newest
set is numbered “1”.
2. Press the button repeatedly to view the
readings stored in memory.
When the memory symbol ( ) appears, first press
the button. Then while holding it down, press the
START/STOP button simultaneously for about 2 - 3
seconds.
Error Display Cause Remedy
Irregular heartbeats
are detected.
Remove the wrist cuff. Wait
2 - 3 minutes and then take
another measurement.
Repeat the steps in section
3.3. If this error continues to
appear, contact your doctor.
Movement during
measurement.
Carefully read and repeat the
steps in section 3.3.
Blink The batteries are
low.
You should replace them with
new ones ahead of time.
Refer to section 2.1.
Lit The batteries are
exhausted.
You should replace them with
new ones at once.
Refer to section 2.1.
Wrist cuff not Apply the wrist cuff correctly
Alternating
date/time
display
First Second
Problem Cause Remedy
The reading is
extremely low
(or high).
The wrist cuff is not
at heart level.
Measure while in the
correct posture.
Refer to section 3.2.
The cuff is not
wrapped snugly
around the wrist.
Wrap the cuff correctly.
Refer to section 3.1.
The arms and
shoulders are
tense.
Relax and try taking the
measurement again.
Refer to section 3.3.
Movement or
talking during
measurement.
Remain still and do not
talk during Measurement.
Refer to section 3.3.
Wrist cuff pressure
does not rise.
Air is leaking from
the wrist cuff.
Consult your OMRON
retail outlet or distributor.
Wrist cuff deflates
too soon.
The wrist cuff is
loose.
Apply the cuff correctly so
that it is firmly wrapped
around the wrist. Refer to
section 3.1.
The blood pressure is different
each time. The reading is
extremely low (or high).
Blood pressure readings
constantly vary with time
of day and how relaxed
you are. Take several
deep breaths and try to
remain relaxed before
taking a measurement.
The unit loses
power during
measurement. The batteries are
exhausted.
Replace the batteries with
new ones.
Refer to section 2.1.
Nothing happens
when you press
the buttons.
The batteries have
been inserted
incorrectly.
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.1.
Other problems.
Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
Product Description Wrist Blood Pressure Monitor
Model OMRON RS6 (HEM-6221-E)
Display LCD Digital Display
Measurement Method Oscillometric method
Measurement Range Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/min.
Accuracy Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation Automatic inflation by pump
Deflation Automatic rapid deflation
Memory 90 Measurements
Power Source 2 x 1.5V (LR03, AAA alkaline batteries)
Battery Life Approx. 300 measurements with new
alkaline batteries at a room temperature of
23°C
Applied Part = Type B
Protection Against
Electric Shock
Internally powered ME equipment
Operating temperature/
Humidity
+10 to +40°C / Maximum: 30 to 85% RH
Storage temperature/
Humidity/Air pressure
-20 to +60°C / Maximum: 10 to 95% RH /
700 to 1060hPa
Console Weight Approximately 85 g without batteries
Outer Dimensions Approximately 87 (w) mm x 64 (h) mm x
14 (d) mm (without the wrist cuff)
Measurable
circumference
Approximately 13.5 to 21.5 cm
Cuff Material Nylon and polyester
Package Content Main unit, storage case, battery set,
instruction manual, guarantee card, blood
pressure pass
5. Technical Data
Important information regarding Electro Magnetic
Compatibility (EMC)
With the increased number of electronic devices such as PC’s and
mobile (cellular) telephones, medical devices in use may be
susceptible to electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of
the medical device and create a potentially unsafe situation.
Medical devices should also not int erfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic
Compatibility) with the aim to prevent unsafe product situations, the
EN60601-1-2:2007 standard has been implemented. This standard
defines the levels of immunity to electromagnetic interferences as
well as maximum levels of electromagnetic emissions for medical
devices.
This medical device manufactured by OMRON HEALTHCARE
conforms to this EN60601-1-2:2007 standard for both immunity and
emissions.
Nevertheless, special precautions need to be observed:
Do not use mobile (cellular) telephones and other devices, which
generate strong electrical or electromagnetic fields, near the
medical device. This may result in incorrect operation of t he unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is
available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its
literature, indicates that it should not be
disposed of, with other household wastes at the
end of its working life.
To prevent possible harm to the environment or human health from
uncontrolled waste disposal, please separate this product from
other types of wastes and recycle it responsibly to promote the
sustainable reuse of material resources.
Household users should contact either the retailer where they
purchased this product, or their local government office, for details
of where and how they can return this item for environmentally safe
recycling.
Business users should contact their supplier and check the terms
and conditions of the purchase contract. This product should not be
mixed with other commercial wastes for disposal.
Manufacturer OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
EU-representative OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
www.omron-healthcare.com
Production Facility OMRON (DALIAN) CO., LTD.
Dalian, CHINA
Subsidiary OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG
U.K.
OMRON MEDIZINTECHNIK
HANDELSGESELLSCHAFT mbH
John-Deere-Str. 81a 68163 Mannheim, GERMANY
www.omron-medizintechnik.de
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois
Cedex, FRANCE


Produktspezifikationen

Marke: Omron
Kategorie: Blutdruckmessgerät
Modell: RS6
Display-Typ: LCD
Akku-/Batterietyp: AA
Typ: Automatisch
Menge pro Packung: 1 Stück(e)
Anzahl unterstützter Akkus/Batterien: 4
Tragetasche: Ja
Akkus/Batterien enthalten: Ja
Unterstützung von Positionierung: Oberarm
Speicherfunktion: Ja
Speicherkartensteckplätze: 90
Pulsfrequenzmessgerät: Ja
Datumsaufdruck: Ja
Handschellengröße: 13.5 - 21.5 cm
Diastolischer Blutdurck: Ja
Systolischer Blutdruck: Ja
Mittleren arteriellen Blutdruck: Ja

Brauchst du Hilfe?

Wenn Sie Hilfe mit Omron RS6 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten




Bedienungsanleitung Blutdruckmessgerät Omron

Bedienungsanleitung Blutdruckmessgerät

Neueste Bedienungsanleitung für -Kategorien-