Omron M3 IT Bedienungsanleitung
Omron
Blutdruckmessgerät
M3 IT
Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Omron M3 IT (2 Seiten) in der Kategorie Blutdruckmessgerät. Dieser Bedienungsanleitung war für 7 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet
Seite 1/2
Introduction
Thank you for purchasing the OMRON M3 IT Automatic Blood Pressure Monitor.
The OMRON M3 IT is a compact, fully automatic blood pressure monitor, operating
on the oscillometric principle. It measures your blood pressure and pulse rate simply
and quickly. For comfortable controlled inflation without the need of pressure
pre-setting or re-inflation the device uses its advanced “IntelliSense” technology.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure and
pulse rate in adult patient population who can understand this instruction manual
with the arm circumference range printed on the arm cuff. The device detects the
appearance of irregular heartbeats during measurement and gives a warning signal
with the measurement result.
Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your own blood
pressure, CONSULT YOUR PHYSICIAN.
Important Safety Information
Warning: Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
(General Usage)
Consult your physician before using the device in pregnancy including
pre-eclampsia, or if diagnosed with arrhythmia or arteriosclerosis.
Do not use the device on the injured arm or the arm under medical treatment.
Do not apply the arm cuff on the arm while being on an intravenous drip or blood
transfusion.
Consult your physician before using the device on the arm with an arterio-venous
(A-V) shunt.
Do not use the device with other medical electrical (ME) equipment
simultaneously.
Do not use the device in the area the HF surgical equipment, MRI, or CT scanner
exists, or in the oxygen rich environment.
The air tube or the AC adapter cable may cause accidental strangulation in
infants.
Contained small parts that may cause a choking hazard if swallowed by infants.
(AC Adapter (optional) Usage)
Do not use the AC adapter if the device or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
Plug the AC adapter into the appropriate voltage outlet. Do not use in a
multi-outlet plug.
Never plug in or unplug the power cord from the electric outlet with wet hands.
Caution: Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
(General Usage)
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
People with severe blood flow problems, or blood disorders, should consult a
physician before using the device, as the arm cuff inflation can cause bruising.
If there are any abnormalities during the measurement, remove the arm cuff.
Do not use this device on infants or persons who cannot express their intentions.
Do not inflate the arm cuff more than necessary.
Do not use the device for any purpose other than measuring blood pressure.
Use only the approved arm cuff for this device. Use of other arm cuffs may result
in incorrect measurement results.
Do not use a mobile phone or other devices that emit electromagnetic fields,
near the device. This may result in incorrect operation of the device.
Do not disassemble the monitor or arm cuff.
Do not use in a location with moisture, or a location where water may splash on
the device. This may damage the device.
Do not use the device in a moving vehicle (car, airplane).
Do not take measurements more than necessary. It may cause bruising due to
blood flow interference.
Consult your physician before using the device if you had a mastectomy.
(AC Adapter (optional) Usage)
Fully insert the power plug into the outlet.
When disconnecting the power plug from the outlet, do not pull the power cord.
Be sure to pull from the power plug safely.
When handling the power cord, take care not to do the following:
Wipe the dust off from the power plug.
Unplug monitor when not in use.
Disconnect the power plug before cleaning.
Use only the original adapter designed for this device. Use of unsupported
adapters may damage and/or may be hazardous to the device.
(Battery Usage)
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 “AA” alkaline or manganese batteries with this device. Do not use
other types of batteries. Do not use new and used batteries together.
Remove the batteries if the device will not be used for three months or more.
General Precautions
• Do not forcibly crease the arm cuff or the air tube excessively.
• Do not press the air tube while taking a measurement.
• To unplug the air plug, pull on the air plug at the connection with the monitor, not
the tube itself.
• Do not drop the monitor or subject device to strong shocks or vibrations.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not use the device outside the specified environment. It may cause an
inaccurate result.
• Read and follow the "Important information regarding Electro Magnetic
Compatibility (EMC)" in the "6. Specifications".
• Read and follow the "Correct Disposal of This Product" in "6. Specifications" when
disposing of the device and any used accessories or optional parts.
1. Know Your Device
Contents:
Monitor, arm cuff, instruction manual, storage case, battery set,
USB cable, blood pressure pass
Monitor:
Arm cuff:
Display:
Display symbols:
Irregular Heartbeat Symbol ( )
When the monitor detects an irregular
rhythm two or more times during the
measurement, the irregular heartbeat
symbol ( ) will appear on the display
with the measurement values.
An irregular heartbeat rhythm is defined
as a rhythm that is 25% less or 25%
more than the average rhythm detected
while the monitor is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol ( ) displays with your measurement results,
we recommend you consult your physician. Follow the directions of your
physician.
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2 - 3 minutes.
Take another measurement, remain still during measurement.
Average Value Symbol ( )
The average value symbol is displayed when you press and hold the memory
button for more than 3 seconds. The most recent average value appears on
the display screen.
Cuff Wrap Guide Lamp ( )
If the cuff was wrapped too loosely, it may cause unreliable values. If the
wrapping of cuff is too loose, the cuff wrap guide lamp does not light.
Otherwise lights in green. This is the function which is used as an aid in
determining if the cuff is wrapped snugly enough.
Blood Pressure Colour Indicator
If your systolic or diastolic pressure is above the
standard range (135 mmHg for the systolic blood
pressure and/or 85 mmHg for the diastolic blood
pressure), the blood pressure colour indicator will light
in orange when the measurement result is displayed.
If they are within the standard range, no light will
appear.
The JNC7* Guideline recommends the following guideline.
These are form statistical values for blood pressure.
* JNC7: The Seventh Report, 2003 Dec, of the Joint National Committee on
Prevention, Detection, Evaluation, and Treatment of Hi Blood
Pressure.
2. Preparation
2.1 Battery Installation
1. Remove the battery cover.
2. Insert 4 “AA” batteries as
indicated in the battery
compartment.
3. Replace the battery cover.
Notes:
• When the low battery symbol ( ) appears on the display, turn the monitor
off, then replace all batteries at the same time. Long life alkaline batteries
are recommended.
• The measurement values continue to be stored in memory even after the
batteries are replaced.
• The supplied batteries may have a shorter life.
Disposal of used batteries should be carried out in accordance with the
national/local regulations for the disposal of batteries.
2.2 Setting the Date and Time
Set the monitor to the correct date and time before taking a measurement for
the first time.
1. Press the button.
2. Push or to change the year
Push button to confirm the year and then the month flashes. Repeat the
same steps to change the month, day, hour, and minutes.
3. Press the START/STOP button to turn the monitor off.
Notes:
• If the batteries have been replaced, the date and time setting will need to be
reset.
• If the date and time are not set, “-:--” appears during or after measurement.
3. Using the Device
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Do not place the arm cuff over thick clothes.
1. Insert the air plug into the air jack
securely.
2. Apply the arm cuff to your left
upper arm.
The bottom edge of the arm cuff
should be 1 to 2 cm above the elbow.
Air tube is on the inside of your arm
and aligned with your middle finger.
3. Secure closed with the
fabric fastener.
Notes:
• When you take a measurement on the right
arm, the air tube will be at the side of your
elbow. Be careful not to rest your arm on the
air tube.
• The blood pressure can differ between the right arm and the left arm, and
therefore also the measured blood pressure values can be different.
OMRON recommends to always use the same arm for measurement. If the
values between both arms differ substantially, please check with your
physician which arm to use for your measurement.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. Avoid
bathing, drinking alcohol or caffeine,
smoking, exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm at
the same level as your heart.
3.3 Taking a Measurement
Notes:
• To cancel a measurement, press the START/STOP button to release the air
in the arm cuff.
• Remain still while taking a measurement.
The monitor is designed to take measurements and store the measurement
values in the memory for 2 people using USER ID 1 and USER ID 2.
1. Select your User ID (1 or 2).
2. Press the START/ STOP button.
The arm cuff will start to inflate automatically.
3. Remove the arm cuff.
4.
Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement result in its memory.
It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes before taking another measurement. Waiting between
measurements allows the arteries to return to the condition prior to
taking a measurement.
Using the Guest Mode
The monitor stores measurement values for 2 users in the memory.
The guest mode can be used to take a single measurement for another user.
No measurement values are stored in the memory when the guest mode is
selected.
1. Press and hold the START/STOP
button for more than 3 seconds.
The USER ID symbol and the Date/Time
display will disappear.
2. Release the START/STOP button when the Date/Time display
turns off.
The arm cuff will start to inflate automatically.
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
3.4 Using the Memory Function
The monitor automatically stores the results up to 60 sets for each user (1 and
2). It can also calculate an average value based on the last 3 measurement
values taken within 10 minutes.
Notes:
• If there are only 2 measurement values in the memory for that period, the
average will be based on these 2 values.
• If there is 1 measurement value in the memory for that period, this is
displayed as the average.
• If the memory is full, the monitor will delete the oldest value.
• When viewing the measurement value taken without setting the date and
time, “-:--” is displayed instead of the date and time.
To View the Measurement Values Stored in Memory
1. Select your User ID (1 or 2).
2. Press the button.
The Memory number appears for a second
before the pulse rate is displayed.
The newest set is numbered “1”.
Note: The cuff wrap guide result appears on
the display with the measurement
values. The cuff wrap guide lamp will not
light.
: The arm cuff was wrapped snugly enough.
: The arm cuff was loose or not wrapped correctly.
3. Press the or button to view the values stored in memory.
: To the older values
: To the more recent values
To View the Average Value
1. Select your User ID (1 or 2).
2. Press and hold the button for
more than 3 seconds.
Notes:
• If the previous measurement was taken without setting the date and time,
the average value is not calculated.
• If there are no measurements results stored in the memory,
the screen to the right is displayed.
To Transfer the Measurement Values
To begin using the Medical Link software for the first time, please visit
www.omron-healthcare.com for the initial set-up instructions.
1. Open the USB port cover.
2. Connect the USB cable to a PC.
3. Connect the USB cable to the monitor.
To Delete All the Values Stored in Memory
1. Select your User ID (1 or 2).
2. Press the Memory button, while the memory symbol ( )
appears.
3. While holding the button
down, press the
START/STOP button for
more than 3 seconds.
Note: You cannot partially delete the values stored in the memory.
All values for the user you select will be deleted.
Do not damage. Do not break it.
Do not tamper with it. Do not forcibly bend or pull.
Do not twist. Do not bundle during use.
Do not pinch. Do not place under heavy objects.
Automatic Blood
Pressure Monitor
Model M3 IT
Instruction Manual
IM-HEM-7131U-E-EN-01-08/2013
2298868-3A
A. Display
B. Memory button
C. START/ STOP button
D. Date/ Time setting button
E. USER ID selection switch
F. Cuff wrap guide lamp
G. Up/ Down buttons
H. Blood pressure colour indicator
I. Battery compartment
J. AC adapter jack
(for optional AC adapter)
K. USB port
L. Air jack
P. OK symbol
Q. Transfer indicator
R. USER ID symbol
S. DATA/FULL symbol
T. Systolic blood pressure
U. Diastolic blood pressure
V. Low battery symbol
W. Cuff wrap guide
X. Heartbeat symbol
(Flashes during measurement.)
Y. Memory symbol
Z. Average value symbol
AA. Date/Time display
AB. Movement error symbol
AC.Irregular heartbeat symbol
AD.Blood pressure level indicator
AE. Deflation symbol
AF. Pulse display / Memory number
General Guidelines for Blood Pressure
Prehypertension at Office Hypertension at Home
Systolic Blood Pressure 120 - 139 mmHg 135 mmHg
Diastolic Blood Pressure 80 - 89 mmHg 85 mmHg
I
L K
J
D
A
B
C
G
F
E
DD
H
M
O
N
M. Arm cuff (Arm circumference 22 - 42 cm)
N. Air plug
O. Air tube
S
T
U
V
Y
R
Q
P
Z
AA
W
X
AB
AC
AD
AE
AF
Normal Heartbeat
Irregular Heartbeat
LongShort
Blood pressure
Blood pressure
Pulse
Pulse
Minute
Change Confirm
Change Confirm
Change Confirm
Change Confirm
Change Confirm
Year
Month
Day
Hour
: Forward
: Back
1 - 2 cm
Cuff Wrap Guide Lamp
START INFLATE DEFLATE COMPLETE
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/ STOP
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Alternating date/ time
display
4. Error Messages and Troubleshooting
4.1 Error Messages
4.2 Troubleshooting
5. Maintenance and Storage
5.1 Maintenance
To protect your device from damage, please observe the following:
• Store the device and the components in a clean, safe
location.
• Do not use any abrasive or volatile cleaners.
• Do not wash the device and any components or immerse
them in water.
• Do not use petrol, thinners or similar solvents to clean the
device.
• Use a soft and dry cloth, or a soft and moistened cloth and
neutral soap to clean on the monitor and the arm cuff.
• Changes or modification not approved by the manufacturer
will void the user warranty. Do not disassemble or attempt to
repair the device or components. Consult your authorised
OMRON retail outlet or distributor.
Calibration and Service
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the device inspected
every 2 years to ensure correct functioning and accuracy.
Please consult your authorised OMRON retail outlet or
distributor.
5.2 Storage
Keep the device in its storage case when not in use.
1. Unplug the air plug from the air jack.
2. Gently fold the air tube into the arm cuff.
Note: Do not bend or crease the air tube excessively.
3. Place the monitor and the arm cuff in the
storage case.
Do not store the device in the following situations:
•If the device is wet.
•Locations exposed to extreme temperatures,
humidity, direct sunlight, dust or corrosive
vapours.
•Locations exposed to vibrations, shocks or
where it will be at an angle.
5.3 Optional Medical Accessories
(within the scope of EC Medical Device Directive 93/42/EEC)
5.4 Other Optional/Replacement Parts
Using the Optional AC Adapter
1. Insert the AC adapter plug
into the AC adapter jack on
the rear side of the monitor.
2. Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the electrical outlet
first and then remove the AC adapter plug from the monitor.
6. Specifications
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, the 5th phase was used on 85 subjects for
determination of diastolic blood pressure.
• This device has not been validated for use on pregnant patients.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and
Part 3: Supplementary requirements for electromechanical blood pressure
measuring systems.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure
monitors, which is the Pressure Sensor, is produced in Japan.
7. Warranty
Thank you for buying an OMRON product. This product is constructed of high
quality materials and great care has been taken in its manufacturing. It is designed
to give you every satisfaction, provided that it is properly operated and maintained
as described in the instruction manual.
This product is guaranteed by OMRON for a period of 3 years after the date of
purchase. The proper construction, workmanship and materials of this product is
guaranteed by OMRON. During this period of guarantee OMRON will, without
charge for labour or parts, repair or replace the defect product or any defective
parts.
The guarantee does not cover any of the following:
a. Transport costs and risks of transport.
b. Costs for repairs and / or defects resulting from repairs done by unauthorised
persons.
c. Periodic check-ups and maintenance.
d. Failure or wear of optional parts or other attachments other than the main device
itself, unless explicitly guaranteed above.
e. Costs arising due to non-acceptance of a claim (those will be charged for).
f. Damages of any kind including personal caused accidentally or from misuse.
g. Calibration service is not included within the guarantee.
h. Optional parts have a one (1) year warranty from date of purchase. Optional parts
include, but are not limited to the following items: Cuff and Cuff Tube, AC Adapter,
USB Cable.
Should guarantee service be required please apply to the dealer whom the product
was purchased from or an authorised OMRON distributor. For the address refer to
the product packaging / literature or to your specialised retailer.
If you have difficulties in finding OMRON customer services, contact us for
information.
www.omron-healthcare.com
Repair or replacement under the guarantee does not give rise to any extension or
renewal of the guarantee period.
The guarantee will be granted only if the complete product is returned together with
the original invoice / cash ticket issued to the consumer by the retailer.
Made in China
Error Display Cause Solution
Irregular heartbeats are
detected.
Remove the arm cuff. Wait 2 - 3
minutes and then take another
measurement. Repeat the steps
in section 3.3. If this error
continues to appear, contact
your physician.
Movement during measurement. Carefully read and repeat the
steps in section 3.3.
The batteries are low.
You should replace the batteries
with new ones ahead of time.
Refer to section 2.1.
The batteries are exhausted.
You should replace the batteries
with new ones at once.
Refer to section 2.1.
Air plug disconnected. Insert the plug securely.
Refer to section 3.1.
Arm cuff is applied too loosely. Apply the arm cuff tighter.
Refer to section 3.1.
Air is leaking from the arm cuff. Replace the cuff with a new one.
Refer to section 5.3.
Movement during measurement
and the arm cuff has not been
inflated sufficiently.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the cuff manually until it is
30 to 40 mmHg above your
previous measurement result.
Refer to section 3.3.
The arm cuff was inflated above
299 mmHg when inflating the
cuff manually.
Do not inflate the cuff above 299
mmHg.
Refer to section 3.3.
Movement during measurement.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
Clothing is interfering with the
arm cuff.
Remove any clothing interfering
with the arm cuff.
Refer to section 3.1.
Device error. Contact your OMRON retail
outlet or distributor.
Problem Cause Solution
The measurement
result is extremely
high (or low).
Arm cuff is applied too
loosely.
Apply the arm cuff tighter.
Refer to section 3.1.
Movement or talking during
measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Clothing is interfering with
the arm cuff.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Arm cuff pressure
does not rise.
The air connector is not
securely connected into the
air jack.
Make sure that the air tube
is connected securely.
Refer to section 3.1.
Air is leaking from the arm
cuff.
Replace the arm cuff with a
new one.
Refer to section 5.3.
The cuff wrapping
guide lamp does not
light.
Arm cuff deflates too
soon.
The arm cuff is loose.
Apply the cuff correctly so
that it is firmly wrapped
around the arm.
Refer to section 3.1.
Cannot measure or
the results are too low
or too high.
The arm cuff has not been
inflated sufficiently.
Inflate the cuff so that it is 30
to 40 mmHg above your
previous measurement
result.
Refer to section 3.3.
Nothing happens
when you press the
buttons.
The batteries are empty.
Replace the batteries with
new ones.
Refer to section 2.1.
The batteries have been
inserted incorrectly.
Insert the batteries with the
correct (+/-) polarity.
Refer to section 2.1.
Other problems.
• Press the START/STOP button and repeat
measurement.
• Replace the batteries with new ones.
If the problem continues, contact your OMRON retail outlet
or distributor.
Arm cuff AC adapter
Arm circumference
22 - 42 cm
Easy Cuff L
9911729-4
(Model: HEM-RML31)
Adapter S
9515336-9
Adapter UK
9983666-5
USB cable
USB cable
9517499-4
Product description Automatic Blood Pressure Monitor
Model OMRON M3 IT (HEM-7131U-E)
Display LCD Digital Display
Measurement
method
Oscillometric method
Measurement range Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/min.
Accuracy Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Memory 60 measurements with date and time for each user (1 and 2)
Rating DC6V 4W
Power source 4 “AA” batteries 1.5V or optional AC adapter
(Adapter S-9515336-9, INPUT AC100-240V 50/60Hz 0.12A)
(Adapter UK-
9983666-5
, INPUT AC100-240V 50/60Hz 15VA)
Battery life Approx. 1000 measurements (using new alkaline batteries)
Applied part = Type BF
Protection against
electric shock
Internally powered ME equipment (When using only the
batteries)
= Class II ME equipment (Optional AC adapter)
Operating
temperature/
humidity
+10 to +40°C / 30 to 85% RH
Storage
temperature/
humidity/
air pressure
-20 to +60°C / 10 to 95% RH / 700 to 1060 hPa
IP classification IP 20
Weight Monitor: Approx. 280 g without batteries
Arm cuff: Approx. 170 g
Outer dimensions Monitor: Approx. 107 (w) mm x 79 (h) mm x 141 (l) mm
Arm cuff: Approx. 145 mm x 594 mm
Cuff circumference 22 to 42 cm
Cuff/ Tube material Nylon, polyester, polyvinyl chloride
Package contents Monitor, arm cuff, instruction manual, storage case,
battery set, USB cable, blood pressure pass
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile (cellular)
telephones, medical devices in use may be susceptible to electromagnetic
interference from other devices. Electromagnetic interference may result in
incorrect operation of the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with
the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has
been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate strong
electrical or electromagnetic fields, near the medical device. This may result in
incorrect operation of the device and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of
the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
Manufacturer OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
EU-representative OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
www.omron-healthcare.com
Production facility OMRON (DALIAN) CO., LTD.
Dalian, CHINA
Subsidiary OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
Gottlieb-Daimler-Strasse 10, 68165 Mannheim, GERMANY
www.omron-healthcare.de
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
Produktspezifikationen
Marke: | Omron |
Kategorie: | Blutdruckmessgerät |
Modell: | M3 IT |
Display-Typ: | LCD |
Akku-/Batterietyp: | AA |
Typ: | Automatisch |
Menge pro Packung: | 1 Stück(e) |
Zahl der Benutzer: | 2 Benutzer |
Anzahl unterstützter Akkus/Batterien: | 4 |
Tragetasche: | Ja |
Akkus/Batterien enthalten: | Ja |
Unterstützung von Positionierung: | Oberarm |
Speicherfunktion: | Ja |
Speicherkartensteckplätze: | 120 |
Pulsfrequenzmessgerät: | Ja |
Datumsaufdruck: | Ja |
Handschellengröße: | 22 - 42 cm |
Diastolischer Blutdurck: | Ja |
Systolischer Blutdruck: | Ja |
Mittleren arteriellen Blutdruck: | Ja |
Brauchst du Hilfe?
Wenn Sie Hilfe mit Omron M3 IT benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten
Bedienungsanleitung Blutdruckmessgerät Omron
30 August 2024
30 August 2024
30 August 2024
25 August 2024
25 August 2024
25 August 2024
24 August 2024
24 August 2024
24 August 2024
23 August 2024
Bedienungsanleitung Blutdruckmessgerät
- Blutdruckmessgerät Clatronic
- Blutdruckmessgerät Tristar
- Blutdruckmessgerät Medion
- Blutdruckmessgerät Nedis
- Blutdruckmessgerät SilverCrest
- Blutdruckmessgerät Panasonic
- Blutdruckmessgerät Beper
- Blutdruckmessgerät Camry
- Blutdruckmessgerät MX Onda
- Blutdruckmessgerät Trisa
- Blutdruckmessgerät AEG
- Blutdruckmessgerät Ardes
- Blutdruckmessgerät Melissa
- Blutdruckmessgerät Taurus
- Blutdruckmessgerät Technaxx
- Blutdruckmessgerät ADE
- Blutdruckmessgerät Beurer
- Blutdruckmessgerät Cresta
- Blutdruckmessgerät König
- Blutdruckmessgerät Medisana
- Blutdruckmessgerät Thomson
- Blutdruckmessgerät Braun
- Blutdruckmessgerät Inventum
- Blutdruckmessgerät Solac
- Blutdruckmessgerät Pyle
- Blutdruckmessgerät Topcom
- Blutdruckmessgerät Nokia
- Blutdruckmessgerät Hyundai
- Blutdruckmessgerät Orava
- Blutdruckmessgerät Withings
- Blutdruckmessgerät Geratherm
- Blutdruckmessgerät ProfiCare
- Blutdruckmessgerät Homedics
- Blutdruckmessgerät Lanaform
- Blutdruckmessgerät Soehnle
- Blutdruckmessgerät Citizen
- Blutdruckmessgerät A And D Medical
- Blutdruckmessgerät Aponorm
- Blutdruckmessgerät Boso
- Blutdruckmessgerät Ecomed
- Blutdruckmessgerät Elta
- Blutdruckmessgerät Fysic
- Blutdruckmessgerät Heine
- Blutdruckmessgerät Laica
- Blutdruckmessgerät Levita
- Blutdruckmessgerät Microlife
- Blutdruckmessgerät Rossmax
- Blutdruckmessgerät Salvatec
- Blutdruckmessgerät Sanitas
- Blutdruckmessgerät Sanotec
- Blutdruckmessgerät Tensoval
- Blutdruckmessgerät Terraillon
- Blutdruckmessgerät Veroval
- Blutdruckmessgerät Visomat
- Blutdruckmessgerät Sencor
- Blutdruckmessgerät IHealth
- Blutdruckmessgerät Scala
- Blutdruckmessgerät Optimum
- Blutdruckmessgerät Hartmann
- Blutdruckmessgerät Medel
- Blutdruckmessgerät OBH Nordica
- Blutdruckmessgerät Kogan
- Blutdruckmessgerät Innoliving
- Blutdruckmessgerät TrueLife
- Blutdruckmessgerät Neno
- Blutdruckmessgerät Gima
- Blutdruckmessgerät Oregon Scientific
- Blutdruckmessgerät Duronic
- Blutdruckmessgerät Eks
- Blutdruckmessgerät Inovalley
- Blutdruckmessgerät Kinetik
- Blutdruckmessgerät Magic Care
- Blutdruckmessgerät MTLogic
- Blutdruckmessgerät My Life My Shop
- Blutdruckmessgerät Sendo
- Blutdruckmessgerät Silvergear
- Blutdruckmessgerät SunTech
- Blutdruckmessgerät Time 2
- Blutdruckmessgerät Transtek
- Blutdruckmessgerät Uebe
- Blutdruckmessgerät Welby
- Blutdruckmessgerät Welch Allyn
- Blutdruckmessgerät Etekcity
- Blutdruckmessgerät A&D
- Blutdruckmessgerät Oro-Med
- Blutdruckmessgerät Dr. Senst
- Blutdruckmessgerät AnD
- Blutdruckmessgerät Bioland
- Blutdruckmessgerät Bintoi
Neueste Bedienungsanleitung für -Kategorien-
16 Oktober 2024
4 Oktober 2024
28 September 2024
28 September 2024
28 September 2024
28 September 2024
17 September 2024
17 September 2024
15 September 2024
12 September 2024