Omron HEM-7121 Bedienungsanleitung

Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für Omron HEM-7121 (2 Seiten) in der Kategorie Blutdruckmessgerät. Dieser Bedienungsanleitung war für 56 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet

Seite 1/2
1 to 2 cm
D
F
E
G
I
H
L
M
N
KQ
O
P
R
S
T
A
B
C
Thank you for purchasing the OMRON HEM-7121 Automatic Blood Pressure
Monitor.
The OMRON HEM-7121 is a compact, fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the device uses its advanced
“IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use.
Please follow this instruction manual thoroughly for your safety.
Please keep for future reference. For specific information about your own blood
pressure, CONSULT YOUR DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Important Safety Information
Introduction
1. Overview
Monitor:
K. Memory symbol
L. Systolic blood pressure
M. Diastolic blood pressure
N. Low battery symbol
O. Cuff wrapping guide
P. Heartbeat symbol
1. Flashes during measurement.
2. If flashing after measurement
completed or when viewing
results stored in the memory,
indicates blood pressure out of
recommended range*.
2.1 Installing/Replacing the Batteries
2. Preparation
1.
2.
Remove the battery cover.
Insert four “AA” batteries
as indicated in the battery
compartment and then
replace the battery cover.
Notes:
If the low battery symbol ( ) appears on the display, turn the
monitor off then replace all batteries at the same time.
The measurement values continue to be stored in memory even after
the batteries are replaced.
The supplied batteries may have a shorter life.
Disposal of used batteries should be carried out in accordance with the
national regulations for the disposal of batteries.
The bottom edge of the arm
cuff should be 1 to 2 cm
above the elbow. The air
tube is centred on the
middle of your inner arm.
3.
3.3 Taking a Measurement
3.4 Using the Memory Function
Notes:
When you take a measurement on the right
arm, the air tube will be at the side of your
elbow. Be careful not to rest your arm on
the air tube.
Notes:
To cancel a measurement, press the START/STOP button to release
the air in the arm cuff.
Remain still and do not talk while taking a measurement.
Press the START/STOP button.
The arm cuff will start to inflate automatically.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated at
comfortable room temperature.
Sit on a chair with your legs uncrossed and
your feet flat on the floor.
Sit with your back and arm being supported.
The arm cuff should be placed on your arm
at the same level as your heart.
The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
measurement. If the values between both arms differ substantially,
please check with your doctor which arm to use for your
measurement.
1.
The monitor automatically stores the result up to 30 sets.
Note: If the memory is full, the monitor will delete the oldest reading.
START INFLATE DEFLATE
COMPLETE
Automatic Blood
Pressure Monitor
Model HEM-7121
Instruction Manual
A. Display
B. START/STOP button
C. Memory button
D. Battery compartment
E. AC adapter jack (for optional
AC adapter)
F. Air Jack
Q. Irregular heartbeat symbol
R. Movement error symbol
S. Pulse display and Memory
number
T. Deflation symbol
* Note: If your systolic or diastolic
pressure is outside the standard
range (above 135/85 mmHg)
the Heartbeat symbol ( ) will
blink. Please refer to Section
3.3.
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Notes:
The monitor will not inflate above 299 mmHg.
Do not apply more pressure than necessary.
Cuff Wrapping Guide
If you move during measurement, the movement
error symbol ( ) will appear on the display.
Keep still and repeat the measurement.
Display:
5342192-4C
G. Arm Cuff (Arm circumference
22 - 42 cm)
H. Air plug
I. Air tube
Arm cuff:
Your blood pressure monitor includes an
irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement.
The irregular heartbeat algorithm automatically
determines if the measurement is usable or
needs to be repeated. If the measurement
3. Using Your Monitor
1. Insert the air plug into the air jack
securely.
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your
upper arm. Do not place the arm cuff over thick clothes.
2. Remove the arm cuff.
3. Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries to
return to the condition prior to taking the blood pressure
measurement.
Note: This reading is NOT reliable due to the incorrect wrapping of the
cuff. Please wrap the cuff again, taking care to wrap it correctly
and take the measurement again. When the is displayed, the
cuff is correctly wrapped tightly enough on the arm and the
reading is accurate and reliable.
The Cuff Wrapping Guide is a unique feature that indicates if the cuff
is not wrapped tightly enough around the arm. Even when the
appears on the display, a blood pressure reading will be taken.
2. Put your arm through the cuff
loop.
Close the fabric fastener
FIRMLY.
Press the button.
The Memory number appears for a second before the pulse rate is
displayed. The newest set is numbered “1”.
1.
Press the button repeatedly to view the readings
stored in memory.
2.
To View the Readings Stored in Memory
Cuff Wrapping Guide:
To Delete All the Values Stored in Memory
Note: You cannot partially delete the stored readings.
When the memory symbol ( ) appears, first press the Memory
button. Then while holding it down, press the START/STOP
button simultaneously for more than 3 seconds.
results are affected by irregular heartbeats but the result is valid, the
result is shown together with the irregular heartbeat symbol ( ). If
the irregular heartbeats cause the measurement to be invalid, no
result is shown. If the irregular heartbeat symbol ( ) is shown after
you have taken a measurement, repeat the measurement. If the
irregular heartbeat symbol ( ) is shown frequently, please make
your doctor aware of it.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
If your systolic or diastolic pressure is outside
the standard range, the heartbeat symbol will
blink when the measurement result is displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Important:
This criteria is for home blood pressure measurement.
Always consult your doctor. Self-diagnosis of measurement
results and self-treatment are dangerous.
Notes:
If your systolic or diastolic pressure is outside the
standard range, the heartbeat symbol will blink
when the measurement result is displayed.
The cuff wrapping guide result appears on the
display with the measurement values.
If there are no measurements results stored in
the memory, the screen to the right is displayed.
Warning: Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
(General Usage)
• DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your doctor. Only a
doctor is qualified to diagnose and treat High Blood Pressure.
• Consult your doctor before using this monitor for any of the following
conditions: common arrhythmias such as atrial or ventricular premature beats
or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age,
pregnancy, pre-eclampsia, renal diseases.
Note that PATIENT motion, trembling, shivering may affect the measurement
reading.
• Do not use this monitor on the injured arm or the arm under medical
treatment.
• Do not apply the arm cuff on the arm while being on an intravenous drip or
blood transfusion.
• Consult your doctor before using this monitor on the arm with an arterio-
venous (A-V) shunt.
• Do not use this monitor with other medical electrical (ME) equipment
simultaneously. This may result in incorrect operation of the monitor and/or
cause an inaccurate reading.
• Do not use this monitor in the area the HF surgical equipment, MRI, or CT
scanner exists, or in the oxygen rich environment. This may result in incorrect
operation of the monitor and/or cause an inaccurate reading.
• The air tube or the AC adapter cable may cause accidental strangulation in
infants.
• Contains small parts that may cause a choking hazard if swallowed by
infants.
• Stop using this monitor and consult your doctor if you experience skin
irritation or other troubles.
(AC Adapter (optional) Usage)
• Do not use the AC adapter if this monitor or the power cord is damaged. Turn
off the power and unplug the power cord immediately.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
multiple-tap.
• Never plug in or unplug the power cord from the electric outlet with wet
hands.
Caution: Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• Consult your doctor before using this monitor for any of the following
conditions:
- If you have had a mastectomy.
- People with severe blood flow problems or blood disorders as cuff inflation
can cause bruising.
• Do not take measurements more than necessary. It may cause bruising due
to blood flow interference.
• If there are any abnormalities during the measurement, remove the arm cuff.
• Do not use this monitor on infants or persons who cannot express their
intentions.
• Do not inflate the arm cuff more than necessary.
• Do not use this monitor for any purpose other than measuring blood
pressure.
• Use only the approved arm cuff for this monitor. Use of other arm cuffs may
result in incorrect measurement results.
• During measurement, make sure that no mobile phone or any other electrical
devices that emit electromagnetic fields is within 30cm of this monitor. This
may result in incorrect operation of the monitor and/or cause an inaccurate
reading.
• Do not disassemble the monitor or arm cuff. This may cause an inaccurate
reading.
• Do not use in a location with moisture, or a location where water may splash
on this monitor. This may damage this monitor.
• Do not use this monitor in a moving vehicle (car, airplane).
(Battery Usage)
• Do not insert the batteries with their polarities incorrectly aligned.
• Use only four “AA” alkaline or manganese batteries with this monitor. Do not
use other types of batteries. Do not use new and used batteries together.
• Remove the batteries if this monitor will not be used for 3 months or more.
• Use the battery within recommended period mentioned to it.
(AC Adapter (optional) Usage)
• Fully insert the power plug.
• When disconnecting the power plug, do not pull the power cord. Be sure to
hold the power plug.
• When handling the power cord, observe the following:
Do not damage. Do not break it.
Do not tamper with it. Do not forcibly bend or pull.
Do not twist. Do not bundle during use.
Do not pinch. Do not place under heavy objects.
• Wipe the dust off from the power plug.
• Disconnect the power plug if the product will not be used for a long period of
time.
• Disconnect the power plug before starting maintenance.
• Use only the original AC adapter designed for this monitor. Use of
unsupported adapters may damage and/or may be hazardous to this monitor.
General Precautions
• Do not forcibly bend the arm cuff or bend the air tube excessively.
Do not fold or kink the air tube while taking a measurement. This may cause
harmful injury by interrupting blood flow.
• To unplug the air plug, pull on the air plug at the connection with the monitor,
not the tube itself.
• Do not apply strong shocks and vibrations to or drop the monitor and arm
cuff.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the “6. Technical Data”.
• Read and follow the “Correct Disposal of This Product” in the Technical Data
Section when disposing of this monitor and any used accessories or optional
parts.
Do not use this monitor outside the specified environment. It may cause an
inaccurate reading.
• Dispose of this monitor, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
• Please check (for example, by observation of the limb concerned) if this
monitor is not causing a prolonged impairment of PATIENT blood circulation.
• If this monitor is stored at the maximum or minimum storage and transport
temperature and is moved to an environment with a temperature of 20°C, we
recommend waiting for approx. 2 hours before using the monitor.
Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and
eating for 30 minutes before taking a measurement.
2. Rest for at least 5 minutes before taking the measurement.
3. Stress raises blood pressure. Avoid taking measurements during
stressful times.
4. Measurements should be taken in a quiet place.
5. Remove tight-fitting clothing from your arm.
HEM-7121-AU_C_M05_180628.pdf
Important information regarding Electro Magnetic Compatibility (EMC)
HEM-7121 manufactured by OMRON HEALTHCARE Co., Ltd. conforms
to EN60601-1-2:2015 Electro Magnetic Compatibility (EMC) standard.
Further documentation in accordance with this EMC standard is available at
http://www.omronhealthcare-ap.com/emc-information.
Refer to the EMC information for HEM-7121 on the website.
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
waste for disposal.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
4. Troubleshooting and Maintenance 6. Technical Data
5. Optional Parts
4.1 The Icons and Error Messages
4.2 Troubleshooting
4.3 Maintenance
Problem
Calibration and Service
Cause Remedy
Do not subject the monitor and the cuff to extreme
temperatures, humidity, moisture or direct sunlight.
Do not fold the cuff or tubing tightly.
Do not inflate the arm cuff over 299 mmHg.
Do not disassemble the monitor or components.
Do not subject this monitor to strong shocks or vibrations
(for example, dropping the monitor on the floor).
Do not use volatile liquids to clean this monitor.
Do not wash the arm cuff or immerse it in water.
Do not use petrol, thinners or similar solvents to clean
the arm cuff.
Do not carry out repairs of any kind yourself. If a defect
occurs, consult your local OMRON representative.
The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
It is generally recommended to have the device
inspected every two years to ensure correct functioning
and accuracy. Please consult your local OMRON
representative.
• The monitor should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and neutral soap to clean
the arm cuff.
Automatic Blood Pressure Monitor
HEM-7121
LCD Digital Display
Oscillometric method
Pressure: 0 to 299 mmHg
20 to 280 mmHg
40 to 180 beats/ min.
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
30 Measurements
Monitor: IP20
Optional AC adapter: IP22
DC6V 4W
Continuous operation
4 “AA” batteries 1.5V
or Optional AC adapter (INPUT AC100-240V
50-60Hz 0.12-0.065A)
Approx. 1000 measurements
(using new alkaline batteries)
Monitor: 5 years
Cuff: 1 year
Optional AC adapter: 5 years
Type BF (Cuff)
Internally powered ME equipment (When using only
the batteries)
Class II ME equipment (Optional AC adapter)
+10 to +40°C / 15 to 90% RH (non-condensing) /
700 to 1060 hPa
-20 to +60°C / 10 to 95% RH (non-condensing) /
700 to 1060 hPa
Approx. 250g without batteries
Approx. 170g
Approx. 103 (w) mm × 80 (h) mm × 129(l) mm
Approx. 145 mm × 594 mm
(Cuff: arm circumference 22 to 42 cm)
Nylon, polyester, polyvinyl chloride
Main unit, arm cuff, instruction manual, battery set
Note: Please check with your local OMRON representatives for the
appropriate optional part models.
The reading is
extremely low (or
high).
Arm cuff deflates too
soon.
Arm cuff pressure
does not rise.
Cannot measure or
readings are too low
or too high.
Arm cuff not applied
correctly.
Movement or talking
during measurement.
Clothing is interfering
with the arm cuff.
The air tube is not
securely connected
into the air jack.
Air is leaking from the
arm cuff.
The arm cuff is loose.
The arm cuff has not
been inflated
sufficiently.
Apply the arm cuff
correctly.
Refer to section 3.1.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Make sure that the air
tube is connected
securely.
Refer to section 3.1.
Replace the arm cuff
with a new one.
Refer to Chapter 5.
Apply the cuff correctly
so that it is firmly
wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it
is 30 to 40 mmHg above
your previous
measurement result.
Refer to section 3.3.
1. Unplug the air plug from the air jack.
2. Gently fold the air tube into the arm cuff.
4.4 Storage
1.
2.
Using the Optional AC Adapter
Insert the AC adapter plug
into the AC adapter jack on
the rear side of your monitor.
Plug the AC adapter into an
electrical outlet.
Product Description
Model
Display
Measurement Method
Measurement Range
Blood pressure
measurement range
Pulse measurement
range
Accuracy
Inflation
Deflation
Memory
IP classification
Rating
Operation mode
Power Source
Battery Life
Durable Period
Applied Part
Protection Against
Electric Shock
Operating conditions
Storage/ Transport
conditions
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Cuff/ Tube Material
Package Contents
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then remove the AC adapter plug from the
monitor.
The batteries are low.
The batteries are
exhausted.
Air plug disconnected.
The arm cuff was inflated
above 299 mmHg
when inflating the cuff
manually.
Movement during
measurement.
Clothing is interfering with
the arm cuff.
Device error.
Irregular heartbeats are
detected.
Cuff is not applied correctly.
Remove the arm cuff. Wait
2 - 3 minutes and then take
another measurement.
Repeat the steps in section
3.3. If this error continues to
appear, contact your doctor.
Movement during
measurement.
Carefully read and repeat
the steps in section 3.3.
Apply the arm cuff correctly.
Refer to section 3.1.
You should replace them
with new ones ahead of
time.
Refer to section 2.1.
You should replace them
with new ones at once.
Refer to section 2.1.
Insert the plug securely.
Refer to section 3.1.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.1.
Air is leaking from the arm
cuff.
Replace the cuff with the
new one.
Refer to Chapter 5.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the cuff manually
until it is 30 to 40 mmHg
above your previous
measurement result.
Refer to section 3.3.
Do not inflate the cuff above
299 mmHg.
Refer to section 3.3.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Contact your local OMRON
representative.
Cause RemedyError Display
• Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your local OMRON representative.
Nothing happens
when you press the
buttons.
Other problems.
The batteries are
empty.
The batteries have
been inserted
incorrectly.
Replace the batteries
with new ones.
Refer to section 2.1.
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.1.
Notes:
Do not bend the air tube excessively.
Do not store your monitor in the following situations:
- If your monitor and other components are wet.
- Locations exposed to extreme temperatures, humidity, direct
sunlight, dust or corrosive vapours.
- Locations exposed to vibrations, shocks or where it will be at an
angle.
Note: Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the monitor and any
used accessories or optional parts.
Movement during
measurement and the arm
cuff has not been inflated
sufficiently.
Made in Vietnam
Manufacturer
EU-representative
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, KYOTO
617-0002 JAPAN
Production facility
OMRON HEALTHCARE MANUFACTURING VIETNAM
CO., LTD.
Binh Duong Province, VIETNAM
Asia Pacific HQ
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
www.omronhealthcare-ap.com
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
www.omron-healthcare.com
Wide Range Soft Cuff
HEM-RML31
Arm circumference
22 - 42 cm
Small Cuff
HEM-CS24
Arm circumference
17 - 22 cm
In case of any of the below problems occur during measurement, first
check that no other electrical device is within 30cm. If the problem
persists, please refer to the table below.
To protect your monitor from damage, please observe the
following:
AC Adapter
HHP-OH01
Note: Make sure to use an easily accessible power socket in which to
connect and disconnect the AC adapter.
Notes:
Subject to technical modification without prior notice.
• In the clinical validation study, the 5th phase was used on 85 subjects for
determination of diastolic blood pressure.
• This monitor is clinically investigated according to the requirements of ISO
81060-2:2013.
• This monitor has not been validated for use on pregnant patients.
• IP classification is degrees of protection provided by enclosures in
accordance with IEC 60529.
This monitor and optional AC adapter are protected against solid foreign
objects of 12.5 mm diameter and greater such as a finger. The optional AC
adapter is protected against oblique falling water drops which may cause
issues during a normal operation.
Symbols description
Applied part - Type BF
Degree of protection against electric shock (leakage
current)
Class II equipment. Protection against electric shock
CE Marking
Serial number
LOT number
Temperature limitation
Humidity limitation
Atmospheric pressure limitation
Indication of connector polarity
For indoor use only
OMRON's trademarked technology for blood
pressure measurement
Identifier of cuffs compatible for the device
Cuff positioning indicator for the left arm
Marker on the cuff to be positioned above the artery
Range pointer and brachial artery alignment position
Not made with natural rubber latex
Range indicator of arm circumferences to help
selection of the correct cuff size.
Need for the user to consult this instruction manual.
Need for the user to follow this instruction manual
thoroughly for your safety.
,
,
,
,
,
Symbols description
Alternating current
Date of manufacture
Arm circumference
Product production date is integrated in the Serial number, which is placed
on the product and/or sales package: the first 4 digits mean year of
production, the next 2 digits mean month of production.
Direct current
Depending on the product, some of the above symbols may not be applicable
on the product.
Efficiency Level of power supply
SMPS (Switch mode power supply unit)
RCM compliance mark, which indicates
compliance with electrical safety, EMC, EME &
telecommunications requirements in Australia, as
applicable to the product.
SMPS incorporating a short-circuit-proof safety
isolating transformer
(inherently or non-inherently)
,

Produktspezifikationen

Marke: Omron
Kategorie: Blutdruckmessgerät
Modell: HEM-7121
Display-Typ: LCD
Breite: 103 mm
Tiefe: 129 mm
Gewicht: 250 g
Produktfarbe: Silber
Akku-/Batterietyp: AAA
Höhe: 80 mm
Menge pro Packung: 1 Stück(e)
Zahl der Benutzer: 1 Benutzer
Anzahl unterstützter Akkus/Batterien: 4
Unterstützung von Positionierung: Oberarm

Brauchst du Hilfe?

Wenn Sie Hilfe mit Omron HEM-7121 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten




Bedienungsanleitung Blutdruckmessgerät Omron

Bedienungsanleitung Blutdruckmessgerät

Neueste Bedienungsanleitung für -Kategorien-