KERN TGC 1K-3 Bedienungsanleitung

KERN Waage TGC 1K-3

Lesen Sie kostenlos die 📖 deutsche Bedienungsanleitung für KERN TGC 1K-3 (3 Seiten) in der Kategorie Waage. Dieser Bedienungsanleitung war für 4 Personen hilfreich und wurde von 2 Benutzern mit durchschnittlich 4.5 Sternen bewertet

Seite 1/3
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E- -Mail: info@kern sohn.com
Tel.: +49- - -0 [0]7433 9933
Fax: 9933 149 - -+49 [0]7433 -
internet: - www.kern sohn.com
Operating manual
Electronic pocket scales
KERN TGC
Version 1.0
2018-07
GB
TGC-BA-e- 1810
TGC- -e- 2BA 1810
GB
Version 1.0 2018- 07
Operating manual
1. Technical data
KERN -2 -1-A -3 TGC 150 TGC 500 TGC 1K
Item no./ Type -2-A -1-A -3-A TTGC 150 TGC 500 TGC 1K
Readability (d) 0.1 g 1 g 0.01 g
Weighing range 000 g(max) 500 150 g g 1
Adjustment weight, not added (class) / 150 g 00 g (M3) 0 g (M3) 50 g (M2) 5 100
Operating temperature 0+ °....+ 40°C
Humidity of air 20% 8 - ∼0% air humidity (non condensing)
Batteries 2 x CR2032
2. Scope of delivery
• Pocket scale
• Balance cover
• Batteries (2 x CR2032)
• Operating manual
3. Keyboard overview
Switch-over weighing unit
Background illumination of the display on/off
Taring / Setting to zero
Switch on/off
4. Operation
Weighing:
 Place the balance on a firm, level surface.
 Press ON/OFF button, the balance completes a self check. As soon as the weight display appears,
the balance is ready for weighing. If required, set to zero by using TARE button.
 Place load on pan and read the weighing result.
 If necessary, use the button to switch over into the available weighing units.UNIT
3 - -e- TGC BA 1810
Taring:
 Deposit empty weighing container.
 Press the TARE button, the zero display disappears.
 Weigh the material, the net weight will be displayed.
 When the balance is unloaded the saved taring value is displayed with negative sign.
 To delete the stored tare value, remove load from weighing pan and press the TARE button.
5. Adjustment
 Arrange adjustment weight (see chap. 1).
Info about adjustment weights can be found on the Internet at: http://www.kern-sohn.com
 Ensure that there are no objects on the weighing pan.
 In weighing mode keep the button pressed until UNIT <CAL> appears.
 Press the ON/OFF button again and wait until the weight value of the required adjustment weight
will appear flashing.
 Place adjustment weight. Wait until <PASS>, followed by the weight value of the placed
adjustment weight will be displayed. After that the adjustment will be finished successfully.
 In case of a diverging display the adjustment has failed. Repeat adjustment process.
6. Important Notes
• Do not overload balance.
• The occurrence of electromagnetic fields may result in major display deviations. Then change
location.
• Avoid high air humidity, vapours and dust.
• Do not bring the balance directly in contact with liquids.
• Do not leave permanent load on the weighing pan.
• Never operate balance in explosive environment. The serial version is not
explosion protected.
• Check the balance regularly with external test weights.
• Warranty claims will be voided when the housing is opened.
7. Declaration of conformity
The current EC/EU Conformity declaration can be found online in:
www.kern-sohn.com/ce


Produktspezifikationen

Marke: KERN
Kategorie: Waage
Modell: TGC 1K-3

Brauchst du Hilfe?

Wenn Sie Hilfe mit KERN TGC 1K-3 benötigen, stellen Sie unten eine Frage und andere Benutzer werden Ihnen antworten




Bedienungsanleitung Waage KERN

Bedienungsanleitung Waage

Neueste Bedienungsanleitung für -Kategorien-